1. Catalogs
  2. RIGEL Medical
  3. Rigel 62353 Electrical Safety Analyser

Rigel 62353 Electrical Safety Analyser

Rigel 62353 Electrical Safety Analyser

Rigel 62353 Electrical Safety Analyser

Product catalog summary
Overview: The Rigel 62353 is a portable medical electrical safety tester designed for in-service and post-repair testing of medical electronic devices, adhering to the IEC 62353 standard. It combines automatic and manual testing capabilities with data logging and asset management features, making it suitable for multi-site use.
Key Features:
  • Portability: Compact and lightweight design with a long-life battery for reduced downtime and enhanced portability.
  • Standards Compliance: Tests according to IEC 62353, AAMI/NFPA, and AS/NZ3551 standards.
  • Data Management: Large internal memory for storing up to 10,000 test records, with easy data transfer to PC-based systems.
  • Custom Test Settings: Users can configure or modify test sequences to meet specific needs, including visual and acceptance test routines.
  • Med-eBase Software: Compatible with Med-eBase software for asset management, test routine configuration, and certificate generation.
Specifications:
  • Earth Continuity: Uses a 2-wire technique with a test current of >200mA DC, measuring range from 0.001 to 19.9 ohms.
  • Insulation Resistance: Measures from 0.01Mohms to 100Mohms with a test voltage of 250V or 500V DC.
  • Leakage Measurements: Direct, differential, and alternative methods with a range of 4µA to 9999µA.
  • Power Measurement: VA rating method with a range of 0.1KVA to 4KVA.
Applications: Suitable for routine testing of medical electrical equipment, performance testing, asset management, and fast testing of IEC leads and earth bond testing.
Included Accessories: Comes with a calibration certificate, carrying case, earth bond test probe, patient applied lead, adaptors, mains cable, instruction manual, and application software.
Design Philosophy: The Rigel 62353 is designed to meet the demand for smaller, comprehensive test equipment in healthcare, combining portability with the capabilities of larger bench-mounted analysers.
Innovations: Features patented technology for earth continuity testing, overcoming contact resistance issues with a dual current high-intensity test.
See more

Catalog excerpts

Rigel 62353 Electrical Safety Analyser-2

Tried. Tested. Trusted. The Rigel 62353 is a hand-held medical electrical safety tester to combine an automatic / manual tester with a data logging / asset management facility. The Rigel 62353’s compact lightweight design and long life battery power reduces downtime between tests, making the instrument practical and highly portable for multi-site use. The Rigel 62353 tests to the new standard for hand-held medical safety testers has in-service and after repair testing of medical changed. The globally recognised IEC 62353 has become the general guideline for routine The Rigel 62353’s large internal...

 Open the catalog to page 2
Rigel 62353 Electrical Safety Analyser-3

Tried. Tested. Trusted. Trusted. Tried. Tested. Tried. Tested . Trusted. uide tical g A Pra A Prac 2353 ctical g uide to IEc 6 to IEc 62353 includes a unique feature to allow visual or Create customer test routines including test acceptance test routines to be configured protocols for testing patient monitors (NIBP, prior to an electrical safety test. These could be simple instructions to the user or protocols can be easily uploaded to the observations required for your own maintenance procedures such as checking Use the Rigel 62353 during functional for certain labels, software versions and...

 Open the catalog to page 3
Rigel 62353 Electrical Safety Analyser-4

Tried. Tested. Trusted. Rigel 62353 Design Philosophy domestic appliances. The Rigel 62353 has been designed to The Seaward Group has filed a patent address the increasing demand for smaller application for the new test technology that more comprehensive test equipment within uses a dual current high intensity test to the healthcare industry. What better way than overcome contact resistance problems or to combine such a tester with the other situations where weak conductor International In-Service Test required as per connections may inhibit protective earth testing with conventional test currents...

 Open the catalog to page 4
Rigel 62353 Electrical Safety Analyser-5

Tried. Tested. Trusted. SPECIFICATIONS Additional Information Unique use of Icons Earth Continuity Method 2 wire technique, using 'zero' lead function. Test Current Measuring Range (low range) 0.001 – 0.999 ohms @ 0.001 ohms resolution Measuring Range (mid range) 1.00 – 9.99 ohms /@ 0.01 ohms resolution Measuring Range (high range) 10.0 – 19.9 ohms @ 0.1 ohms resolution The Rigel 62353 features a hi-resolution graphic backlit display which not only provides highly visible and easy to follow menu structures but also allows the user to operate the tester using intuitive icons to speed up their...

 Open the catalog to page 5
Rigel 62353 Electrical Safety Analyser-6

Tried. Tested. Trusted Also available Power Measurement Mains Outlet Test Input voltage range: I EC Mains Lead Test Test Duration: Mains power Operating conditions Storage environment Environmental Protection Continuity of all conductors, Earth bond, Insulation & Polarity From Rigel Medical ■ Rigel 266 Plus Manual Safety Analyser ■ Rigel 277 Plus Automatic Safety Analyser ■ Rigel 288 Electrical Medical Safety ■ Rigel BP-SiM NIBP Simulator ■ Rigel SP-SiM SP02 Simulator ■ Rigel UNI-SiM Vital Signs Simulator ■ Rigel 333 Patient Simulator ■ Rigel 344 Defibrillator Tester ■ Rigel 355 Ventilator Tester...

 Open the catalog to page 6

All RIGEL Medical catalogs and technical brochures

  1. Med-eBase

    2  Pages

  2. SafeTest 60

    3  Pages

  3. CITREX H3

    1  Page

  4. SafeTest 99

    3  Pages

  5. CITREX H5

    2  Pages

  6. IEC 62353

    29  Pages

  7. IEC 60601

    18  Pages

  8. Global Brochure

    17  Pages

  9. CITREX

    6  Pages

  10. Rigel 62353+

    5  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.