IMAGING EXCELLENCE A/B/S/UBM Ultrasound Platform
Open the catalog to page 1INNOVATIVE ANNULAR IMAGING Quantel Medical has made a decisive leap forward with a new 5 ring annular technology on a 20 MHz probe. The principle is to emit alternating ultrasounds by 5 concentric transducers located in a single probe. The entire eye is now visible with an exceptional level of detail. MACULAR CYSTOID EDEMA IN DIABETES MELLITUS ©Photos courtesy Prof. Dr. med. Mario de La Torre MACULAR PSEUDOTUMOR, BLEEDING CHOROIDAL NEOVASCULAR MEMBRANE, KUHNT JUNIUS DISEASE This technology: • increases the depth of field, • improves the lateral resolution from 250 to 200 µm i.e. 25% • maintains...
Open the catalog to page 2• improved signal processing for enhanced resolution and penetration, • linear transducer motion to optimize image quality, • electromagnetic technology to increase speed acquisition and comfort of use, • ClearscanTM compatible for fast and comfortable examination. IOL IN SITU PLUS ANTERIOR SYNECHIA Semi-automatic quantification tools are available on ABSolu® (AOD, TIA, IT, ARA, LV) to facilitate examination and to understand the mechanisms of the iris, the lens and the ciliary bodies in glaucoma patients. PHACOMORPHIC GLAUCOMA PUPILLARY BLOCK IRIS TUMOR GLAUCOMA MODULE © Photos courtesy Prof....
Open the catalog to page 3The A-scan biometry and B mode biometry modules facilitate measurement of the axial length on all types of eyes: • moderate to dense cataract, • long eyes or posterior staphyloma. The ProBeam™ probe* (biometric probe with projected laser beam) facilitates measurement and enables better patient cooperation during the examination. The ABSolu's user interface is intuitive and easy to use. It shortens the learning curve and makes • A wide range of measuring tools. • A dual mode display for comparison of exams. • Fully configurable patient report generator. <j The B15, B20-5A and UBM probes are equipped...
Open the catalog to page 4TECHNICAL SPECIFICATIONS Glaucoma quantifying semi-automated tools: Cineloop in B mode: Adjustable gain: Adjustable Time Gain Control (TGC): 11 MHz Probe Transducer frequency: 11 MHz Tip diameter: 7 mm (0.28’’) Electronic resolution: 0.04 mm (0.0016’’) Depth of exploration: 40/80 mm (1.57’’/3.15’’) on 2048 points Aiming beam: LED or laser beam ProBeamTM Contact and immersion techniques compatible Axial length measurements Ultrasound propagation velocity adjustable per segment (anterior chamber, lens, vitreous) and IOL and vitreous material Built-in pattern recognition: Phakic, Dense/Long, Aphakic,...
Open the catalog to page 55 Pages
6 Pages