Technical Specifications for SimCube SC3-4
Physical Dimensions: The device measures 3”x 3”x 3.5” (7.6cm x 7.6cm x 8.9cm) and weighs 2.5 lbs.
Power: It can be powered by an external A/C adaptor (6VDC / 1.8amps) or 4 AA batteries with a Battery Boost Option.
Connections: Features a quick disconnect male NIBP connection and 10 ECG snaps for ECG/Resp connection (SC-4).
Manometer: The range is 0 to 400 mmHg with a resolution of 0.1 mmHg and accuracy of ± 1.0 mmHg.
User Interface: Operated with a single button, supporting various modes including Adult NIBP, Neo NIBP, Hypertensive NIBP, Manometer, Peak Detect, Alarm Test/Heart Rate Sequence, ECG Asystole, and ECG Pacer ON.
NIBP Simulations: - Adult: 120/80 (97) mmHg, 70 bpm, 1 ml pulse volume.
- Neonatal: 70/40 (51) mmHg, 95 bpm, 0.5 ml pulse volume.
- Hypertensive: 190/120 (142) mmHg, 70 bpm, 1 ml pulse volume.
ECG Simulation (SC-4): Isolated and synchronized with NIBP, featuring a 1mV R wave size, 35 ms R wave width, and QRS wave shape. Connection via 10 snaps with simulation rates of 70 bpm (95 bpm for neonatal), including Asystole, Pacer, and Alarm Test/HR Sequence.
Pacer Simulation (SC-4): Isolated and synchronized with NIBP, with a pacer size of 3 mV and width of 1.2 ms.
Respiration Simulation (SC-4): Isolated with a square wave shape, 4 Ohm size, and simulation rates of 35 bpm (47 for neonatal), including Apnea and a sequence of rates.
Peak Detect/Overpressure: Resolution of 1 mmHg.
Environmental Specifications: Operates within a voltage range of 100-240 VAC, 50-60 Hz, at temperatures from 10°C to 35°C (50°F to 95°F), with relative humidity of 10 to 80% non-condensing, and altitude up to 3,000m (9,843ft). Storage conditions range from -20°C to +60°C (-4°F to +140°F).