Scroll Medical Air Package
1 /12Pages

Scroll Medical Air Package

Scroll Medical Air Package
1 /12Pages

Catalog excerpts

Scroll Medical Air Package-1

Scroll Medical Air Package Please read and save these instructions. Read carefully before attempting to assemble, install, operate or maintain the product described. Protect yourself and others by observing all safety information. Failure to comply with instructions could result in personal injury and/or property damage! Retain instructions for future reference. Description The Powerex scroll medical air package is designed to provide medical breathing air for hospital and medical institutes. This system meets NFPA 99 requirements for Level 1 breathing air. Safety Guidelines This manual contains information that is very important to know and understand. This information is provided for SAFETY and to PREVENT EQUIPMENT PROBLEMS. To help recognize this information, observe the following symbols. Danger indicates an imminently hazardous situation which, if not avoided, WILL result in death or serious injury. Warning indicates a potentially hazardous situation which, if not avoided, COULD result in death or serious injury. Caution indicates a potentially hazardous situation which, if not avoided, MAY result in minor or moderate injury. Grease-lled Bearing Operating Voltages Compression Cycle Motor Overload Protection Maximum System Pressure Standard: 116 psi High Pressure: 145 psi Overpressure Protection ASME Safety Valve Factory Set and Sealed 1/2 through 1-1/2 inch NPT 80, 120, 200 Gallon ASME Rated 200 psig Meets Requirements of this Ordinance Control Panel Powerex • 150 Production Drive • Harrison, OH 45030 • USA 1-888-769-7979 • www.powerexinc.com Tank Isolation Do not operate unit if damaged during shipping, handling or use. Damage may result in unsafe conditions and cause injury or property damage. Immediately upon receipt of the system, inspect for any damage which may have occurred during shipment. Repair or replace damaged items before use. The name-plate should be checked to verify the correct model and voltage. Performance Specications Tank Sizes MS Series Powerex Multiplex Air Compressors Notice indicates important information, that if not followed, may cause damage to equipment.

 Open the catalog to page 1
Scroll Medical Air Package-2

Scroll Medical Air Package Specications (Continued) Compressor Performance (per base mount) HP 3 Discharge Pressure (PSIG) Motor Voltage NOTE: For system electrical information, see wiring diagram shipped with unit. All electrical hook-ups must be performed by a qualied electrician. Installations must be in accordance with local and national electrical codes. Disconnect, tag and lock out power source then release all pressure from the system before attempting to install, service, relocate or perform any maintenance. Use solderless terminals to connect the electric power source. PIPING Do not...

 Open the catalog to page 2
Scroll Medical Air Package-3

Scroll Medical Air Package Installation (Continued) ELECTRICAL WIRING OF UNIT SHIPPED IN SEPARATE MODULES SAFETY VALVES Provide electrical power in accordance to NEC and local codes. Connection of wiring should be performed by a qualied electrician. Our tanks are shipped from the factory with safety valves installed in the tank. The ow capacity of the safety valve is equal to or greater than the capacity of the compressor system. 1. Safety valves should be placed ahead of any possible blockage point in the system, i.e. shutoff valve. Connect each temperature switch from the compressors to contacts...

 Open the catalog to page 3
Scroll Medical Air Package-4

MEDICAL AIR SCHEMATIC (THIS IS A REPRESENTATIVE SCHEMATIC DIAGRAM FOR A LOW PRESSURE MULTIPLE SCROLL MEDICAL SYSTEM) Scroll Medical Air Package

 Open the catalog to page 4
Scroll Medical Air Package-5

Scroll Medical Air Package Maintenance Schedule COMPRESSOR Operating Hours Action needed Inspect, Replace (145 psig scroll units) Every 2500 hours or less ▲ ▲ Replace both at 5000 hours; alternate monthly Bearings (145 psig scroll units) Service Center ONLY Service Center ONLY; EVERY 5000 hours Dust Seal Inspect, Replace Pressure switch/ Transducer Conrm operation Magnetic starter Check valve Heat Insulation Pipe Replace is contact point is deteriorated PIPING SYSTEM Operating Hours Action needed Safety valve Conrm operation Every 2500 hours or less Pressure gauge Every 2500 hours or less Drain...

 Open the catalog to page 5
Scroll Medical Air Package-6

Scroll Medical Air Package Date Maintenance Required Maintenance Performed

 Open the catalog to page 6
Scroll Medical Air Package-7

Scroll Medical Air Package Maintenance Required Maintenance Performed

 Open the catalog to page 7
Scroll Medical Air Package-8

Scroll Medical Air Package Replacement Parts Illustration HIGH TEMPERATURE HIGH TEMPERATURE LAMP TEST

 Open the catalog to page 8
Scroll Medical Air Package-9

Scroll Medical Air Package Replacement Parts List Ref. No. 1 Part Number Scroll pump 2HP Scroll: 3HP Scroll: 5HP Scroll: For 145 psig Optional Units 5HP Scroll Vertical tank 80 Gallon 120 Gallon 200 Gallon Dew point monitor / probe dessicant Carbon monoxide monitor / sensor Control panel (Consult factory for parts and availability. Call 1-888-769-7979) Motor (ODP) 2HP Motor: 3HP Motor: 5HP Motor: Belt guard Front (all models) Back (all models) Dryer MSD0203 MSD0303, MTSD0303 MSD0503, MTSD0503 MST0503, MSD1004 MST1005 MST1505 MSQ1005 MSQ0504, MSD1504 MSO0505 MSH0504 MSH1506 MSN0505, MSQ1505 MSL0505...

 Open the catalog to page 9
Scroll Medical Air Package-10

Scroll Medical Air Package

 Open the catalog to page 10
Scroll Medical Air Package-11

Scroll Medical Air Package Powerex Limited Warranty – Applicable to Non-OEM Customers in the U.S. & Canada Only Warranty and Remedies. (a) General. Powerex warrants each Compressor System, Vacuum System, Vacuum Pump, Compressor Air-End, or Powerex branded Accessory (collectively “Products”, individually each a “Product”) to be free from defects in material and workmanship (“Defects”) at the date of shipment. This warranty shall apply only to Products that are purchased and used in the United States of America and in Canada. EXCEPT AS SET FORTH BELOW, NO OTHER WARRANTY, WHETHER EXPRESS OR IMPLIED,...

 Open the catalog to page 11
Scroll Medical Air Package-12

Scroll Medical Air Package (i) The warranty provided herein shall not apply to: (i) any defects arising from corrosion, abrasion, use of insoluble lubricants, or negligent attendance to or faulty operation of the Products; (ii) ordinary wear and tear of the Products; (iii) defects arising from abnormal conditions of temperature, dirt or corrosive matter; or (iv) any OEM component which is shipped by Powerex with the original manufacturer’s warranty, which shall be the sole applicable warranty for such component. Limitation of Liability. NOTWITHSTANDING ANYTHING TO THE CONTRARY HEREIN, TO THE...

 Open the catalog to page 12