SafeTest 60
1 / 3Pages

Catalog excerpts

SafeTest 60 - 1

SafeTest 60 An entry-level electrical safety analyzer for medical devices The Rigel SafeTest 60 is a robust, reliable and dedicated medical safety analyzer. It is ideal for testing high volumes of medical and laboratory equipment that does not require patient lead testing, such as beds, hoists, infusion pumps, CPAP’s and centrifuges etc. The SafeTest 60 includes a range of safety tests to enable compliance with a range of international safety standards, including leakage testing to IEC 60601, 62353, 61010 and NFPA-99, earth bond testing to 62353, 61010 and NFPA-99, and insulation testing to IEC 62353. > The SafeTest 60 is available in a wide range of power configurations to suit local and international markets > Compact, robust and portable design > With a large colour display and a colour-coded user interface, it’s easy to select the required tests with a single key press, whilst a fast step-through of the test routine makes the testing process speedy and dependable. > Manual control of fault conditions and tests > Accurate high current, low energy earth bond testing > Automatic warning of secondary earth/ground paths Passionate about patient safety. T: +44 (0) 191 587 8730 E: sales@rigelmedical.com See the SafeTest 60 for yourself, request your FREE live online demonstration today at rigelmedical.com/safetest6

Open the catalog to page 1
SafeTest 60 - 2

Technical Specification: Earth Continuity Pre-pulse 65-25A peak current, (0.1 to 0.8Ω respectively) Pulse Shape Exponential Decay Decay Time 200 – 550µs to 5% of peak (0.1 to 0.8Ω respectively) Method 2 Wire Measurement Current >± 200mADC into 2Ω Max Test Voltage 4-24Vrms o/c Measuring Range (Low Range) 0.001 – 0.999Ω Resolution 0.001Ω Measuring Range (Mid Range) 1.00 – 9.99Ω Resolution 0.01Ω Measuring Range (High Range) 10.0 – 19.9Ω Resolution 0.1Ω Accuracy ± 1% of value, ± 5mΩ ± 1% of value, ± 5mΩ Circuit Protection Test inhibited if ≥ 30VAC or DC at 4mm inputs > Detachable 2m mains lead...

Open the catalog to page 2
SafeTest 60 - 3

Technical Specification (Continued): The SafeTest 60 comes with a free 12-month warranty, subject to terms and conditions, available at Voltage Measurement Application L-N, L-E, N-E and touch voltage (IEC 61010) Range 0.0V – 300VAC Resolution 0.1V Accuracy ± 2% ± 5 digits (between 10V – 270VAC) Mains frequency 45.0 – 66.0Hz Resolution 0.1Hz Accuracy Unspecified EUT Load Current Measurement Range 0.0A – 20.0A Resolution 0.1A Accuracy ±5% ± 2 digits Power Source Maximum Current Rating Duty cycle (@21°C ambient) Mains Power General Specifications: Weight Size (L x W x D) Operating Temperature...

Open the catalog to page 3

All RIGEL Medical catalogs and technical brochures

  1. CITREX H3

    1 Pages

  2. SafeTest 99

    3 Pages

  3. CITREX H5

    2 Pages

  4. IEC 62353

    29 Pages

  5. IEC 60601

    18 Pages

  6. CITREX

    6 Pages

  7. Med-eBase

    2 Pages

  8. UniPulse 400

    2 Pages

  9. Rigel 62353+

    5 Pages

  10. Rigel 288+

    5 Pages

  11. PULS-R

    2 Pages

Archived catalogs

  1. PatSim200

    2 Pages

  2. Uni-Pulse

    6 Pages

  3. RIGEL

    20 Pages