1. Catalogs
  2. Sakura Finetek Europe
  3. Imagine... simple with faster performance TDE 30 SAKUPA Decalcifier System
video corpo

Imagine... simple with faster performance TDE 30 SAKUPA Decalcifier System

Imagine... simple with faster performance TDE 30 SAKUPA Decalcifier System

Imagine... simple with faster performance TDE 30 SAKUPA Decalcifier System

Product catalog summary
Overview: The document describes the SAKURA TDE™ 30 Decalcifier System, a device designed to decalcify bones using an electrolytic principle combined with the TDE™ 30 Reagent. This system significantly reduces the time required for decalcification, making it three times faster than traditional methods.
Specifications:
  • Product: TDE™ 30 Decalcifier
  • Power: 220V, 50Hz
  • Capacity: Up to 30 cassettes
  • Dimensions: Bench System - 30 x 21 x 11 cm; Solution Reservoir - 16 x 12.5 x 6 cm
  • Weight: 3 kg
  • Solution Volume: Up to 750 mL
Procedure:
  1. Pour approximately 750 mL of the SAKURA TDE™ 30 Decalcifier Solution into the reservoir.
  2. Place tissue cassettes in the tray, close the lid, and insert electrodes close to the cassettes.
  3. Connect the unit to the power supply and switch it on.
  4. Ensure the procedure is conducted under a fume hood due to irritating vapors.
  5. Wait approximately 15 minutes for biopsies to float, then verify decalcification.
  6. Rinse biopsies with water and place them in a fixative for further treatment.
Recommendations:
  • Use fixatives such as Formaldehyde, Bouin, or Dubosq Brazil.
  • Place electrodes as close as possible to the bone.
  • Decalcification duration varies from 30 minutes to 3 days depending on bone size and mineralization.
  • Change the decalcifier solution if the process exceeds 12 hours.
Warnings:
  • Conduct the procedure in a corrosive-resistant fume hood.
  • Avoid placing metal objects in the tray.
Company Information: Sakura Finetek Europe B.V. is committed to improving laboratory efficiency by providing reliable products for histopathology. Contact details include phone, fax, and email, with further information available on their website.
See more

Catalog excerpts

Imagine... simple with faster performance TDE 30 SAKUPA Decalcifier System-1

Imagine... simple with faster performance I Decalcifier System

 Open the catalog to page 1
Imagine... simple with faster performance TDE 30 SAKUPA Decalcifier System-2

The SAKURA TDE™ 30 Decalcifier System is based on a simple electrolytic principle that, in combination with the TDE™ 30 Reagent, decalcifies bones. After decalcification by the SAKURA TDE™ 30, bones can be sectioned more easily. By using this technique the duration of the decalcification process is reduced by 60%, which is three times faster than classical techniques (for example, a femur bone section can be processed in just 15 minutes). 1. our approximately 750 mL of the SAKURA P TDE™ 30 Decalcifier Solution (Item Code 1428) in the solution reservoir. 2. lace the cassettes with the tissue in...

 Open the catalog to page 2
Imagine... simple with faster performance TDE 30 SAKUPA Decalcifier System-3

5. ait approximately 15 minutes W for the biopsies to float. If the biopsies do not float verify the decalcification with a needle or a knife or use an identification method. 6. inse the biopsies with running water and R place them in the fixative for further exterior treatment. Decalcification of Bone We recommend to use the following fixatives: ● Formaldehyde ● Bouin ● Dubosq Brazil The method for fixation is identical to the one used for the biopsies. ● Place the electrodes as close as possible to the bone ● Depending on the size of the bone and mineralisation, the duration of the decalcification...

 Open the catalog to page 3
Imagine... simple with faster performance TDE 30 SAKUPA Decalcifier System-4

Dedicated to Histopathology Sakura Finetek improves the laboratory by offering the most reliable products in the market. The histotechnologists can conveniently complete their activities in a safe environment. As the market leader in histopathology, Sakura Finetek is continuously looking for possibilities to improve the laboratory and succeeds in offering solutions for the problems found in the histopathology laboratory. Product Specifications Capacity Dimensions W x D x H Bench Weight Solution Volume Up to 30 cassettes System 30 x 21 x 11 cm Solution Reservoir 16 x 12.5 x 6 cm 3 kg Up to 750mL...

 Open the catalog to page 4
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.