General Specifications:
The RaySafe X2 complies with various international standards including IEC 61326-1 for EMC, IEC 61010-1 for safety, and IEC 61674 for X-ray meters. It requires only one exposure and includes USB cables of varying lengths. The base unit measures 34 x 85 x 154 mm and weighs 521 g. It operates between 15 – 35 °C and can be stored between -25 – 70 °C. The device is powered by a rechargeable Li-ion battery, providing approximately 10 hours of intensive usage. It features a 4.3” LCD touch display and can store around 10,000 exposures. Data handling is facilitated by RaySafe View software, which can export data to Excel. The device holds a PTB certificate and follows the GUM by ISO for uncertainty definition.
X2 mAs Specifications:
The mAs range is 0.001 – 9999 mAs with a resolution of 0.001 mAs and an uncertainty of 1%. The mA range peaks at 0.1 – 1500 mA with a resolution of 0.01 mA. Time measurement ranges from 1 ms to 999 s with a resolution of 0.1 ms and a bandwidth of 1 kHz. Pulse measurements range from 1 to 9999 pulses with a pulse rate range of 0.1 – 200 pulses/s.
X2 R/F Sensor Specifications:
Weighing 42 g, the sensor measures 14 x 22 x 79 mm. It features active compensation for beam quality independence. Dose range is 1 nGy – 9999 Gy with an uncertainty of 5%. Dose rate ranges from 1 nGy/s to 500 mGy/s. The kVp range is 40 – 150 kVp with a minimum dose of 50 µGy and an uncertainty of 2%.
X2 MAM Sensor Specifications:
This sensor also weighs 42 g and measures 14 x 22 x 79 mm. It supports various beam qualities and compensations. Dose range is 1 µGy – 9999 Gy with an uncertainty of 5%. The kVp range is 18 – 50 kVp, with specific requirements for measurements above 40 kVp.
X2 Light Sensor Specifications:
Weighing 136 g, the sensor measures 48 x 60 x 68 mm. It complies with DIN 5032 part 7 class B and other standards. Luminance range is 0.01 – 10,000 cd/m2 with a resolution of 0.001 cd/m2 and an uncertainty of 3%.
X2 Survey Sensor Specifications:
Weighing 140 g, the sensor measures 14 x 66 x 192 mm. It features manual and auto trig modes. H*(10) range is 0 nSv – 9999 Sv with an uncertainty of 10%. Air kerma range is 0 nGy – 9999 Gy with an uncertainty of 5%.
X2 CT Sensor Specifications:
Weighing 86 g, the sensor measures 14 x 22 x 219 mm. It is designed for IEC 60601-2-44 compliance. Dose range is 10 µGy – 999 Gy with an uncertainty of 5%.
X2 Volt Sensor Specifications:
Weighing 55 g, the sensor measures 17 x 23 x 93 mm. It features a BNC connector and measures voltage in the range of ± 16 V with an uncertainty of 1%.
Conclusion:
RaySafe X2 offers a comprehensive suite of sensors and features for radiographic imaging, ensuring high accuracy and compliance with international standards. It is designed to reduce unnecessary radiation exposure and enhance safety culture in medical environments.
See more

Catalog excerpts

RaySafe X2 Specifications-2

X2 GENERAL EMC SAFETY X-RAY METERS STANDARD EXPOSURES NEEDED USB CABLES SIZE BASE UNIT WEIGHT BASE UNIT OPERATING TEMPERATURE STORAGE TEMPERATURE POWER SOURCE BATTERY TIME BATTERY TESTED DISPLAY MEMORY SOFTWARE PTB CERTIFICATE According to IEC 61326-1 According to IEC 61010-1 Complies with IEC 61674 One 2 m (6.6 ft), 5 m (16.4 ft) and 5 m active extender 34 x 85 x 154 mm (1.3 x 3.3 x 6.1 in) 521 g (18.4 oz) 15 - 35 °C (59 - 95 °F) -25 - 70 °C (-13 - 158 °F) Rechargeable Li ion battery ~ 10 hours intensive usage According to UN 38.3 4.3” LCD with capacitive touch ~ 10000 latest exposures RaySafe...

