Overview: The CytoBrite Family offers a high-capacity FISH hybridization system designed to efficiently process multiple batches of slides. It combines a Slide Incubation System and a Slide Oven to handle up to 72 FISH slides simultaneously, equivalent to six ThermoBrite instruments, but at less than half the cost and bench space.
Key Features:
  • Capacity: Can denature and hybridize 72 slides at once.
  • Cost Efficiency: Provides the same capacity as six ThermoBrite instruments at a reduced cost.
  • Space Efficiency: Uses 50% less bench space compared to ThermoBrite instruments.
  • Temperature Calibration: Built-in system eliminates the need for external calibration, reducing CLIA certification costs.
  • Slide Handling: Removable 6-slide trays facilitate probe addition, coverslipping, and transfers without the need for humidity strips or chambers.
Technical Specifications:
  • Slide Incubation System:
    • Electrical: 115/230V AC, 50/60Hz, 6A/3A
    • Dimensions: 8x10x15 inches (20x25x38 cm)
    • Weight: 20 lbs (9 kg)
    • Temperature Range: 15 to 90°C
    • Temperature Regulation: ± 0.2°C from set point
    • Heating/Cooling Rate: 37 to 75°C in < 2 minutes, 75 to 37°C in < 2 minutes
  • Slide Oven:
    • Electrical: 115/230V AC, 50/60Hz, 6A/3A
    • Dimensions: 22x18x17 inches (55x45x42 cm)
    • Weight: 100 lbs (45 kg)
    • Temperature Range: Ambient +5 to 90°C
    • Temperature Regulation: ± 0.5°C from set point
    • Slide-to-Slide Temperature Variation: < 0.5°C
    • Open Door Temperature Drop: 0.2°C @ 37°C after 30 seconds
Additional Information: The system is intended for research use only and includes user manuals for both the Slide Incubation System and the Slide Oven.
See more

Catalog excerpts

High Capacity FISH HybridizaƟon System-1

CYTOBRITE FAMILY Denatures and Hybridizes 72 Slides in Half the Cost and Bench Space of SIX ThermoBrite Instruments A single CytoBrite Slide Incubation System can perform the short denaturation step on multiple batches of up to 12 FISH slides with each batch then transferred to the CytoBrite Slide Oven for overnight incubation. The combined instruments hybridize up to 72 FISH slides simultaneously; the equivalent capacity of SIX ThermoBrite instruments at less than half the cost and bench space. Removable slide trays streamline work by holding slides for probe addition, coverslipping and transfers between instruments. No humidity strips or chambers are required since coverslips are sealed with CytoBond Removable Coverslip Sealant that doesn't dry out and removes easily and cleanly in one-piece post-hybridization. Built-in temperature calibration system eliminates the cost of factory or third party calibration required for CLIA compliance. Products Catalog # CytoBrite Slide Incubation System, 115/230V. 2019-00-1 CytoBrite Slide Oven, 115V. 2019-70-1 CytoBrite Slide Oven, 230V. 2019-70-2 Accessories Additional CytoBrite 6-slide tray. 2019-10-0 CytoBond Removable Coverslip Sealant, 100 ml. 2020-00-1 Digital Thermometer, handheld, with NIST certificate and cable. 1051-52-0 System Specifications Electrical Cat # 2019-00-1 Dimensions (HxWxD) Instrument Weight Performance Temperature range: Temperature regulation: Temperature accuracy: Heating rate: Cooling rate: Description 115/230V AC, 50/60Hz, 6A/3A 8x10x15 inches (20 x25x38 cm) 20 lbs (9 kg) net 15 to 90°C ± 0.2°C from set point ± 0.5°C 37 to 75°C in < 2 minutes 75 to 37°C in < 2 minutes High Capacity System: • Denatures and hybridizes 72 slides • Equivalent to 6 Thermobrites at under half the cost • Uses 50% less bench space than Thermobrites • Built-in temperature calibration eliminates CLIA certification cost • Removable 6-slide trays streamline slide set up and transfers • Slides incubated dry - No humidity strips or chambers needed Oven Specifications 520 CytoBrite Slide Incubation System User Manual 523 CytoBrite Slide Oven User Manual For Research Use Only Type Manual Manual After denaturation, transfer batches of slides to the CytoBrite Oven for overnight incubation. RoHS SciGene • 1287 Reamwood Avenue, Sunnyvale, CA 94089 USA • Tel 408-733-7337 • Fax 408-733-7336 • [email protected]

 Open the catalog to page 1
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.