Product catalog summary
Overview: The document provides detailed specifications and features of Priorclave's QCS range of top-loading autoclaves, available in 100 and 150-litre chamber capacities. These autoclaves are designed for efficient and reliable sterilization, suitable for laboratories with limited space.
Specifications: The autoclaves are available with electric or direct steam heating options. They feature a Quickseal single-action door with thermal and pressure locks, Tactrol 2 microprocessor control, automatic timed free-steaming, forced air cooling, and vacuum-assisted drying. The pressure vessel has a 10-year warranty.
Models: The document details four models: PS/QCS/EV100, PS/QCS/EV150, PS/QVA/EV100, and PS/QVA/EV150. Each model varies in chamber size, loading height, and weight, with operating ranges up to 140°C and 2.4 bar.
Installation and Support: Installation specifications include external dimensions and minimum installation space. The document emphasizes the importance of allowing access for routine servicing. Priorclave offers dedicated pre- and after-sales support, including maintenance programs, calibration, and performance qualification services.
Compliance: All QCS and QVA autoclaves comply with international standards such as BS2464-1993, PD5500:2000 Cat.3, European PED 97/23/EC, and are CE marked for BS EN 61010-2-41.
Key Features: The autoclaves are equipped with features like media warming, delayed start functions, and are supplied with stainless steel baskets. They are constructed with epoxy-coated panels treated with an anti-bacterial agent effective against bacteria and fungi, including MRSA.
See more

Catalog excerpts

QCS EV 100/150 QVA EV100/150 TOP LOADING AUTOCLAVES-1

TOP LOADING AUTOCLAVES The Priorclave QCS range of top loading autoclaves, available with 100 and 150 litre chamber capacities, has amongst its many design features simple easy loading of the media. These versatile autoclaves offer fast, reliable sterilisation and have a chamber size that will appeal to most laboratories, it is versatile enough to sterilise small and tall objects side-by-side in a single process. The standard QCS models have a compact footprint which makes this range of top loading autoclaves ideal for locations where space is at a premium and high yield is required. Depending on application, Priorclaves are available with a choice of heating options - electric or direct steam heating. Quickseal single-action door fitted with thermal and pressure locks to prevent opening at load temperature above 80oC and pressures above 0.2 bar. Tactrol 2 microprocessor control with simple and fully variable setting of process time, temperature with graphic indication of cycle status. Automatic timed free-steaming. Forced Air Cooling. Pre-cycle vacuum in which multiple Vacuum stages are programmed with interspersed heating to achieve steam penetration.1 Vacuum assisted drying.1 Pulsed freesteaming to maximise efficiency of Pre-cycle vacuum phase.1,2 Epoxy coated panels and frame members are treated with an anti-bacterial agent which is effective against all bacteria and fungi including MRSA. Media Warming and Delayed Start functions keep autoclaved media at a usable temperature ready for immediate use. Vacuum Assisted Cooling.1 Supplied with two (100L) or three (150L) stainless baskets. Pressure vessel carries a 10 year warranty and is insurance approved. Service and support direct from manufacturer (or appointed distributor). chamber size chamber material autoclave capacity polished 316 stainless steel loading height operating range model: PS/QCS/EV150 autoclave capacity chamber size chamber material polished 316 stainless steel loading height operating range model: PS/QVA/EV100 autoclave capacity chamber size chamber material polished 316 stainless steel loading height operating range model: PS/QVA/EV150 autoclave capacity Vacuum models only - PS/QVA/EV100 and EV150 Optional extra on Standard models - PS/QCS/EV100 and EV150 chamber size chamber material polished 316 stainless steel Priorclave Limited 129-131 Nathan Way, West Thamesmead Business Park London, SE28 0AB T: +44 (0)20 8316 6620 F: +44 (0) 20 8855 0616 E: [email protected] loading height operating range For Direct Steam Heated models substitute SV for EV. Integral Steam Generator models also an option.

 Open the catalog to page 1
QCS EV 100/150 QVA EV100/150 TOP LOADING AUTOCLAVES-2

Installation Specification model: PS/QCS/EV100 external dimensions 660 x 840 x 840mm (w x d x h) minimum installation 980 x 1400mm (a x b)* - see diagram 7kW - 30A single-phase heater power 10.5kW - 15A three-phase weight height from floor vent outlet - 22mm copper safety valve outlet - 15mm copper waterfill tank inlet - 15mm copper waterfill tank overflow - 15mm copper vacuum pump1 - 15mm copper drain outlet1 - 15mm copper overflow outlet1 15mm copper waterfill inlet1 alternate position on vacuum PS/QVA models only * installation dimensions include space for doors to open fully and for pipes,...