Electronic X-Ray Ruler
6Pages

{{requestButtons}}

Catalog excerpts

Electronic X-Ray Ruler - 1

QUART nonius Electronic X-Ray Ruler

Open the catalog to page 1
Electronic X-Ray Ruler - 2

X-Ray Field & Fan X-Ray Beam Measurement The QUART nonius is an easy-to-use but very sophisticated measuring instrument to verify size and quality of x-ray fields. It can also be used to analyse the properties of fanned x-ray beams. The digitisation in x-ray technology makes traditional film/screens less available. Originally, they were used for checks on x-ray beam properties. Today, the QUART nonius performs the same task. And, it provides even more substantial features. The QUART nonius can be used to verify if the light visor matches the actual x-ray field. In addition, the nonius provides...

Open the catalog to page 2
Electronic X-Ray Ruler - 3

Technical Data Applications – Dental Accuracy / Resolution +/- 0.1 mm Exposure Threshold Dose ≥ 200 µGy / Dose Rate ≥ 20 µGy/s Minimum Exposure (ref. Application Sections) Sensor Area 40 mm Length (16 Active Sensor Elements) Temperature Range 15 – 40 °C Plug & Play Component Connectivity Standard USB (2.0) Operating System Windows 7, Vista, XP Weight 190 g (without USB Cable) To accurately assess field properties of intraoral equipment, a special holder is provided. 4 exposures are required for a substantial evaluation. Size of Head Unit 55 x 75 x 15 mm (W x H x D) Minimum Exposure...

Open the catalog to page 3
Electronic X-Ray Ruler - 4

Applications – Radiology Computed Tomography To evaluate width and dose profiles of fanned x-ray beams in CT, the device is positioned on the patient table towards the rotational direction of the scanner. Only one pass is sufficient to adequately assess the parameters. Minimum Exposure CT: 100 kV / 50 mA / SID 100 cm At mammography installations the nonius is mainly used to check the field alignment towards the thorax wall side. The measuring precision exceeds the one acquired with standard phantoms or other tools. Depending on positioning, the dose rate profile within the radiation field (heel...

Open the catalog to page 4
Electronic X-Ray Ruler - 5

y ld Equivalenc Radiation Fie iation field. Light Field & the actual rad Evaluation of is 2.7mm off e, the light field For this field sid QUART nonius The nonius can be applied in all fields of radiography: Dental, CR/DR, Mammography, CT, ... Quick Operation Align the light field with the nonius center line, or position the nonius directly on the edge of the image receiver. Expose. Read the results from the display of the connected computer. Save or print out protocol. The use of the QUART nonius is incredibly easy. Only 3 steps are required:

Open the catalog to page 5
Electronic X-Ray Ruler - 6

% We help to help others QUART is a proud Supporter of Medecins sans Frontieres

Open the catalog to page 6

All QUART X-Ray QA Solutions catalogs and technical brochures

  1. QUART X-Ray

    8 Pages

  2. Densitometry

    4 Pages