Common Conditions:
1. Solutions used: Water and Methanol in a 7:3 ratio.
2. Initial pulse input before testing: 40 pulses.
1. Reproducibility/ TS-MP18SPAC2020:
Volume: 5.08 uL
Additional pulse input: 400 pulses
Plunger stroke: 2 mm
Intake and dispensing speed: 100 pps
PTFE Tubing: 0.3 mm ID, 10 mm L, 0.7065 uL
Results: The mean volume (M.V.) ranged from 5.0 to 5.119 uL across different pumps with a coefficient of variation (C.V.) between 0.08% and 0.22%, indicating high reproducibility.
2. Reproducibility/ TS-MP55SPAC2020:
Volume: 47.48 uL
Additional pulse input: 400 pulses
Plunger stroke: 2 mm
Intake and dispensing speed: 200 pps
PTFE Tubing: 0.6 mm ID, 10 mm L
Results: M.V. ranged from 47.316 to 47.725 uL with C.V. between 0.025% and 0.150%, showing excellent accuracy and reproducibility.
3. Reproducibility/ TS-MP64SPAC2020:
Volume: 64.32 uL
Additional pulse input: 400 pulses
Plunger stroke: 2 mm
Intake and dispensing speed: 200 pps
PTFE Tubing: 0.6 mm ID, 10 mm L, 2.826 uL
Results: M.V. ranged from 63.400 to 64.700 uL with C.V. between 0.036% and 0.083%, confirming high reproducibility.
4. Reproducibility/ TS-MP78SPAC2020:
Volume: 95.56 uL
Additional pulse input: 200 pulses
Plunger stroke: 1 mm
Intake and dispensing speed: 200 pps
PTFE Tubing: 0.6 mm ID, 10 mm L, 2.826 uL
Results: M.V. ranged from 95.085 to 95.663 uL with C.V. between 0.010% and 0.045%, indicating unparalleled accuracy and reproducibility.
5. Dispensing Speed Data/ TS-MP78SPAC2020P0.5:
Result: 300 uL dispensing achieved within 0.5 seconds using a unipolar motor with specific wiring and driver settings. The motor's maximum current should be kept at 70% or lower of the rated current.
Contact Information:
SANWA TSUSHO CO., LTD.
Nishi-Gotanda 3-13-2 Shinagawa-ku Tokyo 141-0031 Japan
Tel: +81-3-3492-6300
Fax: +81-3-3492-6311
URL: www.sanwatsusho-global.com
Email: [email protected]
See more

Catalog excerpts

ts_series_reproducibility_data.pdf-1

Unparalleled Accuracy and Reproducibility The following data are taken under the conditions shown below: Common Conditions: 1) Solutions: Water : Methanol/ 7 : 3 2) Pulse input prior to the test: 40 pulses 1. Reproducibility/ TS-MP18SPAC2020: Volume: Additional pulse input: Plunger stroke: Intake speed: Dispensing speed: PTFE Tubing: 2. Reproducibility/ TS-MP55SPAC2020: Volume: Additional pulse input: Plunger stroke: Intake speed: Dispensing speed: PTFE Tubing: 3. Reproducibility/ TS-MP64SPAC2020: Volume: Additional pulse input: Plunger stroke: Intake speed: Dispensing speed: PTFE Tubing:

 Open the catalog to page 1
ts_series_reproducibility_data.pdf-2

Unparalleled Accuracy and Reproducibility 4. Reproducibility/ TS-MP78SPAC2020: Volume: Additional pulse input: Plunger stroke: Intake speed: Dispensing speed: PTFE Tubing: 5. Dispensing speed data/ TS-MP78SPAC2020P0.5: Result: Driver: Motor wiring: Thread pitch: Applied pressure: Micro step: Initial speed: Acceleration: Speed: 16,000 pps 300 uL dispensing has been effected within 0.5 second (Linear travel: 6.278 mm/ 0.5 second) TS-2A-SDB-220B at 0.8 A Wiring No.1* 0.5 mm 1 MPa 1/2 20 pps 100,000 pps/ second *Wiring No.1 with a unipolar motor Stepping motor A phase coil A phase coil B phase coil...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.