Warning Indicators and Fault Codes
1 /5Pages

Warning Indicators and Fault Codes

Warning Indicators and Fault Codes
1 /5Pages

Catalog excerpts

Warning Indicators and Fault Codes-1

Priorclave Tactrol Controller Warning Lights and Fault Codes Warning Lights The control panel has a series of ‘hidden until active’ warning indicators that appear in the black bar beneath the Temperature and Time displays on the control panel. (Some of these indicators will appear in conjunction with a fault code in the temperature display) SERVICE 500 cycles, or six months have passed since the autoclave was last serviced. The engineer will cancel the message when the autoclave is serviced. This indicator does not signify a machine fault, nor does it affect the Autoclave function. Indicated along with a Low Water fault code F004 Indicates a vessel Over-heat fault along with fault code F003 Indicates a function fault the fault code will be shown in the Red Temperature Display Window. This warning will light when the thermal lock key-switch is in the override position. Indicates a Load Sense Probe fault along with F001 Fault Codes are displayed in the Red Temperature Display Window on the control panel. Fault Codes As options are developed new fault codes are added so not all of these codes are applicable dependant on the version of the Tactrol System and the options installed on the autoclave. Fault codes F000 to F012 are common to all versions. If two or more faults occur at the same time, the one with the highest priority is displayed. (F000 is the highest priority and F019 is the lowest.) If a higher priority fault is cleared it will be replaced by the next active fault, unless this too is cleared by the same action. Note: Some of these fault codes may be cleared as described at the end of this list of codes. If the fault code will not clear, or continues to re-appear then the fault cannot be corrected and you should contact Priorclave Service or your local Priorclave approved service agent for expert assistance. H:\Technical\MANUALS\Fault Indicators and Codes UK.docx

 Open the catalog to page 1
Warning Indicators and Fault Codes-2

H:\Technical\MANUALS\Fault Indicators and Codes UK.docx [email protected] Page 2 129-131 Nathan Way, London SE28 0AB.

 Open the catalog to page 2
Warning Indicators and Fault Codes-3

H:\Technical\MANUALS\Fault Indicators and Codes UK.docx [email protected] Page 3 129-131 Nathan Way, London SE28 0AB.

 Open the catalog to page 3
Warning Indicators and Fault Codes-4

H:\Technical\MANUALS\Fault Indicators and Codes UK.docx [email protected] Page 4 129-131 Nathan Way, London SE28 0AB.

 Open the catalog to page 4
Warning Indicators and Fault Codes-5

Priorclave Tactrol Controller Fault Codes and Warning Lights - from September 2018 Code Condenser Water Fault* Where the autoclave is fitted with a drain condenser this fault is triggered by detection of no water or insufficient water supply to the condenser. This acts to protect heat sensitive external drain pipework from exposure to high temperature autoclave exhaust. Activation of this fault would indicate a problem with water supply, water pressure or water inlet valve operation. Note: This fault will abort the cycle and NOT complete the sterilization process. Optional Feature – when the...