Technical Specifications for SimCube SC-5
Physical Dimensions: The device measures 3”x 3” x 3.5” (7.6cm X 7.6cm X 8.9cm) and weighs 2.5 lbs. It can be powered by an external A/C adaptor (6VDC / 1.8amps) or 4 AA batteries with a Battery Boost Option.
Connections: It includes a quick disconnect male NIBP connection, 10 ECG snaps for ECG/Resp connection, and a Mini-DIN for IBP connection.
Manometer: The range is from -400 to +400 mmHg with a resolution of ±0.1 mmHg for positive pressures and ±1 mmHg for negative pressures. Accuracy is ±1.0 mmHg.
User Interface: Operable via a single button or Mobilize App Control. Modes include Adult, Neo, Hypertensive, Hypotensive NIBP, Manometer, Over Pressure, HR Sequence/Alarm Test, ECG Pacer ON, Arrhythmia Sequence, and Invasive BP modes.
Simulation Modes: - NIBP Adult Simulation: 120/80 (97) mmHg, 70 bpm, 1 ml pulse volume.
- NIBP Neonatal Simulation: 70/40 (51) mmHg, 95 bpm, 0.5 ml pulse volume.
- NIBP Hypertensive Simulation: 190/120 (142) mmHg, 70 bpm, 1 ml pulse volume.
- NIBP Hypotensive Simulation: 80/40 (58) mmHg, 70 bpm, 1 ml pulse volume.
ECG Simulation: Isolated and synchronized with NIBP, featuring a 1mV R Wave size and 35 ms width. Simulation rates include 70 bpm, asystole, arrhythmia, pacer, and HR sequence.
Pacer Simulation: Isolated and synchronized with NIBP, with a pacer size of 3 mV and width of 1.2 ms.
Respiration Simulation: Isolated and synchronized with NIBP, featuring a square wave shape and 4 Ohm size. Rates include 20 bpm, apnea, and a sequence of various rates.
Arrhythmia Simulation: Includes a cardiac failure sequence with normal beats, PVCs, VTAC, VFIB, and asystole.
Invasive Blood Pressure Simulation: Isolated and synchronized with NIBP, with a pressure range of 0-250 mmHg and static pressure accuracy of ±1 mmHg.
Environmental Specifications: Operates at 100-240 VAC, 50-60 Hz, within a temperature range of 10°C to 35°C (50°F to 95°F), and relative humidity of 10 to 80% non-condensing. Storage temperature ranges from -20°C to +60°C (-4°F to +140°F).