Specifications:
The Porter Vanguard Manifold Model 4222CXB is designed for a flow capacity ranging from 0 to 100 SLPM (nitrogen) with a pressure drop not exceeding 5 PSIG. It requires a power supply of 110Vac.

Manifold Features:
The manifold system incorporates the cross + protection system, adhering to NFPA 99C standards for nonflammable medical gas systems. It is suitable for Level II facilities using nitrous oxide and oxygen. Key features include:
  • Preset operating regulators for seamless switch over.
  • CGA gas specified inlets and outlets with O-ring seals and integral check valves.
  • High/low pressure switches preset at 20% above and below normal operating pressures.
  • Compliance with NFPA-99C standards.
  • Visual indicators for current supply tank status.
  • Automatic changeover to reserve tank with memory and empty tank indicator.
  • Failsafe protection to shut down nitrous oxide supply if oxygen pressure is lost.
  • Cross + Protection diameter indexed tubing connections.
  • Swivel copper tubing connections for flexible piping configurations.
  • Pressure relief valves set at 50% of normal operating pressure.
  • Audible and visual alerts for normal or high/low system pressure conditions.
  • Normally open electronic control valve ensuring continuous gas flow during power loss.
  • Manual override option available.

Contact Information:
For more details, contact Porter Instrument at 888-723-4001 or visit www.porterinstrument.com or email [email protected].
See more

Catalog excerpts

Porter Vanguard Manifold System Spec Sheet FM-1370 Rev--1

Porter Instrument Vanguard Manifolds, Zone Valves and Outlet Stations utilize the cross + protection system. The system meets or exceeds the guidelines established by the National Fire Protection Association for Nonflammable Medical Gas Systems, NFPA 99C. Product Application: Manifold Gas System for Nitrous Oxide and Oxygen for Level II facility. Porter Vanguard Wall Alarm Panel Flow Capacity: Any flow range from 0-100 SLPM (nitrogen) with a pressure drop of no greater than 5 PSIG Power Requirements: 110Vac Manifold Features: • Preset operating regulators for easy sw i t c h o ver • CGA gas specified inlets and outlets with O-ring seal back up and integral check valves • High/Low pressure switches preset 20% above and below normal operating pressures • Manifold complies with NFPA-99C • Visual indicator of current supply tank • Automatic changeover to reserve tank with memory of tank usage and empty tank indicator • Failsafe protection (loss of oxygen supply pressure will shut down Nitrous Oxide supply) • Cross + Protection diameter indexed tubing connections •Swivel, copper tubing connections to allow easy connection of piping in the up or down configuration • Pressure relief valves downstream of check valves pre-set at 50% of normal operating pressure • Audible and visual display indicating normal or high/low system pressure condition • Normally open electronic control valve that provides continuous gas flow with loss of electrical power. • Manual over-ride option www.porterinstrument.com [email protected]

 Open the catalog to page 1
Porter Vanguard Manifold System Spec Sheet FM-1370 Rev--2

ENGINEERING YOUR SUCCESS 888-723-4001

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.