VETERINARY

VETERINARY

VETERINARY

Product catalog summary
Technical Specifications:
Dimensions: 310 (L) x 260 (W) x 210 (H) mm
Weight: 7 kg
Power Supply: 85-264 V; 47-63 Hz, AC input power: 45 to 100 VA
Battery: 2 hours autonomy, Built-in charger
Environmental Conditions: Operating temperature: 0 ºC to 40 ºC; 32 ºF to 104 ºF, Relative humidity: 5% to 90% non-condensing
Measurement Procedures:
Oscillometric method with different cuff sizes
Measurement time: Typical 25 to 45 s, Maximum 120 s
Blood Pressure Range: 10 to 250 mmHg, Overpressure limit: 300 mmHg
Pulse Rate: Range 40 to 250 bpm, Resolution 1 bpm, Accuracy ±3%
Operation Modes:
Manual and Automatic (1 to 120 min.)
Fully isolated, defibrillator protected > 5 kV
Display and Recording:
10.4 inches TFT colour display, Resolution: 640 x 480 pixels
Sweep speed: 12.5 / 25 / 50 mm/s
Up to seven traces on screen
Complies with IEC 601 standards
Temperature Measurement:
Range: 15 to 45 ºC; 59 to 113 ºF, Resolution: 0.1 ºC; 0.2 ºF, Accuracy: ±0.1 ºC
Use of oesophageal, rectal or skin probes
ECG and Heart Rate:
Leads: I, II, III, V, aVL, aVF, aVR
Heart rate range: 20 to 800 bpm, Accuracy: ±2%
Signal features: Diagnostic mode: 0.05 to 100 Hz, Monitor mode: 0.5 to 25 Hz
Alarms and Trends:
Optical and acoustic alarms for all parameters with three priority levels
Trends up to 96 hours, graphical and tabular display
Ordering Information:
Various configurations available, including options for IBP, CO (Sidestream/Mainstream), and Printer
Contact Information:
RGB Medical Devices, Alfonso Gómez, 42, 28037 Madrid SPAIN
Phone: +34 913 042 011
Website: www.rgb-medical.com
See more

Catalog excerpts

VETERINARY-1

VISION VET COMPLETE VITAL SIGNS VETERINARY MONITOR • ECG 7 LEADS • NIBP Non Invasive Blood Pressure * ST SEGMENT • SPO2 PULSE OXIMETRY OPTIONS • INVASIVE BLOOD PRESSURE (2) • ANAESTHETIC AGENTS • C02 Capnography • THERMAL PRINTER Omicrom Vision VET incorporates latest innovations for small, medium and big size animals monitoring. Heart Rate up to 800 bpm. Specific algorithm for QRS and Respiration detection. Optional module for “gating” to PET, SPECT, CT and US devices.

 Open the catalog to page 1
VETERINARY-2

Technical Data Omicrom® Vision VETGENERAL Dimensions: 310 (L) x 260 (W) x 210 (H) mm Weight: 7 kg Power Supply: 85-264 V; 47-63 Hz AC input power: 45 to 100 VA Battery: 2 hours autonomy Built-in charger Environmental conditions: Operating temperature: 0 °C to 40 °C; 32 °Fto104°F Relative humidity: 5% to 90% non-condensing Display: TFT colour display. Up to seven (7) traces on screen. Size: 211 x 158 mm Diagonal 264 mm /10.4 inches Resolution: 640 x 480 pixels Sweep speed : 12.5 / 25 / 50 mm/s Recorder: 4 channel (optional). Safety Standard: Complies with IEC601 standards. Patient leakage current...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.