
Group: Randox

Catalog excerpts

QC Explained 1. 0 - Quality Control for Point of Care Testing QUALITY CONTROL QC Explained 1.0 - Quality Control for Point of Care Testing QC Explained 1.0 - Quality Control for Kee, Sarah., Adams, Lynsey., Whyte, Carla J., McVicker, Louise. Point of Care Testing QC Explained 1. 0 - Quality Control for Point of Care Testing Kee, Sarah., Adams, Lynsey., Whyte, Carla J., McVicker, Louise. Background Point of care testing (POCT) refers to testing that is performed near or at the site of a patient with the result leading to a possible change in the care of the patient. Over the past few years, the Point of care testing (POCT) refers to popularity and demand of POCT has been testing that is performed near or at the growing rapidly. This should come as no site of a patient with the result leading surprise as there are many advantages to to a possible change in the care of the POCT, for example, the added patient. Over the past few years, the convenience of being able to obtain a popularity and demand of POCT has been rapid result at the patient’s bedside, thus growing rapidly. This should come as no allowing immediate action, saving time surprise as there are many advantages to and improving the potential outcome for POCT, for example, the added the patient. convenience of being able to obtain a Although there the patient’s bedside,using rapid result at are many benefits of thus POCT devices in action, saving their terms of time allowing immediate convenience, these benefits are only true if and improving the potential outcome for the patient. produced are both accurate results and reliable. Ensuring accuracy and Although there are many benefits of using POCT devices in terms of their convenience, these benefits are only true if 1 the results produced are both accurate and reliable. Ensuring accuracy and reliability is the primary responsibility of Quality Control. Quality control is composed of two key elements; internal quality control (IQC) and external quality assessment (EQA). IQC reliability is the primary responsibility of involvesControl. quality control material Quality running that contains analytes of known Quality control is composed of two key concentration to monitor the precision of elements; internal quality control (IQC) the analytical process over time. Whereas and external quality assessment (EQA). IQC EQA involves running blind patient-like involves running quality control material samples, comparing your results to peer that contains analytes of known results, in order to retrospectively monitor concentration to monitor the precision of the accuracy of reporting. EQA samples the analytical process over time. Whereas should be treated as if they were a patient EQA involves running blind patient-like sample and therefore must be run by samples, comparing your results to peer personnel who would be using the device. results, in order to retrospectively monitor This provides confidence in the reliability the accuracy of reporting. EQA samples of patient test results. should be treated as if they were a patient sample and therefore must be run by personnel who would be using the device. This provides confidence in the reliability of patient test results
Open the catalog to page 1
QC Explained 1. 0 - Quality Control for Point of Care Testing QUALITY CONTROL Section Title Page 1.0 Importance of Quality Control for POCT Devices 3 3.0 Choosing an Appropriate QC Procedure for your POCT Device 4 3.1 Balancing risk with the stringency of your POCT quality 4 control procedure 3.1.1 Choosing a QC procedure for High Throughput Analyser 5 Type POCT Instruments 3.1.2 Choosing a QC procedure for Cartridge Based Instruments 5 3.1.3 Choosing a QC procedure for Strip Based Instruments 6 3.1.4 Choosing a QC procedure for manually read...
Open the catalog to page 2
QC Explained 1. 0 - Quality Control for Point of Care Testing QUALITY CONTROL ISO 15189 states that "Quality Control materials shall be periodically examined with a frequency that is based on the stability of the procedure and the risk of harm to the patient from an erroneous result" (1). Therefore, when implementing a QC strategy for POCT devices the risk of harm to the patient should be the foundation of the plan: where and why do errors occur and what are the consequences of an erroneous result to the patient? It is important to balance the risk of harm to the patient with the...
Open the catalog to page 3
QC Explained 1. 0 - Quality Control for Point of Care Testing QUALITY CONTROL on patient outcome was 14.7% (3), demonstrating that for POCT devices there is larger room for error and a greater need for quality control. Some POCT device manufacturers advise that QC procedures similar to those found in the laboratory are unnecessary for POCT devices. However, this is not the opinion of the majority of laboratory professionals who believe a place undoubtedly remains for traditional IQC and EQA practices for POCT. This view is supported by the introduction of the new ISO POC...
