Human High Sensitivity Cytokine Base Kit A

Product catalog summary
Introduction
Cytokines are essential proteins for intercellular signaling, playing roles in growth regulation, cellular proliferation, and immune response modulation. The Human High Sensitivity Cytokine Base Kit A is designed for quantifying multiple human cytokines in serum and plasma using Luminex® technology.

Principle of the Assay
The assay uses analyte-specific antibodies on magnetic microparticles with fluorophores, analyzed by a Luminex system. Analytes bind to antibodies, followed by washes and a biotinylated antibody cocktail. Streptavidin-phycoerythrin is added, and dual laser analysis measures bound analytes.

Limitations of the Procedure
This research-only kit is unsuitable for diagnostics. It must be used before expiration, and reagents from different lots should not be mixed. Sample handling and assay conditions can affect results, and untested factors may interfere.

Materials Provided & Storage Conditions
The kit includes cytokine standards, diluents, and wash buffer, stored at 2-8°C. Reconstituted materials should be freshly prepared for each assay.

Other Supplies Required
Necessary equipment includes a Luminex analyzer, microplate magnet, pipettes, and a microplate shaker. Safety equipment is recommended.

Precautions
Some components contain preservatives that may cause allergies. Protective gear should be worn, and hands washed after handling.

Technical Hints
Avoid foaming when mixing, change pipette tips between additions, and protect microparticles from light for accurate results.

Sample Collection & Storage
Serum and plasma samples should be collected and stored properly to maintain stability, avoiding repeated freeze-thaw cycles.

Sample Preparation
Samples need centrifugation and dilution, with a 2-fold dilution recommended using Calibrator Diluent RD6-40.

Reagent Preparation
Reagents should reach room temperature before use. Wash buffer and standards must be prepared as instructed for proper assay performance.

Diluted Microparticle Cocktail Preparation
Microparticle concentrates should be centrifuged, vortexed, and diluted, used within 30 minutes, and protected from light.

Biotin-Antibody Concentrates Preparation
Biotin-Antibody Concentrates are prepared by centrifuging, vortexing, and diluting in Biotin Antibody Diluent 2, with specific volumes for different well counts.

Instrument Settings
Instructions for MAGPIX® and Luminex® analyzers include sample volume, microparticle region assignment, and parameters like flow rate and reporter gain.

Assay Procedure
Reagents and samples are prepared, added to wells, and incubated with a microparticle cocktail. The process includes washing steps, adding Biotin-Antibody Cocktail and Streptavidin-PE, followed by incubation and analysis. Light protection for certain components is crucial.

Calculation of Results
Results are calculated using standard concentrations and a standard curve with a five-parameter logistic curve-fit. Concentrations must be adjusted by the dilution factor.

Calibration
The assay is calibrated against recombinant human cytokines from R&D Systems®.

Specificity
The assay shows less than 0.5% interference with various recombinant proteins.

Plate Layout
A plate layout is provided for recording standards and samples.

Notes
The assay is for research use only, not for diagnostic procedures, with trademark information included.
See more

Catalog excerpts

Human High Sensitivity Cytokine Base Kit A-1

Luminex® Performance Assay Human High Sensitivity Cytokine Base Kit A Catalog Number LHSCM000 For the simultaneous quantitative determination of multiple human cytokine concentrations in serum and plasma. This package insert must be read in its entirety before using this product. For research use only. Not for use in diagnostic procedures.

 Open the catalog to page 1
Human High Sensitivity Cytokine Base Kit A-2

TABLE OF CONTENTS SECTION Manufactured and Distributed by: USA R&D Systems, Inc. 614 McKinley Place NE, Minneapolis, MN 55413 TEL: 800 343 7475 612 379 2956 FAX: 612 656 4400 E-MAIL: [email protected] Distributed by: Europe | Middle East | Africa Bio-Techne Ltd. 19 Barton Lane, Abingdon Science Park Abingdon OX14 3NB, UK TEL: +44 (0)1235 529449 FAX: +44 (0)1235 533420 E-MAIL: [email protected] China Bio-Techne China Co., Ltd. Unit 1901, Tower 3, Raffles City Changning Office, 1193 Changning Road, Shanghai PRC 200051 TEL: +86 (21) 52380373 (400) 821-3475 FAX: +86 (21) 52371001 E-MAIL:...

 Open the catalog to page 2
Human High Sensitivity Cytokine Base Kit A-3

INTRODUCTION Cytokines are intercellular signaling proteins released from a wide variety of cells and tissues. They play an integral role in regulating growth and cellular proliferation as well as modulating host response to infection, injury, and inflammation. Cytokines also influence reproduction and bone remodeling. A large number of cytokines are pleiotropic and share similar functions. In addition, many cytokines influence the production of other cytokines. Analysis and quantification of cytokines in biological fluids has thus become increasingly important. Methods such as bioassay, enzyme-linked...

 Open the catalog to page 3
Human High Sensitivity Cytokine Base Kit A-4

PRINCIPLE OF THE ASSAY Luminex® Performance Assay multiplex kits are designed for use with any Luminex analyzer including the MAGPIX®, Luminex® 100/200™, FLEXMAP 3D®, xMAP INTELLIFLEX®, or Bio-Rad® Bio-Plex®, dual laser, flow-based sorting and detection platforms. Analyte-specific antibodies are pre-coated onto magnetic microparticles embedded with fluorophores at set ratios for each unique microparticle region. Microparticles, standards and samples are pipetted into wells and the immobilized antibodies bind the analytes of interest. After washing away any unbound substances, a biotinylated antibody...

