The RI 2000 series of Differential Refractive Index Detectors are designed for high sensitivity, stability, and reproducibility in refractive index detection. Key features include a thermally isolated optic with a counter-current heat exchanger and programmable temperature control, ensuring a stable baseline and optimal signal-to-noise ratio. The detectors also offer autopurge, autozero capabilities, and RS232 communication for direct data acquisition.

Technical Specifications
  • Detection Method: Deflection
  • Refractive Index Range: 1.00 to 1.75
  • Flow Rate: RI 2000M and RI 2000A: 0.2 – 3.0 ml/min; RI 2000P: 5 – 50 ml/min
  • Flow Cell Volume: RI 2000M: 4 μl, 45° angle; RI 2000A: 9 μl, 45° angle; RI 2000P: 13 μl, 5° angle
  • Flow Cell Pressure: 6 kg/cm²
  • Linearity Range: RI 2000M: 0 - 500 μRIU; RI 2000A: 0 - 1000 μRIU; RI 2000P: 0 - 20000 μRIU
  • Noise Level: RI 2000M and RI 2000A: 5 x 10-9 RIU; RI 2000P: 5 x 10-8 RIU
  • Autozero Range: Full Range
  • Drift with 1ml H2O/min: < 1mV/hour for RI 2000M and RI 2000A; 1mV/hour for RI 2000P
  • Purge Valve: Yes for RI 2000M and RI 2000A; Yes/No for RI 2000P depending on flow

Output and Control
  • Integrator Output: +/- 1 Volt
  • Recorder Output: +/- 10mV / 100mV / 1Volt
  • Recorder Offset: 0 mV/10mV/ 100mV
  • Recorder Range: 8 steps (1:8) - (16:1)
  • Marker: Yes/No

Order Information
  • RI2000, M: Refractive index detector RI 2000M, micro
  • RI2000, A: Refractive index detector RI 2000A, analytical
  • RI2000, P: Refractive index detector RI 2000P, semi preparative

Spare Parts
  • RI2000-030: Flow cell, RI 2000, micro
  • RI2000-001: Flow cell, RI 2000, analytical
  • RI2000-031: Flow cell, RI 2000, semi preparative
  • RI2000-012: Lamp RI 2000
  • RI2000-024: Seal kit for flow cell RI 2000
  • RI2000-006: Purge valve RI 2000
See more

Catalog excerpts

RI 2000 Refractive index detector-1

RI 2000 Refractive index detector The Differential Refractive Index Detector RI 2000 series offers the sensitivity, stability and reproducebility required for optimal RI detection. The thermal isolated optic with a counter current heat exchanger and with its programmable temperature control, results in an extremely stable baseline and an optimal Signal / Noise ratio. The Refractive Index Detector RI 2000 provides autopurge and autozero capabilities, as well as RS232 communication to acquire data directly without using any external signal interface. Technical Specifications* RI 2000M, micro RI 2000A, analytical Detection Method: Refractive Index Range: Flow Rate: Flow Cell Volume Flow Cell Pressure Dead Volume Into cell Linearity Range: Noise Level: Autozero Range: Full Range Full Range Full Range Purge Valve ____________________________________________________________________________________ Integrator Output: Recorder Output: Recorder Offset: Recorder Range:

 Open the catalog to page 1
RI 2000 Refractive index detector-2

RI 2000 Refractive index detector Order Information Refractive index detector RI 2000 Part-No.: Refractive index detector RI 2000M, micro Refractive index detector RI 2000A, analytical Refractive index detector RI 2000P, semi preparativ Order Information Spare Parts Refractive index detector RI 2000 Part-No.: Flow cell, RI 2000, micro Flow cell, RI 2000, analytical Flow cell, RI 2000, semi preparativ Seal kit for flow cell RI 2000

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.