UNISTAT® Bilirubinometer

UNISTAT® Bilirubinometer

UNISTAT® Bilirubinometer

Product catalog summary
Overview: The Reichert UNISTAT® Bilirubinometer is a point-of-care device designed for the rapid and accurate measurement of Total Serum Bilirubin (TSB) in newborns. It utilizes a photometric analyzer to provide results without the need for dilution or complex chemistry, making it suitable for all skin types and ethnicities.
Clinical Implications: Monitoring bilirubin levels in newborns is crucial to prevent hyperbilirubinemia and kernicterus. The UNISTAT® Bilirubinometer offers a reliable alternative to visual detection methods, which can be inaccurate due to skin pigmentation.
Key Features:
  • Simple operation with results in 5 seconds.
  • 99% correlation with the diazo Jendrassik-Grof method.
  • Low operating costs with disposable and reusable cuvettes.
  • Complies with the 2004 AAP guidelines for hyperbilirubinemia management.
Principles of Operation: The device uses a direct spectrophotometer to measure bilirubin levels. A light beam is split and passed through filters at 460nm and 550nm to differentiate between bilirubin and oxyhemoglobin, ensuring accurate readings even in hemolyzed samples.
Technical Specifications:
  • Sample Size: 20µL undiluted serum.
  • Turnaround Time: 5 seconds for results, 15 seconds total.
  • Readout Range: 0-40 mg/dL (0-684 µmol/L).
  • Accuracy: ±5% for 0-30 mg/dL, ±10% for 30.1-40 mg/dL.
  • Operating Voltage: 115V or 230V, 50/60 Hz.
  • Approvals: ETL, CE.
Accessories and Options: The device comes with disposable cuvettes, calibration and high-level check cuvettes, and a replacement lamp. Additional accessories include various cuvettes for different assay values.
Contact Information: For more details, visit the Reichert website or contact Charlie Smith at 716-319-4777 or [email protected].
See more

Catalog excerpts

UNISTAT® Bilirubinometer-1

Reichert UNISTAT® Bilirubinometer Chemistry-free, no dilution solution for point-of-care Total Serum Bilirubin measurement • imple to operate. Requires no complicated chemistry or waiting time. S • NISTAT®’s photometric analyzer measures Total Serum Bilirubin (TSB) U concentration easily and accurately within 5 seconds. • NISTAT® provides 99% correlation to widely recognized method U (diazo Jendrassik-Grof ) for accurate TSB measurement.1 • Low operating cost with no expensive slides, carousels, or reagents. UNISTAT® uses an inexpensive disposable specimen cuvette and re-usable glass calibration cuvette. Clinical Implications: Bilirubin is a byproduct of the breakdown of hemoglobin by the liver. Newborns may have difficulty with filtering blood and excreting this excess bilirubin. Because of this risk of bilirubin toxicity, newborns must be monitored to identify those who might develop hyperbilirubinemia and kernicterus. The Total Serum Bilirubin test measures the concentration of bilirubin in the newborn. This detection is needed before a diagnosis and treatment program can be initiated. • study published in the Journal of Perinatology showed that UNISTAT® meets the A 2004 AAP guidelines “Management of Hyperbilirubinemia in the Newborn Infant 35 or More Weeks of Gestation” for physicians taking TSB measurements as a point-of-care test (J Perinatol. 2006; 26:100-105). 2* Point-of-care testing for all skin types and ethnicities. Reichert’s UNISTAT® Bilirubinometer uses a proven photometric analyzer to provide accurate and rapid measurements of total serum bilirubin (TSB) concentration in undiluted serum. The photometric analyzer is the ideal STAT-level analyzer for determining TSB in newborn infants. Other methods use visual detection of skin color. This has proven to be unreliable where skin color pigments can mask the jaundice symptom. continued Contact Charlie Smith: 716-319-4777 [email protected] Reichertai.com H A Barko, G L Jackson, W D Engle; Evaluation of a point-of-care direct spectrophotometric method for measurement of total serum bilirubin in term and near-term neonates; Journal of Perinatology; 01 Feb 2006; 26, 100 - 105 1, 2 *Mention of the AAP guidelines does not imply endorsement of any product by the AAP.

 Open the catalog to page 1
UNISTAT® Bilirubinometer-2

Reichert UNISTAT® Bilirubinometer Principles of Operation: The Reichert UNISTAT® is a direct spectrophotometer that measures total bilirubin in neonatal serum samples. Dilution is not necessary. A beam of light is passed through the sample cuvette. The beam is split by a dichroic mirror and directed toward the 460nm and 550nm filters. The transmittance of the light exiting the filters is measured by photodetectors. The control electronics calculate total bilirubin based on output from the photodetectors. Since peak absorbance is shared by bilirubin and oxyhemoglobin at 460nm, it is necessary...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.