Open the catalog to page 4
QC Explained 1. 0 - Quality Control for Point of Care Testing QUALITY CONTROL • tests used to support the clinician's decision in isolation, • tests acted upon immediately, and • tests performed on specimens that are difficult to collect. The POCT strategy will depend on the POC device that is being used as POCT devices can be broadly split into three categories: 1. "High Throughput Analyser"- Full size instruments used at the point of care, e.g.: blood gas analysers. 2. "Cartridge Based Instruments"- For example HbA1c and INR analysers. 3. "Strip Based Instruments" e.g.: electrochemical...
Open the catalog to page 5
QC Explained 1. 0 - Quality Control for Point of Care Testing QUALITY CONTROL Each time the laboratory receives a new batch of cartridges they should be evaluated using multi-level, third party controls to ensure that the cartridges have not been damaged during transport. It is also a good idea to periodically test quality control samples over the shelf life of a particular batch. Ideally a third party control should be used every day of patient testing to ensure the stability of the on-board reagents and the accuracy of the results obtained, as recommended by ISO 15189. However,...
Open the catalog to page 6All Randox Laboratories catalogs and technical brochures
-
LT735 Vivalytic Overview
32 Pages
-
Acusera New Controls
8 Pages
-
Acusera 24.7
24 Pages
-
Acusera Third Party Controls
108 Pages
-
Acusera SMART Controls
12 Pages
-
Company Overview
40 Pages
-
Reagents Brochure
64 Pages
-
sdLDL-C
8 Pages
-
RX SERIES ANALYSER OVERVIEW
24 Pages
-
RX modena
20 Pages
-
RX misano
20 Pages
-
RX Daytona Plus
20 Pages
-
Rx Imola
20 Pages
-
RX Monaco
20 Pages
-
2024 Product List
64 Pages
-
LT107 Evidence Investigator
32 Pages
-
VeraSTAT-V
16 Pages
-
LT033 RIQAS Explained
64 Pages
-
VERASTAT
16 Pages
-
Endocrine Array
4 Pages
-
Cerebral Arrays
4 Pages
-
Thyroid Arrays
4 Pages
-
Respiratory Multiplex Array
8 Pages
-
LT253 Molecular Diagnostics
20 Pages
-
LT367 FH Array Brochure
4 Pages
-
The role of EQA in QC
8 Pages
-
Basic QC Stastics
8 Pages
-
Commutability Guide
4 Pages
-
How to measure uncertainty
8 Pages
-
ISO 15189 Educational Guide
8 Pages
-
Troubleshooting QC Errors
8 Pages
-
Qnostics
52 Pages
-
Adiponectin LT519
28 Pages
-
Specific Proteins
40 Pages
-
Linearity sets
12 Pages
-
Antioxidants
16 Pages
-
Cardiology & Lipid Testing
28 Pages
-
Diabetes Portfolio
28 Pages
-
Total Bile Acids
4 Pages
-
VIVALYTIC
30 Pages
-
RANDOX DISCOVERY
36 Pages
-
HbA1c
2 Pages
-
Preparing QC
1 Pages
-
LT394 Using QC Multirules
1 Pages
-
Guide to running QC
1 Pages
-
Tumour Marker Arrays
4 Pages
-
Which QC is the Right QC
8 Pages
-
How often is right for QC
6 Pages
-
Cardiac Risk Multiplex Array
4 Pages
-
LT241 Metabolic Array MAY15
8 Pages
-
KRAS / BRAF / PIK3CA Array*
4 Pages
-
Custom Arrays for Biochip
12 Pages
-
LT169 Cardiac Array
4 Pages
Archived catalogs
-
ACUSERA
108 Pages
-
Evidence Evolution
28 Pages
-
Evidence
16 Pages
-
Metabolic Syndrome Arrays
8 Pages
-
STI Multiplex Array
8 Pages
-
Molecular Testing
16 Pages
-
Evidence Investigator
20 Pages
-
Fertility Array
4 Pages
-
Rx Daytona
16 Pages
-
Evidence Investigator
16 Pages
-
Evidence
16 Pages