Luminex® Performance Assay - Human Cytokine Base Kit A

Product catalog summary
Introduction
This document describes a cytokine assay kit designed for the simultaneous quantification of multiple cytokines in biological samples. It addresses the limitations of traditional methods like ELISA and PCR, which cannot measure multiple cytokines at once.
Principle of the Assay
The assay utilizes Luminex® technology with magnetic microparticles coated with analyte-specific antibodies. These are analyzed using lasers to identify microparticle regions and quantify analytes. The process involves binding analytes, adding a biotinylated antibody cocktail, and detecting with streptavidin-phycoerythrin.
Limitations of the Procedure
The kit is intended for research use only and not for diagnostic purposes. It should not be used after expiration, and reagents from different lots should not be mixed. Variations in sample handling can affect results, and while designed to minimize interference, this cannot be guaranteed without testing.
Materials Provided & Storage Conditions
The kit includes reagents like standard cocktails and wash buffer, which should be stored at 2-8°C and used before expiration. Fresh solutions are required for each assay.
Other Supplies Required
Necessary equipment includes a Luminex analyzer, microplate magnet, pipettes, and a microplate shaker. Optional controls are available.
Precautions
Some components contain preservatives that may cause allergic reactions. Protective gear is recommended, and hands should be washed after handling.
Technical Hints
Avoid foaming when mixing, change pipette tips to prevent contamination, and protect reagents from light.
Sample Collection & Storage
Guidelines are provided for collecting and storing samples like cell culture supernates, serum, and plasma, which should be stored at ≤ -20°C to avoid freeze-thaw cycles.
Sample Preparation
Serum and plasma samples require centrifugation and dilution, with specific instructions for different analytes.
Reagent Preparation
Reagents should reach room temperature before use. Instructions are given for preparing wash buffer and calibrator diluents, with standards reconstituted as per value cards.
Standard Preparation
A 3-fold dilution series is prepared using reconstituted Standard Cocktail vials, with working standard 1 as the high standard and a calibrator diluent as the blank.
Microparticle Cocktail Preparation
Microparticle Concentrates are centrifuged, vortexed, and diluted, with specific volumes for different well numbers. Protection from light is crucial.
Biotin-Antibody Cocktail Preparation
Similar preparation steps as microparticles, with specified volumes based on well numbers.
Streptavidin-PE Preparation
Involves centrifuging, vortexing, and diluting in Wash Buffer, with specific volumes for different well numbers, emphasizing light protection.
Instrument Settings
Detailed settings for analyzers are provided, including sample volume, bead type, and calibration requirements.
Assay Procedure
The procedure includes reagent preparation, sample addition, incubation, washing, and reading results with analyzers, stressing light protection.
Calculation of Results
Results are calculated using standard concentrations and a 5-parameter logistic curve-fit, considering dilution factors.
Specificity
The assay shows less than 0.5% cross-reactivity with various recombinant proteins.
Plate Layout
A layout is provided for recording standards and samples.
See more

Catalog excerpts

Luminex® Performance Assay - Human Cytokine Base Kit A-1

Luminex® Performance Assay Human Cytokine Base Kit A Catalog Number LUHM000 For the simultaneous quantitative determination of multiple human cytokine concentrations in cell culture supernates, serum, and plasma. This package insert must be read in its entirety before using this product. For research use only. Not for use in diagnostic procedures.

 Open the catalog to page 1
Luminex® Performance Assay - Human Cytokine Base Kit A-2

TABLE OF CONTENTS SECTION Manufactured and Distributed by: USA R&D Systems, Inc. 614 McKinley Place NE, Minneapolis, MN 55413 TEL: 800 343 7475 612 379 2956 FAX: 612 656 4400 E-MAIL: [email protected] Distributed by: Europe | Middle East | Africa Bio-Techne Ltd. 19 Barton Lane, Abingdon Science Park Abingdon OX14 3NB, UK TEL: +44 (0)1235 529449 FAX: +44 (0)1235 533420 E-MAIL: [email protected] China Bio-Techne China Co., Ltd. Unit 1901, Tower 3, Raffles City Changning Office, 1193 Changning Road, Shanghai PRC 200051 TEL: +86 (21) 52380373 (400) 821-3475 FAX: +86 (21) 52371001 E-MAIL:...

 Open the catalog to page 2
Luminex® Performance Assay - Human Cytokine Base Kit A-3

INTRODUCTION Cytokines are intercellular signaling proteins released from a wide variety of cells and tissues. They play an integral role in regulating growth and cellular proliferation as well as modulating host response to infection, injury, and inflammation. Cytokines also influence reproduction and bone remodeling. A large number of cytokines are pleiotropic and share similar functions. In addition, many cytokines influence the production of other cytokines. Analysis and quantification of cytokines in biological fluids and cell culture supernates has thus become increasingly important. Methods...