 Open the catalog to page 2
RaySafe X2 Specifications-3

kVp TF DOSE RANGE UNCERTAINTY DOSE RATE RANGE RESOLUTION TRIG LEVEL UNCERTAINTY Beam quality independent for the following ranges: DOSE/DOSE RATE 40 - 150 kVp, 1 - 14 mm Al HVL 40 - 150 kVp, up to 1 mm Cu 60 - 120 kVp, up to 1 mm Cu 1 nGy - 9999 Gy (0.1 pR - 9999 R) 5 % or 5 nGy (0.5 pR) 1 nGy/s - 500 mGy/s (5 pR/min - 3400 R/min) 1 nGy/s (5 pR/min) 50 nGy/s (340 pR/min) 5 % or 10 nGy/s (70 pR/min) x duty cycle MINIMUM DOSE RATE (PEAK) 10 pGy/s (70 mR/min) UNCERTAINTY 2 % MINIMUM DOSE RATE (PEAK) 0.5 pGy/s (3.5 mR/min) at > 70 kV 2.5 pGy/s (17 mR/min) at 50 kV TOTAL FILTRATION RANGE 1.5 - 35...

 Open the catalog to page 3
RaySafe X2 Specifications-4

ACTIVE COMPENSATION Beam quality independent for the following ranges: DOSE/DOSE RATE & HVL No selections needed. With or without paddle, with or without phantom. Mo/Mo, Mo/Rh 20 - 40 kVp W/Al, Rh/Rh, Rh/Al Mo/Cu, Rh/Cu, W/Cu, W/Ti 40 - 50 kVp kVp User selectable beam qualities. Paddle compensation available when relevant. W/Ag 20 - 40 kVp Measuring above 40 kVp requires an X2 R/F Sensor + 2 mm Al (incl.) W/Rh 18 - 40kVp Mo/Rh 32 - 40 kVp using + 2 mm Al (incl) Mo/Cu, W/Cu, W/Ti 40 - 50 kVp, using the X2 R/F Sensor DOSE RATE RANGE 10 gGy/s - 300 mGy/s Measuring above 40 kVp requires an X2 R/F...

 Open the catalog to page 4
RaySafe X2 Specifications-5

X2 LIGHT SENSOR ILLUMINANCE Complies with relevant parts of AAPM TG18, IEC 62563-1 and IEC 61223-2-5. RANGE RESOLUTION MEASUREMENT AREA DEVIATION FROM HUMAN EYE V(λ) (f 1’) < 3 % (see figure Cosine Response) APERTURE ANGLE < 3 % (see figure Photopic Response) DEVIATION FROM HUMAN EYE V(λ) (f 1’) < 3 % (see figure Photopic Response) Photopic Response Cosine Response

 Open the catalog to page 5
RaySafe X2 Specifications-6

X2 SURVEY SENSOR AIR KERMA RATE ACTIVE COMPENSATION H*(10) – when selecting Sv Air kerma – when selecting Gy or R 5 % or 0.3 µGy/h (0.03 mR/h), RQA 50 – 150 kV 10% or 0.3 µGy/h (0.03 mR/h), N-series 40 – 150 kV TRIG MODES MANUAL Manual start and stop of measurement Trig level (N80): 10 µGy/h (1.2 mR/h) or 20 µSv/h MEAN ENERGY RANGE UNCERTAINTY RANGE RESOLUTION DEFINING STANDARD MINIMUM DOSE RATE TIME RANGE RANGE UNCERTAINTY MINIMUM DOSE RATE

 Open the catalog to page 6
RaySafe X2 Specifications-7

X2 VOLT SENSOR WEIGHT SIZE SIZE 0 STANDARD ACTIVE LENGTH ENERGY DEPENDENCE AUTOMATIC ENVIRONMENTAL COMPENSATION DOSE RANGE UNCERTAINTY DOSE LENGTH PRODUCT RANGE UNCERTAINTY DOSE RATE RANGE UNCERTAINTY TIME RANGE RESOLUTION BANDWIDTH UNCERTAINTY WAVEFORMS RESOLUTION BANDWIDTH 86 g (3.0 oz) 14 x 22 x 219 mm (0.5 x 0.9 x 8.6 in) 12.0 mm (0.47 in) For measurements in accordance with IEC 60601-2-44 100 mm (3.94 in) < 5 % for 70 - 150 kV (RQR, RQA and RQT beam qualities) 55 - 110 kPa, 15 - 35 °C (59 - 95 °F) WEIGHT SIZE INPUT TERMINAL VOLTAGE RANGE UNCERTAINTY TRIG LEVEL TIME RANGE RESOLUTION UNCERTAINTY...

 Open the catalog to page 7
RaySafe X2 Specifications-8

User-friendly state of the art technology and maximum accuracy form the backbone of the RaySafe product range. All solutions meet the company’s mission to reduce unnecessary radiation exposure and to establish a greater safety culture. RaySafe offers a variety of comprehensive solutions for radiographic imaging rooms. Our devices add value through the collection of radiation information and the ability to easily share results with relevant stakeholders.

 Open the catalog to page 8