Product catalog summary
Overview: The Dewpoint Monitor is a device designed to measure and control the dew point in air systems. It features a range of models suitable for various applications, including medical and laboratory settings.
Specifications:
  • Dew Point Range: Varies by model, from -148°F to +54°F.
  • Sensor Output: 0V-5V.
  • Alarm Indication: Audible and visual alarms with a red alarm light and push-to-test/acknowledge features.
  • Alarm Output: 10 AMP @ 115 VAC rated normally closed contact.
  • Alarm Horn: 95+ dB(A) at 120 VAC, measured at 24 inches.
  • Operating Pressure: 10-150 psig with a minimal flow rate.
  • Operating Temperature: 32°F to 131°F.
  • Power Requirements: 110-120VAC with a Type FNM 1.25-amp fuse.
  • Dimensions and Weight: 9.1” x 5.5” x 5”, weighing 5 lbs.
  • Enclosure: UL Listed Polycarbonate NEMA-4X Electrical Box.
Features:
  • Patented auto calibration for accuracy.
  • Polymer sensor technology for fast response times.
  • Sensor is resistant to water, particulate contamination, oil vapor, and most chemicals.
  • Visual and audible alarms for high dew point, including dry contacts.
  • Dew point demand purge control feature for desiccant air dryers.
Model Specifics:
  • PDPM1001JJ: Dew point range of -60°F to +54°F, high alarm at +35°F, with purge modes.
  • PDPM1001AJ: Similar to JJ model but without a power cord.
  • PDPM1001JJL: Commonly used with refrigerated dryers, high alarm at +40°F.
  • PDPM2001AJ: Dew point range of -148°F to +32°F, specified high alarm.
  • PDPM1004AJ: Dew point range of -60°F to +54°F, high alarm at -20°F.
Application Specifics:
  • Medical: Dew point range of -70°F to +54°F, high alarm at +35°F.
  • Laboratory: Dew point range of -70°F to +54°F, high alarm at -20°F.
Interface and Controls:
  • Main page displays pressure and dew point levels, with color changes indicating alarm status.
  • System status page shows operating status, run hours, NFPA alarms, and service alerts.
See more

Catalog excerpts

Dewpoint Monitor-1

Dewpoint Monitor Engineering Specifications Dew Point Range: see chart below Patented auto calibration to ensure accuracy Polymer sensor technology for fast response times Sensor is immune to water, particulate contamination, oil vapor and most chemicals Visual and audible alarms for high dew point (includes dry contacts) Dew point demand purge control feature to control desiccant air dryers Dew Point Range Enclosure High Alarm Contact Open for Purge Contact Closed for rate fixed mode Purge Saver mode

 Open the catalog to page 1
Dewpoint Monitor-2

Integrated Dewpoint Monitor For Air Systems with HMI or PBMI controls MAIN PAGE  Displays Pressure level and Dew Point level  Pressure level will change from green to orange when   high Dew Point level will change from green to orange when approaching alarm level; will change from orange to red when in alarm System Status, Alarm Status, Service Info and Trends buttons will redirect to those specific pages Patented auto calibration to ensure accuracy Polymer sensor technology for fast response times Sensor is immune to water, particulate contamination, oil vapor and most chemicals...

 Open the catalog to page 2

All Powerex catalogs and technical brochures

  1. Rotary Tooth

    4  Pages

  2. AS/AD Models

    1  Page

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.