 Open the catalog to page 3
Luminex® Performance Assay - Human Cytokine Base Kit A-4

PRINCIPLE OF THE ASSAY Luminex® Performance Assay multiplex kits are designed for use with any Luminex analyzer including the MAGPIX®, Luminex® 100/200™, FLEXMAP 3D®, xMAP INTELLIFLEX®, or Bio-Rad® Bio-Plex®, dual laser, flow-based sorting and detection platforms. Analyte-specific antibodies are pre-coated onto magnetic microparticles embedded with fluorophores at set ratios for each unique microparticle region. Microparticles, standards and samples are pipetted into wells and the immobilized antibodies bind the analytes of interest. After washing away any unbound substances, a biotinylated antibody...

 Open the catalog to page 4
Luminex® Performance Assay - Human Cytokine Base Kit A-5

Store the unopened kit at 2-8 °C. Do not use past kit expiration date. ^Provided this is within the expiration date of the kit. Note: Additional wash buffer is available for purchase (R&D Systems®, # WA126).

 Open the catalog to page 5
Luminex® Performance Assay - Human Cytokine Base Kit A-6

OTHER SUPPLIES REQUIRED • Luminex® Performance Assay analyte-specific kit(s) (see page 1) • MAGPIX®, Luminex® 100/200™, FLEXMAP 3D®, xMAP INTELLIFLEX®, or Bio-Rad® Bio-Plex® analyzer with X-Y platform • Hand-held microplate magnet or platewasher with a magnetic platform • Pipettes and pipette tips • Deionized or distilled water • Multi-channel pipette, manifold dispenser, or automated dispensing unit • 50 mL and 500 mL graduated cylinders • Horizontal orbital microplate shaker (0.12" orbit) capable of maintaining a speed of 800 ± 50 rpm • Microcentrifuge • Polypropylene test tubes for dilution...

 Open the catalog to page 6
Luminex® Performance Assay - Human Cytokine Base Kit A-7

SAMPLE COLLECTION & STORAGE The sample collection and storage conditions listed below are intended as general guidelines. Sample stability has not been evaluated. Cell Culture Supernates - Remove particulates by centrifugation and assay immediately or aliquot and store samples at ≤ -20 °C. Avoid repeated freeze-thaw cycles. Serum - Use a serum separator tube (SST) and allow samples to clot for 30 minutes at room temperature before centrifuging for 15 minutes at 1000 x g. Remove serum and assay immediately or aliquot and store samples at ≤ -20 °C. Avoid repeated freeze-thaw cycles. Plasma - Collect...

 Open the catalog to page 7
Luminex® Performance Assay - Human Cytokine Base Kit A-8

REAGENT PREPARATION Bring all reagents to room temperature before use. Wash Buffer - If crystals have formed in the concentrate, warm to room temperature and mix gently until the crystals have completely dissolved. Add 20 mL of Wash Buffer Concentrate to 480 mL of deionized or distilled water to prepare 500 mL of Wash Buffer. Calibrator Diluent RD5K (diluted 1:2) - For cell culture supernate samples only. Add 20 mL of Calibrator Diluent RD5K to 20 mL of deionized or distilled water to prepare 40 mL of Calibrator Diluent RD5K (diluted 1:2). Standards - Refer to the Standard Value Cards for the...

 Open the catalog to page 8
Luminex® Performance Assay - Human Cytokine Base Kit A-9

1. Centrifuge each Microparticle Concentrate vial for 30 seconds at 1000 x g prior to removing the cap. 2. Gently vortex the vials to resuspend the microparticles, taking precautions not to invert the vials. 3. Dilute the Microparticle Concentrates in the mixing bottle provided. The volume of the Microparticle Concentrate listed in the table below is for each analyte (e.g. if measuring a full plate of IL-ip and IL-6, add 50 pL of IL-1p Microparticle Concentrate and 50 pL of IL-6 Microparticle Concentrate to 5 mL of Microparticle Diluent).

 Open the catalog to page 9
Luminex® Performance Assay - Human Cytokine Base Kit A-10

INSTRUMENT SETTINGS Note: Adjust the probe height setting on the analyzer to avoid puncturing the plate. Calibrate the analyzer using the proper reagents for superparamagnetic microparticles (refer to instrument manual). MAGPIX® analyzer: a) Sample volume: 50 μL b) Assign the microparticle region for each analyte being measured (see page 1) c) 50 count/region d) Collect Median Fluorescence Intensity (MFI) Luminex® 100/200TM, FLEXMAP 3D®, xMAP INTELLIFLEX®, and Bio-Rad® Bio-Plex® analyzers: Note: Ensure that the instrument flow rate is set to the default of 60 μL/minute (fast) for all flow based...

 Open the catalog to page 10
Luminex® Performance Assay - Human Cytokine Base Kit A-11

ASSAY PROCEDURE Bring all reagents and samples to room temperature before use. It is recommended that all standards, controls, and samples be assayed in duplicate. Note: Protect microparticles and Streptavidin-PE from light at all times. 1. Prepare all reagents, working standards, and samples as directed in the previous sections. 2. Add 50 μL of standard, control, or sample* per well. A plate layout is provided to record standards and samples assayed. 3. Resuspend the diluted Microparticle Cocktail by inversion or vortexing. Add 50 μL of the microparticle cocktail to each well of the microplate....

 Open the catalog to page 11