IPV® 1 Quick Reference Guide-1

IPV® 1 Quick Reference Guide Control Knobs + Starting Positions Digital Display Gas Source knob turns the IPV® 1 on by selecting a connected gas source. Starting Position: OFF Pulse Amplitude displays a 5 second average of peak minus trough pressures sampled at the output of the Phasitron. Frequency knob determines the rate of high frequency percussive pulses delivered to the patient. Starting Position: Rotated fully to the left ean Airway Pressure (MAP) displays a M 5 second average of the pressure sampled at the output of the Phasitron. Session Timer displays total usage time of the current session. Pulse Frequency displays the number of pulses per minute. Amplitude knob determines the pressure delivered to the patient. Starting Position: Rotated fully to the right NOTE: For patients coming off respiratory support, consider starting treatment with some amplitude (e.g. the amplitude knob pointing straight up) then quickly titrate to target the MAP delivered by previous support settings

 Open the catalog to page 1
IPV® 1 Quick Reference Guide-2

Prepare the patient ✓ Place knobs in starting positions (page 1). ✓ Connect IPV® 1 to appropriate gas source. ✓ Assemble and connect Phasitron® to IPV® 1. + If needed, perform a Pre-Use Check (page 5). ✓ Position patient upright as tolerated or elevate head and shoulders using pillows. ✓ Assess patient's heart rate, respiratory rate, and breath sounds. + If using in-line, note current ventilator mode, settings, and alarms. ✓ Turn Gas Source knob to connected source. ✓ Connect Phasitron® to patient interface. + The Digital Display will not awaken until IPV® 1 detects back pressure from the Phasitron®....

 Open the catalog to page 2
IPV® 1 Quick Reference Guide-3

Delivering IPV® Therapy In-Line Refer to the IPV® 1 In-Line Valve IFU for a complete list of warnings and cautions. In-Line Valve Pressure Relief Adjustment Knob 22mm ID Pressure Relief Port *Indications for the use of IPV® therapy in patients under two years of age vary by geography and by device. WARNING: NEVER perform IPV® treatment without liquid in Phasitron® nebulizer during treatment. WARNING: Follow your institutional protocols before disconnecting the ventilator inspiratory limb prior to installation of IPV® In-Line Valve. Internal One-way Check Valve Cap WARNING: The In-Line Valve should...

 Open the catalog to page 3
IPV® 1 Quick Reference Guide-4

• Faster/higher frequencies provide secretion mobilization and quickly improve gas exchange. • Slower/lower frequencies loosen thicker secretions and help recruit the lung. • Aeration of pulses throughout lung fields and visible chest wiggle are primary indicators of effective treatment. • It is recommended to perform therapy for 15-20 minutes every 6-8 hours, or per provider orders. • A trained healthcare professional must monitor the patient throughout the therapy. • Both amplitude and frequency can be adjusted to increase or decrease the amount of chest wiggle. • The IPV In-Line Valve may...

 Open the catalog to page 4
IPV® 1 Quick Reference Guide-5

IPV® 1 Pre-Use Check WARNING: Complete the Pre-Use Check prior to initial use of the IPV® 1 and with each new Phasitron® breathing circuit. CAUTION: Do NOT use the device if any problems are noted. Contact a qualified service technician. NOTE: Displayed values are for illustrative purposes only. NOTE: Saline solution or other liquid is not required to perform a Pre-Use Check. Ensure the Gas Source knob is in the OFF position, before connecting the IPV® 1 to the gas source supply from the wall (AIR or O2). Connect the Unified Connector of the breathing circuit (A) to the connector port on the...

 Open the catalog to page 5
IPV® 1 Quick Reference Guide-6

Product Support Technical Support For technical inquiries, please contact your local Sales Representative. If you need further clarification, Sentec will gladly assist you via the following contact details: IPV® Product Support page Percussionaire 130 McGhee Road, Suite 109 Sandpoint, ID 83864 MMQ-248-V3 | 2026-03