Warning Indicators and Fault Codes

Warning Indicators and Fault Codes

Warning Indicators and Fault Codes

Product catalog summary
Overview: This document provides detailed information on the warning lights and fault codes associated with the Priorclave Tactrol Controller, used in autoclaves. It includes descriptions of warning indicators, fault codes, and procedures for resetting faults.
Warning Lights: The control panel features several warning indicators that appear when active. Key indicators include:
  • SERVICE: Indicates the autoclave requires servicing after 500 cycles or six months.
  • WATER: Accompanied by fault code F004, indicates low water level.
  • O/HEAT: Accompanied by fault code F003, indicates vessel overheat.
  • FAULT: General function fault, with specific code shown in the temperature display.
  • LOCK: Lights when the thermal lock key-switch is in override.
  • LOAD: Indicates a Load Sense Probe fault with F001.
Fault Codes: Fault codes are displayed in the Red Temperature Display Window. Codes F000 to F012 are common across all versions. If multiple faults occur, the highest priority is displayed. Key fault codes include:
  • F000: Measuring system error, critical temperature issue.
  • F001: Load sense probe failure, requires thermocouple replacement.
  • F003: Overheat, temperature exceeds 142°C.
  • F004: Water level alarm, requires topping up water.
  • F006: Power failure during cycle.
  • F010: Excess air detected, affects sterilization.
  • F012: Door fault, microswitch issue.
Resetting Faults: To clear fault messages, correct the fault source first. For F004, refilling water resets the fault. For others:
  • Single Program Machines: Use the manual reset button on the control panel.
  • Multi-program Machines: Use the SETTING key to toggle positions and reset.
Recommendations: Regular servicing and monitoring of warning indicators and fault codes are essential for maintaining autoclave functionality. Contact Priorclave Service for unresolved issues.
See more

Catalog excerpts

Warning Indicators and Fault Codes-1

Priorclave Tactrol Controller Warning Lights and Fault Codes Warning Lights The control panel has a series of ‘hidden until active’ warning indicators that appear in the black bar beneath the Temperature and Time displays on the control panel. (Some of these indicators will appear in conjunction with a fault code in the temperature display) SERVICE 500 cycles, or six months have passed since the autoclave was last serviced. The engineer will cancel the message when the autoclave is serviced. This indicator does not signify a machine fault, nor does it affect the Autoclave function. Indicated along with a Low Water fault code F004 Indicates a vessel Over-heat fault along with fault code F003 Indicates a function fault the fault code will be shown in the Red Temperature Display Window. This warning will light when the thermal lock key-switch is in the override position. Indicates a Load Sense Probe fault along with F001 Fault Codes are displayed in the Red Temperature Display Window on the control panel. Fault Codes As options are developed new fault codes are added so not all of these codes are applicable dependant on the version of the Tactrol System and the options installed on the autoclave. Fault codes F000 to F012 are common to all versions. If two or more faults occur at the same time, the one with the highest priority is displayed. (F000 is the highest priority and F019 is the lowest.) If a higher priority fault is cleared it will be replaced by the next active fault, unless this too is cleared by the same action. Note: Some of these fault codes may be cleared as described at the end of this list of codes. If the fault code will not clear, or continues to re-appear then the fault cannot be corrected and you should contact Priorclave Service or your local Priorclave approved service agent for expert assistance. H:\Technical\MANUALS\Fault Indicators and Codes UK.docx

 Open the catalog to page 1
Warning Indicators and Fault Codes-2

H:\Technical\MANUALS\Fault Indicators and Codes UK.docx [email protected] Page 2 129-131 Nathan Way, London SE28 0AB.

 Open the catalog to page 2
Warning Indicators and Fault Codes-3

H:\Technical\MANUALS\Fault Indicators and Codes UK.docx [email protected] Page 3 129-131 Nathan Way, London SE28 0AB.

 Open the catalog to page 3
Warning Indicators and Fault Codes-4

H:\Technical\MANUALS\Fault Indicators and Codes UK.docx [email protected] Page 4 129-131 Nathan Way, London SE28 0AB.

 Open the catalog to page 4
Warning Indicators and Fault Codes-5

Priorclave Tactrol Controller Fault Codes and Warning Lights - from September 2018 Code Condenser Water Fault* Where the autoclave is fitted with a drain condenser this fault is triggered by detection of no water or insufficient water supply to the condenser. This acts to protect heat sensitive external drain pipework from exposure to high temperature autoclave exhaust. Activation of this fault would indicate a problem with water supply, water pressure or water inlet valve operation. Note: This fault will abort the cycle and NOT complete the sterilization process. Optional Feature – when the...