Tono-Pen AVIA Cleaning Instructions

Tono-Pen AVIA Cleaning Instructions

Product catalog summary
Introduction: The TONO-PEN AVIA® tonometer may experience measurement difficulties or display a 'Fail' message if its tip is dirty. Cleaning is necessary when contaminants are present between the sensor and housing, affecting sensor movement and causing erratic readings.
Cleaning Procedure:
  1. Remove the OCU-FILM® + tip cover if installed.
  2. Use Dust-Off JR canned air to blow air into the sensor end of the tonometer for approximately 3 seconds, ensuring contaminants are expelled from the airspace.
  3. Allow the tip to warm to room temperature after cleaning.
  4. Perform tonometer verification as per the Instruction Manual. Repeat cleaning if verification fails, but do not exceed three consecutive cleanings. Contact Reichert if issues persist.
Storage and Usage Notes:
  • Always store the tonometer with an OCU-FILM+ tip cover to protect it from dirt.
  • Never use the tonometer without the approved OCU-FILM+ tip cover.
Cleaning Schedule: The frequency of cleaning depends on the number of patients per week:
  • 10 patients: Clean every 30 days.
  • 100 patients: Clean every 15 days.
  • 300 patients: Clean every 7 days.
  • 600 patients: Clean daily.
Service and Support: For service, support, and accessories like Dust-Off JR and OCU-FILM+ tip covers, visit www.tonopen.com.
Manufacturer Information: Reichert, Inc., ISO 9001 Certified, ensures quality manufacturing processes. Contact details: 3362 Walden Avenue, Depew, NY 14043, Toll-Free: 888-849-8955, Tel: (716) 686-4500, Fax: (716) 686-4545, www.reichert.com.
See more

Catalog excerpts

Tono-Pen AVIA Cleaning Instructions-1

TONO-PEN AVIA® Tonometer Cleaning Instructions The TONO-PEN AVIA tonometer may have difficulty taking measurements or display “Fail” after a verification when its tip is dirty and requires cleaning. When the tip of the tonometer has dirt and contaminants in the airspace between the sensor and the housing, cleaning of the tip is necessary. When the airspace contains contaminants, the sensor cannot move freely and the tonometer may have erratic readings and then show a “Fail” verification. To clean the tonometer, perform the following steps: 1.) Remove OCU-FILM® + tip cover from the tonometer, if one is installed. 2.) Using Dust-Off JR canned air, place the sensor end of the tonometer against the outlet of the canned air as shown in Figure 1. 3.) Blow the canned air into the tip of the tonometer for approximately 3 seconds. Note: t is necessary to blow canned air directly into the tip, I so that the contaminants are pushed out of the airspace between the sensor and the housing. Figure 1 4.) After cleaning the tip of the tonometer with compressed air, the tip will be cold. Allow the tip of the tonometer to warm to room temperature. 5.) Perform the tonometer verification as indicated in the Instruction Manual. Note: f the tonometer does not pass verification, then repeat the above cleaning instructions. Do not I clean more than 3 times in a row. If the tonometer still will not pass verification, contact Reichert. Note: lways store the TONO-PEN AVIA tonometer with an OCU-FILM+ tip cover installed to protect A the tonometer tip from dirt and contaminants. Note: Never use the TONO-PEN AVIA tonometer without an OCU-FILM+ tip cover installed. Note: CU-FILM+ tip covers are the only manufacturer approved covers for use with the O TONO-PEN AVIA tonometer. Suggested Cleaning Schedule Number of Patients per Week 10 100 300 600 Number of Days Between Cleaning 30 15 7 1 For Service and Support information and to order TONO-PEN AVIA tonometer accessories, such as Dust-Off JR and genuine OCU-FILM + tip covers, please visit www.tonopen.com Important Notice For your own protection, only have your TONO-PEN AVIA tonometer serviced by Reichert. 230650-104 REV A Reichert, Inc. 3362 Walden Avenue, Depew, NY 14043 Toll Free: 888-849-8955 • Tel: (716) 686-4500 Fax: (716) 686-4545 • www.reichert.com ISO 9001 Certified Reichert products are designed and manufactured under quality processes meeting ISO 9001 requirements. Reichert, TONO-PEN AVIA and OCU-FILM are registered trademarks of Reichert, Inc.

 Open the catalog to page 1

All Reichert catalogs and technical brochures

  1. Tono-Vera®

    8  Pages

  2. Phoroptor® VRx

    12  Pages

Archived catalogs

  1. ML1

    2  Pages

  2. Tono-Pen®

    10  Pages

  3. TONO-PEN VET?

    2  Pages

  4. AP250

    2  Pages

  5. Reflex

    2  Pages

  6. AL700

    2  Pages

  7. Phoroptor®

    4  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.