Prerequisites: Before starting the procedure, ensure the tear is identified and the implant packaging is fully opened. Use the meniscus rasp to create a bleeding surface and the MENIX® probe to determine the insertion depth.
Insertion Procedure: Insert the MENIX® portal skid through the portal and position it under the condyle of the compartment with the tear. Advance the MENIX® Duo system along the portal skid and remove the skid. The device can be bent up to 35° if necessary.
Anchor Insertion: For the first anchor, retract the button, push it slightly to expose the pin tip, and select the entry point. Advance to the desired depth, noting the marks at 17 mm and 22 mm from the pin tip. For the second anchor, ensure the entry point is 5 mm away from the first, and follow a similar procedure.
Tensioning and Evaluation: Tension the repair by alternately pulling the sutures and apply counterpressure with the probe. Cut the suture tails using a vertical insertion of the suture cutter. Evaluate the repair and use additional anchors if necessary.
MENIX® Duo Characteristics: The system includes options with one or two anchors and various suture tapes. Additional non-absorbable sutures are recommended for repair.
Instrumentation: The complete set includes a suture cutter, portal skid, knot pusher, arthroscopic probe, twister, rasp, and scissors. These are reusable and classified as class I medical devices.
Manufacturer Information: The MENIX® and MENIX® Duo are class IIb medical devices manufactured by S.B.M. SAS, located in Lourdes, France. The HS FIBER sutures are also class IIb, manufactured by RIVERPOINT and distributed by S.B.M. SAS.
See more

Catalog excerpts

Surgical Technique - Meniscus Repair - MENIX® DUO-2

SURGICAL TECHNIQUE Prerequisites Once you have identified the tear, open the implant packaging in its entirety Follow the opening direction. Note: Locate chevron seal on packaging. Evaluate the insertion depth Using the provided meniscus rasp, create a bleeding surface and use the MENIX® probe to determine the desired insertion depth. Insert the MENIX® portal skid Insert the MENIX® Duo system Advance and insert the MENIX® portal skid through the portal and place it under the condyle of the compartment with the meniscus tear that needs to be repaired. Advance the MENIX® Duo meniscal suture system...

 Open the catalog to page 2
Surgical Technique - Meniscus Repair - MENIX® DUO-3

Insert the first MENIX® Duo anchor into the joint Slightly push the first button in order to show the tip of the pin out of the cannula and select your entry point. Note: The first mark is 17 mm from the tip of the pin. The second mark is 22 mm from the tip of the pin. Retract the button to its initial position without pulling on the handle. Note: Leave the device inside the joint. MENIX®Duo Surgical Technique | 3

 Open the catalog to page 3
Surgical Technique - Meniscus Repair - MENIX® DUO-4

SURGICAL TECHNIQUE Insert the second MENIX® Duo anchor Slightly push the second button in order to show the pyramidal tip of the pin out of the cannula and select your entry point. Advance the second button to the desired depth. Note: Entry point should be 5mm away from the 1st one. Retract the second button to its initial position without pulling on the handle, then remove the MENIX® Duo device from the joint. Note: You can block your finger on the button opposite to the one in use to keep the pin in the cannula. 4 | MENIX®Duo Surgical Technique

 Open the catalog to page 4
Surgical Technique - Meniscus Repair - MENIX® DUO-5

Tension the repair Alternately pull on the two sutures to tension the repair and control the system reduction. Use the probe to apply counterpressure on the meniscus when reducing the tear and to control suture tension. Evaluate the Meniscus repair Advance the suture cutter into the joint and cut the suture tails. Make sure the fixation is adequate ; if necessary complete the repair with another MENIX® anchor. Note: Prefer vertical insertion of the suture cutter. MENIX®Duo Surgical Technique | 5

 Open the catalog to page 5
Surgical Technique - Meniscus Repair - MENIX® DUO-6

PORTFOLIO MENIX® Menlscal suture system with anchorsReference Description MEN0201902 MENIX® Meniscal suture system with one anchor MEN0201901 MENIX® Duo Meniscal suture system with two anchors Non-absorbable sutures HS FIBER* Reference Description HS286 Passing Loop Suture Tape - WHITE/BLUE - 1.4 mm - Non-needled - L 60 cm (3 cm loop Size) HS287 Passing Loop Suture Tape - BLUE - 1.4 mm - Non-needled - L 60 cm (3 cm loop Size) *Additional non-absorbable sutures recommended for meniscus repair. 6 | MENIX® Duo Surgical Technique

 Open the catalog to page 6
Surgical Technique - Meniscus Repair - MENIX® DUO-7

MENIX® Complete instrumentation set References MENIX® Portal skid MENIX® Suture cutter MENIX® 90 degree arthroscopic probe MENIX® Knot pusher MENIX® Twister MENIX® Arthroscopic scissors MENIX® Rasp Reusable instrumentation Carefully read the instructions on the package leaflet that accompanies the medical device or on the label given to the healthcare provider. MENIX® and MENIX® Duo: class IIb medical device. MENIX® Instrumentation: class I medical devices. Manufacturer: S.B.M. SAS HS FIBER sutures: class IIb medical device. Manufacturer: RIVERPOINT - Distributor: S.B.M. SAS Ref.: MGMENTOEN02_0426...