 Open the catalog to page 4
Human High Sensitivity Cytokine Base Kit A-5

Store the unopened kit at 2-8 °C. Do not use past kit expiration date. PART PART# Standard Cocktail 1 Microparticle Diluent 895529 Biotin Antibody Diluent 2 895832 Wash Buffer Concentrate 895003 Standard Value Card 749255 STORAGE OF OPENED, DILUTED, DESCRIPTION OR RECONSTITUTED MATERIAL Use a fresh standard and control for each assay. Discard after use. May be stored for up to 1 month at 2-8 °C.* Prepare fresh 1Xsolutions at the time of assay. Discard after use. 2 vials of recombinant human cytokines in a buffered protein base with preservatives; lyophilized. 6 mL of a buffered protein base with...

 Open the catalog to page 5
Human High Sensitivity Cytokine Base Kit A-6

OTHER SUPPLIES REQUIRED • Luminex® Performance Assay analyte-specific kit(s) (see page 1) • MAGPIX®, Luminex® 100/200™, FLEXMAP 3D®, xMAP INTELLIFLEX®, or Bio-Rad® Bio-Plex® analyzer with X-Y platform • Hand-held microplate magnet or platewasher with a magnetic platform • Pipettes and pipette tips • Deionized or distilled water • Multi-channel pipette, manifold dispenser, or automated dispensing unit • 50 mL and 500 mL graduated cylinders • Horizontal orbital microplate shaker (0.12" orbit) capable of maintaining a speed of 800 ± 50 rpm • Microcentrifuge • Polypropylene test tubes for dilution...

 Open the catalog to page 6
Human High Sensitivity Cytokine Base Kit A-7

SAMPLE COLLECTION & STORAGE The sample collection and storage conditions listed below are intended as general guidelines. Sample stability has not been evaluated. Serum - Use a serum separator tube (SST) and allow samples to clot for 30 minutes at room temperature before centrifuging for 15 minutes at 1000 x g. Remove serum and assay immediately or aliquot and store samples at ≤ -20 °C. Avoid repeated freeze-thaw cycles. Plasma - Collect plasma using EDTA or heparin as an anticoagulant. Centrifuge for 15 minutes at 1000 x g within 30 minutes of collection. Assay immediately or aliquot and store...

 Open the catalog to page 7
Human High Sensitivity Cytokine Base Kit A-8

REAGENT PREPARATION Bring all reagents to room temperature before use. Wash Buffer - If crystals have formed in the concentrate, warm to room temperature and mix gently until the crystals have completely dissolved. Add 20 mL of Wash Buffer Concentrate to 480 mL of deionized or distilled water to prepare 500 mL of Wash Buffer. Standard - Refer to the Standard Value Card for the reconstitution volume and assigned values. Reconstitute the Standard Cocktail with Calibrator Diluent RD6-40. Allow the standard to sit for a minimum of 15 minutes with gentle agitation prior to making dilutions. Note:...

 Open the catalog to page 8
Human High Sensitivity Cytokine Base Kit A-9

1. Centrifuge each Microparticle Concentrate vial for 30 seconds at 1000 x g prior to removing the cap. 2. Gently vortex the vials to resuspend the microparticles, taking precautions not to invert the vials. 3. Dilute the Microparticle Concentrates in the mixing bottle provided. The volume of the Microparticle Concentrate listed in the table below is for each analyte (e.g. if measuring a full plate of IL-ip and IL-6, add 50 pL of IL-ip Microparticle Concentrate and 50 pL of IL-6 Microparticle Concentrate to 2.5 mL of Microparticle Diluent).

 Open the catalog to page 9
Human High Sensitivity Cytokine Base Kit A-10

INSTRUMENT SETTINGS Note: Adjust the probe height setting on the analyzer to avoid puncturing the plate. Calibrate the analyzer using the proper reagents for superparamagnetic microparticles (refer to instrument manual). MAGPIX® analyzer: a) Sample volume: 50 μL b) Assign the microparticle region for each analyte being measured (see page 1) c) 50 count/region d) Collect Median Fluorescence Intensity (MFI) Luminex® 100/200™, Luminex® FLEXMAP 3D®, Luminex® INTELLIFLEX, and Bio-Rad Bio-Plex analyzers: Note: Ensure that the instrument flow rate is set to the default of 60 μL/minute (fast) for all...

 Open the catalog to page 10
Human High Sensitivity Cytokine Base Kit A-11

ASSAY PROCEDURE Bring all reagents and samples to room temperature before use. It is recommended that all standards, controls, and samples be assayed in duplicate. Note: Protect microparticles and Streptavidin-PE from light at all times. 1. Prepare all reagents, working standards, and samples as directed in the previous sections. 2. Add 100 μL of standard, control, or sample* per well. A plate layout is provided to record standards and samples assayed. 3. Resuspend the diluted Microparticle Cocktail by inversion or vortexing. Add 25 μL of the microparticle cocktail to each well of the microplate....

 Open the catalog to page 11