1. Catalogs
  2. revvity
  3. Tri-Carb 5110TR liquid scintillation counter
video corpo
  • Products
  • Catalogs
  • News & Trends
  • Exhibitions

Tri-Carb 5110TR liquid scintillation counter

Tri-Carb 5110TR liquid scintillation counter

Tri-Carb 5110TR liquid scintillation counter

Product catalog summary
Overview
The Tri-Carb 5110TR is a premium liquid scintillation counter designed for detecting ultra-low-level counts in mixed and unknown samples. It utilizes advanced technologies such as Dual Phase Separation and PSA Histogram for enhanced accuracy.

Key Features
  • TR-LSC: Time-resolved liquid scintillation counting for high sensitivity and low background.
  • QuantaSmart Software: Provides a secure multiuser environment with unlimited assays.
  • Alpha-Beta Separation: Efficient separation of alpha and beta counts.
  • PSA Histogram: Offers higher resolution and reduces counting time by up to 90%.
  • Dual PSA Discriminator: Reduces isotope spillover and increases sensitivity.
  • Ultralow Level Count Mode: Enhances system sensitivity.
  • PAC: Reduces background from optical crosstalk.
  • SpectraWorks 2: Determines optimal counting regions and calculates figures of merit.
  • Enhanced IPA Database: Monitors instrument performance over time.

Additional Features
  • Dynamic color correction, auxiliary spectrum memory, and bi-directional sample conveyer.
  • Direct DPM determination, luminescence detection and correction.
  • Automatic power-fail recovery and anti-jam recovery.
  • Multi-parameter linear multichannel analyzer for extended dynamic quench range.
  • Automatic efficiency control and spectral unfolding for dual label samples.
  • 3D spectral mapping and SpectraBase counting and data management system.
  • Auto QA for automatic quality assurance reporting.
  • Computer-aided diagnostics and customizable data reporting.

Physical and Environmental Specifications
  • Dimensions: Height: 18.5 in, Width: 40.5 in, Depth: 32 in (44 in with temperature control).
  • Weight: 477 lb (523 lb with temperature control).
  • Electrical Requirements: 100-240 Vac, 50/60 Hz.
  • Operating Temperature: 15 to 32 °C (59-90 °F).

Performance Metrics
  • Energy Range: 0-2,000 Kev.
  • Efficiency: 3H: 63%, 14C: 95%.
  • Background: 3H: 17 CPM, 14C: 26 CPM.
  • Figure of Merit (E2/B): Normal mode: 3H: 180, 14C: 360; Ultra low level mode: 3H: 500, 14C: 940.

Compliance and Safety
  • Tested and approved for safety, radiated emissions, and immunity according to UL, IEC61010, and CE standards.
See more

Catalog excerpts

Tri-Carb 5110TR liquid scintillation counter-1

Tri-Carb 5110TR liquid scintillation counter Exclusive standard features • TR-LSC (Time-resolved liquid scintillation counting) for high sensitivity, low background liquid scintillation counting • Single/Dual labeled color-corrected DPM to determine absolute activity level of the sample • QuantaSmart™ Software a robust multitasking, easy networking environment with unlimited assays in a secure multiuser environment • Alpha-Beta Separation for efficient separation of Alpha and Beta counts in mixed samples The Tri-Carb® 5110TR liquid scintillation • Histogram provides higher resolution of Alpha-Beta PSA separation with up to 90% reduction in time required to count alpha/beta standards compared to legacy method counter is a fully loaded premium instrument capable of more accurately detecting Ultra Low Level counts in mixed and unknown samples using proprietary Dual Phase Separation and PSA Histogram technology. • ual PSA Discriminator further reduces isotope spillover and D increases Quality Metric for maximum sensitivity and lowest MDA • Ultralow level count mode increases system sensitivity (E²/B) • (Pulse Amplitude Comparison) decreases the component PAC of background produced by optical crosstalk • SpectraWorks 2 spectral analysis package that determines optimal counting regions, and calculates figure of merit automatically • Enhanced IPA (Instrument Performance Assessment) database monitors efficiencies, backgrounds, E2/B and Chi-square values for 3H and 14C over the l

 Open the catalog to page 1
Tri-Carb 5110TR liquid scintillation counter-2

Tri-Carb 5110TR liquid scintillation counter • Replay reanalyzes sample counts without recounting samples • Dynamic color correction ensures accuracy of tracking lower energy sample spectra over a wide quench range • Auxiliary spectrum memory stores rejected events for future analysis • Bi-directional sample conveyer with a capacity of up to 408 (20 mL vials) or 720 (4 or 7 mL vials) • Varisette™ sample cassette for intermixing vial sizes without special adapters • Operational-status LED indicator for clear assay status updates at a glance Additional standard features • Direct DPM determines...

 Open the catalog to page 2
Tri-Carb 5110TR liquid scintillation counter-3

Tri-Carb 5110TR liquid scintillation counter • Printed header contains instrument serial number, user ID, and drive and path of all electronic stored data for GLP compliance • Password protection prevents unwanted changes to saved assays • Half-life correction adjusts for decay to any date and time • Unit conversion activity can be reported in becquerels, microcuries, or picocuries • Auto QA (Automatic quality assurance) automatically prints reports for backgrounds, efficiencies, E2/B, and Chi- square values, results can be transmitted via RS-232 for archiving • Percent of standard calculations...

 Open the catalog to page 3
Tri-Carb 5110TR liquid scintillation counter-4

Tri-Carb 5110TR liquid scintillation counter Note: The efficiencies, backgrounds, and E2/B values for the Normal Count Mode were determined using Revvity sealed large vial glass standards set P.N. 6008500 verified with NIST standard activity. The ULLCM values are determined using Revvity low level sealed large glass vial standards set P.N. 6018914 verified with NIST standard activity. No maximum is specified for background. Safety, radiated emissions and immunity: The Tri-Carb 5110TR has been tested and approved for safety, radiated emissions and immunity according to the standards of UL, IEC61010...

 Open the catalog to page 4
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.