Product catalog summary
Overview: The 10X6-60DAP is designed for Dose Area Product (DAP) measurement, particularly suitable for small beams such as Pan-Dental or CBCT-Dental. It integrates seamlessly with existing Accu-Gold systems without requiring calibration adjustments.
Key Features:
  • Includes an easy-to-use chamber mounting and alignment fixture to minimize setup time.
  • Offers auto-ranging and unit selection between Gy-m2 or Gy-cm2.
  • Provides a flat energy response across all RQR and RQA beams.
  • Can be used with an Accu-Gold multi-sensor for simultaneous measurement of DAP, kVp, and HVL.
Specifications:
  • DAP Mode Rate Specifications: 11 pGy-m2/s to 110 μGy-m2/s.
  • Exposure Specifications: 56 pGy-m2 to 290 mGy-m2.
  • Exposure Rate Dependence: Less than 5% over the rated rate specification.
  • Calibration Accuracy: ±5% using X-rays at IEC RQR-8, 100 kV & 4.0 mm Al HVL.
  • Energy Dependence: ±5% from RQR-3 to RQR-10 (50 kV-1.8 mm Al HVL to 150 kV-6.6 mm Al HVL).
  • Chamber Beam Attenuation: 6% ±2% over energy range RQR3 to RQR10.
Construction: The chamber is constructed with a parallel plate, polycarbonate wall, and a conductive graphite exterior coating. It has a 60 cm3 active volume and includes a 2-meter low noise triax cable, weighing 0.13 kg.
Energy Response: Correction factors are provided versus HVL (mmAl) for typical use.
See more

Catalog excerpts

10X6-60DAP-1

10X6-60DAP The 'DAP' Chamber Ideal for Dose Area Product (DAP) measurement of small beams such as Pan-Dental or CBCT-Dental. • Easy to use chamber mounting and alignment fixture that minimizes set-up time (Mounting holder included) • Auto ranging and convenient unit selection for Gy-m2 or Gy-cm2 • Flat energy response over all RQR and RQA beams • Add to your existing Accu-Gold system – no calibration adjustments, just update your software and plug it in • Use with an Accu-Gold multi-sensor to simultaneously measure DAP, kVp, and hvl Rate Specifications Exposure Specifications Exposure Rate Dependance Calibration Accuracy: Energy Dependence Chamber Beam Attenuation Construction 11 pGy-m2/s - 110 μGy-m2/s 56 pGy-m2 - 290 mGy-m2 <5% over rated rate specification ±5% using X-rays @ IEC RQR-8 100 kV & 4.0 mm Al HVL ±5%, RQR-3 to RQR-10 (50 kV-1.8 mm Al HVL to 150 kV-6.6 mm Al HVL) 6% ±2% over energy range RQR3 to RQR10 Parallel plate. Polycarbonate wall. Conductive graphite exterior coating; 60 cm3 active volume, 2 meter low noise triax cable; 0.13 kg Mounting Holder Energy Response Correction Factor vs HVL (mmAl) (Typical) apply when used with Accu-Gold series digitizer. THE GOLD STANDARD IN RADIATION MEASUREMENT 426 West Duarte Road Monrovia, CA 91016 USA T: (626) 357-7921 F: (626) 357-8863 www.radcal.