Multi - Parameter Patient Simulator The Rigel 333 is a powerful, comprehensive patient simulator with a number of unique features. It provides a full 12 lead ECG signal with more than 40 arrhythmias, dual blood pressure, respiration, and temperature. It also delivers square, sine, and a triangle performance waveforms. Rigel 333 arrhythmia selections include pacer waveforms, ST segment changes, and both atrial and ventricular waveforms. The dual blood pressure and respiration waveforms are synchronized to the ECG and track with all arrhythmia selections. Multiple static and dynamic pressure readings are selectable as per apnea events Operation of the hand - held Rigel 333 is simple via menu driven tactile keys utilizing a large LCD Display. The instrument may be either AC or Battery operated. Testing Features: • Dual Blood Pressure Operating Features: • Small, Compact Size • Menu Driven Operation • Tactile Keypad
Open the catalog to page 1Specifications: ECG Normal Sinus: Rate: Amplitude: High Level Output: Performance Waveforms: Sine, Triangle, Square Rate: 0.1 to 100 Hz 0.1 to 1 Hz in 0.1 Hz increments 1 to 10 Hz in 1 Hz increments 10 to 100 Hz in 10 Hz increments Tall T Wave Test Waveform@ 70 BPM Atrial Pacer Ventricular Pacer* Paced Rhythms - Variable from 0.1 to 2mS R Wave detection - Variable form 10 to 120mS. Pulse: 2mS Pulse of 1mV amp @ 4 sec intervals Pacers ST Segment Analysis: ST Elevation and Depression ST Depression Positive and Negative Slope Myocardial Infarction Arrhythmias: Atrial Ventricular Blocks Respiration:...
Open the catalog to page 2Blood Pressure: Impedance: 350 Ohms Excitation: 2 to 16 Volts Sensitivity: 5uV/V/mmHg Static: 0, 5, 10, 20, 25, 30, 40, 50, 100, 150, 200, 300 mmHg. Dynamic: 50/10, 60/20, 70/30, 80/40, 100/60, 120/80 mmHg. Accuracy: (±2% FS. ±1 mmHg Synchronized with all normal sinus rates Physiologically track all arrhythmia selections. Temperature: Compatibility: YSI 400/700 25/37 Degree Centigrade Accuracy: ±1% of reading General: Display: Key Pad: Output Connectors ECG: High Level: Pressure: Temperature: Physical Size: Case: Weight: 16 Character 2 Line LCD 8 tactile keys Ð F1, F2, F3, F4, >, <, ON/OFF, Menu...
Open the catalog to page 32 Pages
3 Pages
1 Page
3 Pages
4 Pages
2 Pages
19 Pages
19 Pages
29 Pages
18 Pages
17 Pages
6 Pages
12 Pages
6 Pages
5 Pages
2 Pages
8 Pages
5 Pages
2 Pages
12 Pages
2 Pages
6 Pages
20 Pages
5 Pages
5 Pages
3 Pages
20 Pages
20 Pages
6 Pages