BENCHTOP AUTOCLAVES WITH PREVACUUMS AND DRYING AHS-B SERIES
11Pages

{{requestButtons}}

Catalog excerpts

BENCHTOP AUTOCLAVES WITH PREVACUUMS AND DRYING AHS-B SERIES - 1

BENCHTOP AUTOCLAVES WITH PREVACUUMS AND DRYING AHS-B SERIES CLASSIC LINE EXCELLENT PERFORMANCE, ADVANCED FEATURES, COMPACT FOOTPRINT AND ALL-AROUND VERSATILITY FOR SEVERAL APPLICATIONS

Open the catalog to page 1
BENCHTOP AUTOCLAVES WITH PREVACUUMS AND DRYING AHS-B SERIES - 2

+ STERILIZATION OF LOADS OF COMPLEX GEOMETRIES, POROUS OR HOLLOW OBJECTS, FABRIC LOADS, LABORATORY WASTE BAGS, PLASTICS, GLASSWARE, CULTURE MEDIA AND LIQUIDS Front-loading benchtop autoclaves with prevacuums and drying The AHS-B Series benchtop autoclaves with front-loading access cover all laboratory sterilization needs in many industries and research facilities with the aim of increasing the productivity of the laboratory. A great chamber capacity, the independent integrated steam generator, the touchscreen display, the independent clean water tank and waste water tank, the initial...

Open the catalog to page 2
BENCHTOP AUTOCLAVES WITH PREVACUUMS AND DRYING AHS-B SERIES - 3

Heating by an independent integrated powerful steam generator. Final vacuum drying feature by a heating jacket and a vacuum pump to completely dry solid loads. Sterilization chamber and door made of high quality stainless steel grade AISI-316L extremely resistant to corrosion. <”> Adjustable number of initial prevacuum pulses to guarantee proper steam penetration on items of complex geometries and large loads. Equipment built following all applicable European Union quality, regulatory and safety standards. Control by a PID microprocessor and a 5'' touchscreen, with 50 available programs,...

Open the catalog to page 3
BENCHTOP AUTOCLAVES WITH PREVACUUMS AND DRYING AHS-B SERIES - 4

WORKING PRINCIPLE AHS-B Series autoclaves provide a solution for the multiple sterilization needs of all laboratories, including wrapped and unwrapped solids, fabric loads, porous and hollow objects, plastics, metal utensils, laboratory waste bags, liquids, culture media, glassware and other laboratory items. The load has to be placed into the vessel's trays or basket and, after manually filling the independent clean water tank with purified water, the equipment starts to create the initial prevacuum, automatically feeds water to the independent integrated steam generator, generating...

Open the catalog to page 4
BENCHTOP AUTOCLAVES WITH PREVACUUMS AND DRYING AHS-B SERIES - 5

AHS-B SERIES PROGRAMS AHS-B Series autoclaves have 50 programs, from P1 to P50, and the first fourteen are predefined and protected. PREDEFINED PROGRAMS Program Program name Prevacuum Sterilization Sterilization time Drying time Program mode Core probe N° pulses temperature min min Solids or Liquids regulation P14 Liquid probe 1 121 15' - Liquids Core probe The rest of the programs are editable with the following parameters settings: • Prevacuum pulses number. • Sterilization temperature. • Sterilization time. • Final drying time. • Sterilization mode (Solids or Liquids). • Sterilization...

Open the catalog to page 5
BENCHTOP AUTOCLAVES WITH PREVACUUMS AND DRYING AHS-B SERIES - 6

Front-loading benchtop autoclaves with prevacuums and drying DIFFERENT SIZES BUT SAME FEATURES COMPONENTS SUPPLIED WITH THE EQUIPMENT COMPONENTS SUPPLIED WITH THE EQUIPMENT A. Stainless steel tray support for 5 trays. B. 3 stainless steel wire trays. C. Auxiliary plastic tray for collecting condensed water after opening the door in cycles without final drying. D. Holding clamp to move the trays. E. Silicone tube of 1m with fast connection to drain the independent clean water tank and the independent waste water tank. F. Sainless steel bag support. A. Stainless steel tray support for 5...

Open the catalog to page 6
BENCHTOP AUTOCLAVES WITH PREVACUUMS AND DRYING AHS-B SERIES - 7

STAINLESS STEEL WIRE TRAYSReference BAH-21 BAH-50 B BAH-75 B Dimensions Exterior L x D mm For autoclaves with the following chamber volumes 22 L 55 L 79 L STAINLESS STEEL WIRE HORIZONTAL BASKET Reference RB-AH-21 RB-AHS-50 RB-AHS-75 For autoclaves with 22 L the following chamber 55 L volumes 79 L STAINLESS STEEL BAG SUPPORT* Reference BAP-21 BAP-75 Slots / support For autoclaves with the following chamber volumes 20 22 L 1 *This accessory can be customized in size according to each client needs. For more information please contact us. STAINLESS STEEL CONTAINERS WITH FILTER ON THE...

Open the catalog to page 7
BENCHTOP AUTOCLAVES WITH PREVACUUMS AND DRYING AHS-B SERIES - 8

• After installing this accessory, the temperature regulation of the sterilization cycle can either be controlled by the main chamber temperature sensor or both the main chamber temperature sensor and the temperature sensor of the flexible core temperature probe. • The temperature control by the flexible core temperature probe is especially advantageous for processes involving the sterilization of large volumes of liquids, where the sterilization process is regulated by both the temperature achieved in the center of the liquid sample as well as the temperature achieved in the sterilization...

Open the catalog to page 8
BENCHTOP AUTOCLAVES WITH PREVACUUMS AND DRYING AHS-B SERIES - 9

BOWIE DICK TEST PACK STERILIZATION CONTROL TAPE • Class 1 indicator for steam sterilization. The change of color indicates that the materials have been processed, without being a guarantee of proper sterilization, additional methods are needed such as biological indicators (EN ISO 11138). • Tape roll of 50 m x 19 mm. Reference: TEST-CT • Class B indicator that checks the correct steam penetration in porous loads. • Indicator printed with non-toxic inks and laminated. • Box of 20 tests. Reference: TEST-BD AUTOMATIC WATER FILLING KIT • Water pump to automate the feed of the independent clean...

Open the catalog to page 9
BENCHTOP AUTOCLAVES WITH PREVACUUMS AND DRYING AHS-B SERIES - 10

+: Recommended V: Included O: Optional 10 FRONT-LOADINGBENCHTOPAUTOCLAVESWITHPREVACUUMSANDDRYING - AHS-B SERIES RAY

Open the catalog to page 10
BENCHTOP AUTOCLAVES WITH PREVACUUMS AND DRYING AHS-B SERIES - 11

Specifications References Total/usable volume of the chamber L Usable dimensions of the chamber 0 max. x D mm Volume of the built-in clean water tank L Volume of the built-in waste water tank L Maximum number of trays Tray size L x D mm Net weight Kg Power W Standard voltage* V Frequency Hz *Other voltages and electrical configurations available on request. Safety features • Safety valve. • Safety thermostats with manual rearm for the heating jacket and the steam generator. • Pneumatic door blocking system while positive pressure exists inside the sterilization chamber. • Open door sensor....

Open the catalog to page 11