For QTRAP® Systems System Integration Test and Data Log
Open the catalog to page 1This document is provided to customers who have purchased SCIEX equipment to use in the operation of such SCIEX equipment. This document is copyright protected and any reproduction of this document or any part of this document is strictly prohibited, except as SCIEX may authorize in writing. Software that may be described in this document is furnished under a license agreement. It is against the law to copy, modify, or distribute the software on any medium, except as specifically allowed in the license agreement. Furthermore, the license agreement may prohibit the software from being disassembled,...
Open the catalog to page 2System Integration Test and Data Log RUO-IDV-05-8311-A
Open the catalog to page 3System Integration Test and Data Log RUO-IDV-05-8311-A
Open the catalog to page 4This guide describes the steps for preparing for and performing an LC-MS/MS system integration test for a NanoLC ® System combined with the SCIEX 4000, 4500, 5500/5500+, or 6500 QTRAP mass spectrometer. The system might ® include the cHiPLC System or an external column and one of the following NanoLC Systems: TM • NanoLC Ultra System TM • Tempo System Note: Read the safety instructions in the Operator Guide prior to running the system integration test. ® Note: For information about assembly and operation of the NanoSpray or OptiFlow Turbo V Ion Source, refer ® to the Ion Source Operator Guide....
Open the catalog to page 5Note: The test in this document is written with Gradient 1 as the low-flow channel. If this is not true for the system being tested (for example, a 2D system), then make the appropriate changes throughout the test. Refer to Table 1-1. Table 1-1 Channel Flow Rate NanoLC™ Systems System Integration Test and Data Log
Open the catalog to page 6Table 1-3 Required Materials for a NanoLC System with a cHiPLC® System This column only available for purchase by users from Phenomenex. The equivalent SCIEX part is only available to Field Service Employees (FSEs).
Open the catalog to page 7Create the autosampler and LC methods in the Eksigent Control Software and the acquisition method in the Analyst Software. 1. Plumb the autosampler valve with a 10 pL sample loop. 2. Open the Analyst® Software. 3. In the Autosampler control window, click Method Editor. 4. Create the autosampler method for a trap-elute configuration. Refer to the appropriate appendix for the system. • Methods for NanoLC™ 400 Systems with a cHiPLC® System • Methods for NanoLC™ 400 Systems with an External Trap and Column • Methods for NanoLC Ultra® Systems with a cHiPLC System • Methods for NanoLC Ultra® Systems...
Open the catalog to page 8LC-MS System Integration Test Table 2-1 Mobile Phase Mixtures In this method, the loading pump is the channel with the microflow module. 1. Open the Analyst® Software. 2. In the Acquisition window that opens (from the Eksigent Control Software), click the arrow in the top, right corner of the window to select the high-flow or loading pump. 4. Create the loading pump method. Refer to the appropriate appendix for the system: • Methods for NanoLC™ 400 Systems with a cHiPLC® System • Methods for NanoLC™ 400 Systems with an External Trap and Column • Methods for NanoLC Ultra® Systems with a cHiPLC®...
Open the catalog to page 9LC-MS System Integration Test • Methods for NanoLC™ 400 Systems with a cHiPLC® System • Methods for NanoLC™ 400 Systems with an External Trap and Column • Methods for NanoLC Ultra® Systems with a cHiPLC® System • Methods for NanoLC Ultra® Systems with an External Trap and Column • Methods for NanoLC™ and Tempo™ Systems with a cHiPLC® System • Methods for NanoLC™ and Tempo™ Systems with an External Trap and Column 4. Type the method name as specified in the appropriate appendix. 1. Close the Eksigent Control Software, if it is open. 2. Verify that the Analyst® Software drivers are installed. Refer...
Open the catalog to page 10LC-MS System Integration Test Table 2-2 Acquisition Method Parameters—Analyst® Software (continued) System Integration Test and Data Log NanoLC™ Systems
Open the catalog to page 11Table 2-3 MRM Transitions for Beta-galactosidase (continued) 7. Save the Analyst® software method (for example, System Integration Test). 2. Click the autosampler (for example, Eksigent AS2 or AS3) and then select the appropriate autosampler method. Refer to Create the Autosampler Method.
Open the catalog to page 12LC-MS System Integration Test 3. If Eksigent Gradient 2 appears in the device list, right-click Eksigent Gradient 2 and then clear the "Use" selection as it is not being used. Figure 2-2 Software Application Properties Tab—Autosampler Method 4. Click Eksigent Gradient 1 and then select the appropriate gradient pump method. Refer to Create the Gradient Method. 5. Click Eksigent Loading Pump and then select the appropriate loading pump method. Refer to Create the Loading Pump Method. 6. Save the acquisition method. Plumb the system in trap-elute configuration to perform the pre-column desalting...
Open the catalog to page 13WARNING! Toxic Chemical Hazard. Refer to the chemical product Safety Data Sheets and follow all of the recommended safety procedures when handling, storing, and disposing of chemicals. For health and safety precautions, refer to the Operator Guide. 1. Add 625.0 pL of buffer A (100% water:0.1% formic acid) to the beta-galactosidase vial. 2. Vortex the vial for at least 30 seconds. 3. Using a centrifuge, spin the vial to bring the liquid down to the bottom of the vial before opening. 4. Repeat step 2 to step 3 to confirm dissolution. 5. Aliquot the stock solution into 50 pL volumes and then store...
Open the catalog to page 14LC-MS System Integration Test This step produces a 1:5 dilution, with a concentration of 200 fmol/pL. 3. Transfer the solution to the autosampler vial, making sure there is no bubble on the bottom of the vial. Required Materials • Beta-galactosidase stock solution, prepared in Prepare the 1 pmol/pL Stock Solution of Beta-Galactosidase Digest Solution • Water with 0.1% formic acid (Buffer A) 1. Combine the specified amount of the 1 pmol/pL beta-galactosidase stock solution with Buffer A in clean vial. Refer to Table 2-4. Table 2-4 Sample Dilutions by Mass Spectrometer 2. Mix the solution for at...
Open the catalog to page 1565 Pages
12 Pages
12 Pages
12 Pages
12 Pages
6 Pages
6 Pages
6 Pages
6 Pages
7 Pages
6 Pages
16 Pages
2 Pages
12 Pages
8 Pages
26 Pages
6 Pages
8 Pages
8 Pages
8 Pages
13 Pages
32 Pages
8 Pages