Product catalog summary
Introduction to STATCARE™
STATCARE is a pre-sterilization handpiece maintenance spray designed to lubricate and protect dental handpieces, including turbines, air motors, shanks, and heads, to extend their lifespan. It is compatible with STATIS handpieces and should be used before every sterilization cycle. The spray is applied by holding the can vertically and spraying for one second, followed by running the handpiece for 20 seconds to remove excess oil. Handpieces are then ready for sterilization in a STATIM Cassette Autoclave®.
STATCARE Nozzles
Three reusable metal spray-cap nozzles are available: E-Type Nozzle (Order No. S500E), Multi Nozzle (Order No. S500M), and Tip Nozzle (Order No. S500T). These nozzles ensure precise application of the spray to necessary areas.
High-Speed Handpiece Maintenance
After use, remove the bur or diamond, clean the handpiece externally with a soft brush under cold water, and dry it. Use the Multi spray nozzle #S500M for lubrication. If debris is present, repeat lubrication until clean oil is emitted. Operate the handpiece for 20 seconds to expel excess lubricant before sterilization.
Chuck Maintenance
After cleaning and lubricating the handpiece, lubricate the chuck using the Tip nozzle #S500T.
Low-Speed Handpiece Maintenance
Remove attachments and clean shanks under cold water. Use the E-type nozzle #S500E to spray lubricant into the attachment's back end. Operate the handpiece for 20 seconds to expel excess lubricant before sterilization.
Air Motor Maintenance
Spray lubricant into the drive airline using the Tip nozzle #S500T. Cover the motor's front end to catch debris. Reassemble and operate for 20 seconds to expel excess lubricant before sterilization.
Contact Information
SciCan Ltd., Toronto, Ontario, and SciCan Inc., Canonsburg, PA, provide contact details for further assistance.
See more

Catalog excerpts

STATCARE brochure-1

One step, pre-sterilization handpiece maintenance spray VaPorisateur pre-sterilisation en uneeif Pour I'entretien de piece amain - Provides lubrication to help prolong the* your dental handpieces. bricate prior to every sterilization cycle. 9 the can in a vertical position, spray for one Ml m a vertical position, spray Tor w lubrication to expel excess oil. For c°mpiete maintenance instructions see the han** j manufacture's instruction or maintenance g"1 .Le "aporisateur procure une lubrification qui a* Prolonger la duree de vos pieces a main <**• Lubrifier avant chaque cycle de sterilisation- j ^erisateur en position verticale, *^J$St?\ *ndant une seconde a chaque processus oe : Conner la piece a main pendant 20 e***^ *i6ceD ",«"«uren position veru«~— ■ ^ref„ nt une seconde a chaque processus *fet l'0nner la P^ce a main pendant 20 second Pendant 20 secondes apres I'huile excedentai r°S^,nirles instructions dentretien completes , nS,ructi°nsdu manufactunerou au guided STATCA^ATCARE is 3 <rademari< °* ^fwe^^ Ma<ie in Germany - Fabriqu* en AnemM" Introducing STATCARE From the creators of STAT/M® handpiece maintenance spray STATCARE provides lubrication to help increase the life of dental handpieces. It cleans and protects turbines, air motors, shanks and heads to ensure long lasting service STATCARE can be used as directed by handpiece manufacturer's instruction or mainenance guide FOR STATIS handpieces, lubrication should be done prior to every sterilization cycle SIMPLY HOLD the can in a vertical position and spray for one full second during every lubrication process. Run the handpiece for 20 seconds after lubrication to expel excess oil HANDPIECES are then ready for sterilization in a STATIM Cassette Autoclave®; for fast, gentle sterilization that helps prolong handpiece life Order No. S500-6NA (case of 6 cans) STATCARE NOZZLES Three specifically designed reusable, metal spray-cap nozzles deliver just the right amount of STATCARE spray to the exact areas where E-Type Nozzle Multi Nozzle* STATIM and STATIM Cassette Autoclave are registered trademarks and STATCARE and STATIS are trademarks of SciCan Ltd. Tits KaVo MULTIflex coupling

 Open the catalog to page 1
STATCARE brochure-2

STATIS™ handpiece maintenance guide Lubricate prior to every sterilization cycle. High speed handpieces Preparation: After use of the high speed handpiece, remove the bur or diamond and clean the handpiece externally, using a soft bristle brush under cold tap water. Dry handpiece with gauze or paper towel. Lubricate and operate: To lubricate and clean turbines use Multi spray nozzle #S500M as illustrated (also suitable for low speed air motors equipped with a Multiex® coupling). Hold a tissue, paper towel or an absorbent pad over the head of the instrument to catch dispelled debris. Then inspect...

 Open the catalog to page 2

Archived catalogs

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.