Product catalog summary
Specifications and Adjustments:
The slit lamp tower should be moved 10 to 12 degrees away from the center. An I.R. heat-absorption filter should be selected, and the height of slit illumination adjusted. The magnification selector, joystick, and centering knob are used to achieve an artifact-free image with good color saturation. Slit beam width can be adjusted for wide peripheral illumination for fundus observation or narrow slit illumination for observing topographical changes such as optic nerve heads, tumors, detachments, and depressions.
Distance Measurements:
For 78D lenses, the distance to the lens edge is 110 mm and to the corneal apex is 128 mm. For 90D lenses, these distances are 109 mm and 124 mm, respectively.
Cleaning and Sterilization Procedures:
Clean lenses with alcohol. Disinfect by placing the lens in 2% glutaraldehyde for 25 minutes or in a diluted sodium hypochlorite solution (10:1 ratio) for 25 minutes. Sterilize using ethylene oxide gas and aerate at 50°C to 60°C. Do not use autoclave, boiling, or plasma sterilization methods.
Lens Specifications:
The Righton 78D and 90D lenses are designed for use with slit lamps, providing hassle-free fundus and vitreous imaging. The lenses feature an aplanat design for even illumination and low aberration, multilayer coatings to reduce reflections, and easy-to-hold frames. The 14D and 20D lenses offer excellent working distances and wider fields of view.
Warnings and Recommendations:
Ensure correct usage by reading all manuals carefully. Specifications and equipment are subject to change without notice. The height of the slit beam should not exceed the patient's pupillary dilation.
Contact Information:
RIGHT MFG. CO., LTD. and TOHOKU RIGHT MFG. CO., LTD. provide contact details for further inquiries.
See more

Catalog excerpts

Aspherical Lens-1

Righton 78D and 90D lens holders are attachable for Righton slit lamps, making possible hasslefree fundus and vitreous imaging. Aspherical Lens For observation, simply adjust the slit lamp position and set the lens to infinity. Available with Righton slit lamps. 78D and 90D lenses for use with slit lamps Guide The height of the slit beam should not exceed the patient's pupillary dilation. Slit lamp tower Move tower 10 to 12 degrees away from center Corneal apex 128 mm 124 mm Wide peripheral illumination for fundus observation Select I.R. heat-absorption filter Adjust height of slit illumination Narrow slit illumination for observation of topographical changes (optic nerve heads, tumors, detachments, depressions, etc) Magnification selector Centering knob Decenter and rotate the tower until an artifact-free image with good color saturation is obtained. Joystick Slit beam width adjustment Cleaning: Clean the lens with alcohol. Caution! How to sterilize lenses Disinfecting: Place the lens in 2% glutaraldehyde for 25 minutes. Alternatively, dilute sodium hypochlorite (household bleach) in water at a ratio of 10:1 and place the lens in the bleach solution for 25 minutes. Sterilizing: Sterilize the lens with ethylene oxide gas and allow it to aerate at a temperature of 50°C to 60°C. Warning: Do not use autoclave sterilization, boiling sterilization or plasma sterilization. To ensure correct usage, read all manuals carefully before using equipment Specifications and equipment are subject to change without any notice or obligation on the part of the manufacturer. The information in this brochure is correct as of August 2013. Ophthalmic Sales Ophthalmic Service 1-47-3, Maeno-cho, Itabashi-ku, Tokyo 174-8633, Japan Tel: +81-3-3960-2275 Fax: +81-3-3960-2285 e-mail: [email protected] 45-1, Aza-yashikimae, Nakamura Osato-cho, Kurokawa-gun, Miyagi 981-3521, Japan Tel: +81-22-359-3113 Fax: +81-22-359-3213

 Open the catalog to page 1
Aspherical Lens-2

For larger image observation Effective diameter Extremely comfortable to use Fit snugly into the hand Following a tradition of Japanese-made premium lenses • The aplanat design ensures images with even illumination and low aberration. Focal length Center thickness Working distance Best standard for indirect ophthalmoscopy Effective diameter Center thickness Working distance Focal length • Clear and bright images • Wider effective diameters • Multilayer coating on lenses greatly reduces reflections • Undistorted peripheral visual field Best use for slit lamp Effective diameter Center thickness...

 Open the catalog to page 2

All Righton catalogs and technical brochures

  1. Speedy-K2

    2  Pages

  2. SPEEDY-i

    1  Page

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.