1. Catalogs
  2. Rushwood
  3. Instructions for use

Instructions for use

Instructions for use

Product catalog summary
Overview: The eCrush™ is an electronic pill crusher designed for ease of use, reliability, and cleanliness. It is capable of performing approximately 100,000 crushes and is intended solely for medication crushing.
Package Contents: The package includes the eCrush™ device, an AC/DC adaptor with power cord, a rechargeable Li-Ion battery, a screwdriver, a cup holder, and 100 disposable crushing cups.
Setup and First Use: Ensure the device is undamaged, install the battery, and connect to a power source. The device can operate on battery or direct power. Use two disposable cups per crush, placing medication between them.
Operation: The device is activated by pressing the power button, and crushing is initiated with a separate button. The process takes approximately 22-23 seconds. LED indicators provide status updates, including power and battery levels.
Cleaning Instructions: Disconnect the power before cleaning. Wipe the device with a dry or slightly damp cloth. The cup holder can be washed with water.
Intended Use and Safety: The eCrush™ is for medication crushing only. Misuse voids the warranty. Keep away from children and use in a clean, well-lit environment.
Service and Maintenance: Repairs should be conducted by authorized personnel. The device comes with a 2-year limited warranty, excluding misuse and unauthorized repairs.
Technical Specifications: The device operates on a 100-240V power supply, with a Li-ion battery capacity of 14.8V4400mAh. It has a crushing cycle time of 22 seconds and can handle up to 4g of pills per crush.
Environmental Conditions: Operate within 5 to 40°C, below 2000 meters altitude, and in less than 80% humidity.
Battery and Adaptor Guidelines: Use only the provided battery and adaptor. The battery supports 500 charging cycles, and its performance may decrease thereafter.
See more

Catalog excerpts

Instructions for use-1

Range of conditions of use: • Temperature range: 5 to 40 0C • Altitude: less then 2000 meter • Relative humidity: less then 80% for temperatures below 31 0C, decrease linearly to 50% at 40 0C. • Only use the adaptor provided with the unit in the packaging. • The manufacturer/ distibutor does not hold any legal responsibility incurred by any damage or accidents caused by other adaptors. • Only purchase replacement AC/DC adaptor from our authorised distributors. • Do not use the AC/DC adaptor if any damage is observed. • Do not use the machine, Li-Ion battery, AC/DC adaptor or plug with wet hands. • If the product is used in a manner not specified by Rushwood B.V., the protection provided by the machine may be impaired. • Position the AC/DC adaptor in such a way that it is easy to disconnect from the socket outlet. • Only use the Li-Ion battery provided with the unit in the packaging. • The manufacturer/ distibutor does not hold any legal responsibility incurred by any damage or accidents caused by other batteries. • Only purchase replacement rechargeable Li-Ion battery from our authorised distributors. • Do not use the Li-Ion battery if any damage is observed. • Disconnect and remove the Li-Ion battery from the battery compartment if you only want to operate machine by the AC/DC Adaptor. • The rechargeable Li-Ion battery MUST be disconnected during transport. • Rechargeable Li-Ion Battery guarantees 500 charging cycles with optimized performance. • Charging capacity will decrease after that. Battery Level Display Indications: Disconnect the power source. Wipe unit's surfaces with a dry or slightly wet cloth. Use dry paper or cloth for a second cleaning. Cupholder can be washed with water and wiped with a dry cloth. It is strongly advised to clean the unit after use to prevent contamination, powder getting inside the mechanical assembly, that may lead to malfunction. eCrush™ is developed for the crushing of medication only. Do not use this device and its accessories for other purposes. In case the device or its accessories is used for other purposes, the warrenty will be voided. Always consult your physician first before crushing medication. Not all medication is allowed to be crushed. Keep the eCrush™ away from children. Stop using the unit if any susceptible damage is observed. Consult your physician first before crushing of medication with eCrush™. Repairs should always be performed by qualified technicians or our official representatives. Always place the unit on a flat, steady surface. Do not open the lid during crushing. eCrush™ should be used in room temperature and in a clean environment. Use the eCrush™ in a well-lit room to avoid accidents. The crushing cups are disposable and can be used once. For repairs and replacements of parts, the eCrush™ shall be returned to Rushwood B.V. or your regional distributor. After repair of replacement, always check the unit with a trial use. If you have any uncertainties, please do not use the unit and contact your distributor for assistence. Standard 2-year limited warranty. Under regular use and conditions, the manufacturer warrants this product to be free from defects in quality and materials for two (2) years from the original invoice date. If the disposable crushing cups are re- used or cups of any other brand are used, the warrenty will be voided. Shipping and handling fees are to be paid for by the customer. The manufacturer agrees, at its option during the warranty period, to repair any defect in material or to furnish a rebuilt or refurbished product of equal value in exchange without charge. The customer will incur the fee for shipping, handling, packing, return postage and insurance. Such repair is subject to verification of the defect or malfunction and proof of purchase as confirmed by showing the model number on the original dated sales receipt. Misuse or the device has been used for other purposes Defects and malfunction caused by unauthorized repair or modification. Operation guidelines/ User manual has not been followed. Damage and malfunction caused by accidents, misuse, negligence, careless handling, abnormal voltage supply, power interruptions or other events beyond companies control. Normal wear and tear. Not applicable to accessories and spare parts. Power Supply

 Open the catalog to page 1

All Rushwood catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.