Endodontic Irrigation Device
Open the catalog to page 1
VATEA Irrigation System Instructions for use Indications for use The VATEA system is intended to be attached to dental handpieces to deliver irrigation during endodontic procedures. Device description The VATEA system is a self-contained, fluid delivery unit consisting of a pump console, irrigation reservoir and disposable tubing intended to be used in endodontics to provide irrigation while attached to dental handpieces. The VATEA is an electronically operated device, powered by a rechargeable battery. Flow rate and timer is adjusted via control panel display, and the irrigation is toggled...
Open the catalog to page 2
Description of the front panel Battery Indication LED Flow Rate / timer Control Buttons Fluid Connection Socket (luer-lock) Foot Pedal Connection Socket Figure 1: Front panel of the VATEA
Open the catalog to page 3
Description of the rear panel Liquid Reservoir Master On/ Off Power Switch Electrical Charger Socket Figure 2: Rear panel of the VATEA
Open the catalog to page 4
3. Precautions This device is to be used by dental professionals only. Batteries should be charged until indicating LED light changes to green. See instructions regarding batteries in paragraph 13. Electrical charger (medical AC/DC power supply adapter - complies to UL60601-1) Make sure that the silicone tube and foot pedal cable are properly connected. Use only sodium hypochlorite in the reservoir. 4. Contraindications Do not use without a rubber dam.
Open the catalog to page 5
Getting started Unpack the system carefully and place it on a flat, stable surface. Connect the electrical charger power cable to the electrical charger socket on the rear panel. Charge the batteries until indicating LED light changes to green. Connect the foot pedal plug and the fluids connector to the socket on the front panel: Position the foot pedal plug in a way that the marking arrow will be placed in front of the marking dot, as shown in figure 3(a). Push the plug until you hear a click sound. Tighten the luer-lock connector. To release the foot pedal plug turn it to the left as...
Open the catalog to page 6
How to operate the VATEA system Fill the reservoir with sodium hypochlorite, up to the 400ml indication line. It is required to use a funnel. Secure the cap.
Open the catalog to page 7
6.2 Assemble the silicone ring on the handpiece, as shown in figure 4: Figure 4: Assembly of the silicone ring on the handpiece 6.3 Turn on the system using the master power switch, located on the rear panel. Master On/ Off Power Switch
Open the catalog to page 8
6.4 Flow rate is displayed when the system starts up and/or when adjusting the flow rate. 6.5 Timer will start running when irrigation is switched on with the foot pedal. 6.6 The timer will count 10 minutes and then reset to "00" 6.7 During the first minute, the time is displayed in seconds (1-59 sec). 6.8 The following 9 minutes are presented in minutes (left digit), together with an indication of active irrigation by cyclic moving segments (right digit).
Open the catalog to page 9
A beeping sound will be heard after 2 minutes of operation. Stopping the irrigation during the first 4 minutes stops the timer but does not reset it to "00". Stopping the irrigation after 4 minutes with the foot pedal resets the timer. An integrated buzzer is used for notifications and alarms (see section 9). (if it is desired to mute the buzzer the flow rate should be adjusted to zero and then the "+/-" control buttons pressed simultaneously. To turn on the buzzer again repeat this action). When working with the electrical charger connected – there will be a steady LED notification for the...
Open the catalog to page 11
7. Equipment supplied • VATEA irrigation system. • 10 disposable Silicone tube 1.5m with silicone ring and luer-lock (cat no. RDN-MCH0027). • Foot pedal (cat no. RDN-MCH0035). • Electrical charger (cat no. RDN-MCH0023). • Funnel. • Instructions for use. Use only certified parts supplied by the manufacturer.
Open the catalog to page 12
10.1 Silicone tube and ring are disposable and should be discarded after use. Before initial use the tube should be sterilized in the dental clinic autoclave at 134o C for 3 minutes. 10.2 VATEA outer surfaces - Wipe with disinfection wipes after each treatment. 10.3 Charging - Connect the electrical charger until indicating LED light changes to green. 11. Technical data Power inlet: Supply voltage: Current: Dimensions in mm: (HxWxD) Weight in kg: Liquid reservoir capacity: Storage temperature: Operational temperature Type of protection against electrical shock: Internally powered equipment....
Open the catalog to page 15
Equipment not suitable for use in the presence of flammable anesthetic mixture with air or with oxygen or nitrous oxide. Mode of operation: Intermittent operation: Operation time: 20 minutes; Rest time: 20 minutes. Precautions in Battery Handling Be sure to connect the positive and negative battery terminals to the corresponding terminals of the application device, in order to prevent reverse charging. Never incinerate the battery. Avoid subjecting a battery to strong vibrations, pressure or impact. Never carry a battery with other metallic belongings to avoid shortcircuiting. Never...
Open the catalog to page 16
Description Read Instruction for use prior use Catalogue number Serial number Type BF applied part CAUTION, consult accompanying documents TUVus mark
Open the catalog to page 17
Manufacturer ReDent-Nova Ltd. 15 Ha’taasia Street, POB 4159 Ra’anana 43000, Israel. Tel: +972 9 7445130 / 70, Fax: +972 9 7445190 E-mail: office@redent.co.il European Representative Obelis s.a Boulevard Général Wahis 53 1030 Brussels, BELGIUM Tel: + (32) 2. 732.59.54 Fax: + (32) 2.732.60.03 E-Mail: mail@obelis.net