pITA - 1

pITA Rapid Exchange Neurovascular PTA Balloon Catheter rapid exchange PTA balloon for intracranial stenosis • Long soft distal tip with 0.016” thin profile • Low balloon crossing profile: 0.023” - 0.026” 0.016” entry profile embedded x-ray markers • Compatible with 0.014” guide wire • “Zero profile” x-ray markers • Total length: 160 cm • Diameter sizes from 1.25 to 4.0 mm Balloon lengths from 10 to 40 mm • Nominal pressure at 6 ATM* Rated burst pressure at 14 ATM* • Non-/semi-compliant** *For pITA with diameter of 1.25 mm and 1.5 mm: Nominal pressure at 10 ATM and rated burst pressure at 18 ATM **For pITA with diameter of 1.25 mm and 1.5 mm: Non-compliant; For all others: Semi-compliant

Open the catalog to page 1
pITA - 2

pITA Rapid Exchange Neurovascular PTA Balloon Catheter Catheter Length: 160 cm Distal Marker* Proximal Marker* A Proximal Shaft Diameter: 1.9F / 0.63mm Distal Shaft Diameter: 2.5F / 0.83mm * The pITA with diameter of 1.25 mm x-ray 1.5 mm *The diameter 1.25 mm/1.5 mm balloon catheters have one centrally located and marker only. have only one centrally located x-ray marker only. Rated Burst Pressure [atm] Rated Burst Pressure [atm] phenox GmbH | Lise-Meitner-Allee 31 | D-44801 Bochum, Germany | Tel: +49 (0) 234 - 36 919 - 0 | Fax: +49 (0) 234 - 36 919 - 19 www.phenox.net | info@phenox.info