Technical Parameters
The PE-3020VET is a fully automated hematology analyzer that uses electrical impedance and colorimetric methods to measure cell volume, number, and hemoglobin concentration. It requires a sample volume of ≤20μL and supports venous, capillary, and pre-diluted modes. The analyzer measures 21 parameters including WBC, RBC, HGB, and PLT, and provides three histograms for WBC, RBC, and PLT. It has a throughput of 60 samples per hour and supports CBC analysis mode with L-J/X-B QC mode. The device dimensions are 364mm x 498mm x 431mm, with a net weight of 23.5kg. It operates in temperatures between 10-30℃ and humidity ≤70%.
Reagents and Calibration
The analyzer uses diluent and lyse reagents and supports both auto and manual calibration. It can store up to 50,000 sample results with histograms.
Performance and Display
The device offers high precision with repeatability parameters for WBC, RBC, HGB, PLT, MCV, and HCT. It features a 10.4’’ IPS color touch screen with a resolution of 800x600.
Species Compatibility
The analyzer is compatible with multiple species including dogs, cats, horses, pigs, cattle, sheep, rabbits, monkeys, rats, mice, and 10 user-defined species.
Main Advantages
The PE-3020VET offers a smart software interface with precise and stable testing reports, multiple report formats, and adjustable printing orders. It supports PDF and JPG formats and includes an external barcode scanner for reagent management. The device supports two-way LIS systems and provides abnormal alarms and intelligent maintenance features. It includes four USB ports for connecting peripherals and offers mass data storage capabilities.
Working Principles
The analyzer uses the electrical impedance method for RBC and PLT data, where cell volume is determined by changes in DC resistance as cells pass through a detection hole. The colorimetric method is used for HGB measurement, where HGB derivatives are formed and measured using a 525nm wavelength light.
Contact Information
Produced by Shenzhen Prokan Electronics Inc., located at Yanda Tech-park, Fenghuanggang, Xixiang, Bao'an District, Shenzhen, China. Contact via email at [email protected] or [email protected], or by phone at +86-755-26955166. More information is available on their website www.prokanmed.com.
See more

Catalog excerpts

PE-3020VET-1

PE-3020vet 3 Diff Fully Auto Hematology Analyzer

 Open the catalog to page 1
PE-3020VET-2

Technical Parameters • Working Principles Electrical impedance method to measure the volume and number of cells HGB: colorimetric method Parameters 21 Parameters: WBC, Lym#, Gran#, Mid#, Lym%, Gran%, Mid%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RWD-CV, PLT, MPV, PDW, PCT, P-LCR,P-LCC 3 histograms:WBC/RBC/PLT Analysis Mode CBC Working Environment Temperature: 10-30 'C Humidity<70% Pressure: 70-106kPa Sample Volume Venous mode, Capillary mode, Pre-diluted mode<20pL Reagent Diluent, Lyse Display 10.4’’ IPS color touch screen Resolution: 800*600 Storage 50,000 sample results with histograms QC Mode...

 Open the catalog to page 2
PE-3020VET-3

Main Advantages 3 diff of WBC Precise and stable testing report The analyzer support Two-way LIS system Multiple report formats Abnormal alarms Adjustable printing order of different parameters; PDF and JPG format available The analyzer give alarms while abnormal happens External barcod scanner Intelligent maintenance Convenient for reading the reagents barcode Analyzer is equipped with automatically maintenance setting, operate the auto cleaning and mentain Data batch export Liquid level sensing Support date batch exported to PC side and U disk The analyzer give alarm while insufficient reagent...

 Open the catalog to page 3
PE-3020VET-4

Built-in Independence Operating System • 10.4" IPS color touch screen • No extra PC required Powerful Data Management • Flag info help diagnosis when result abnormal • 50,000 sample result storage with graphics, easy to transfer to PC side for analysis Easy Reagent Management • Only two kind: Diluent+Lyse • Simple management with reagent card Intelligent Monitoring • Real-time displaying reagent residual volume and valid date • Alarm when reagent insufficient, remind to replace • Strictly monitor analyzer temperature, voltage, pressure and current Smart Maintenance • Automatical routine maintance,...

 Open the catalog to page 4
PE-3020VET-5

Working Principle Electrical impedance method for determining the RBC and PLT data Ruby Hole (Aperture) Diluted sample Aperture Negative pressure Circuit When the cells in the diluted sample pass through the detection hole(Ruby hole) under the effect of constant negative pressure, the DC resistance between the electrodes will change, thus forming a proportional to the cell volume at both ends of the electrode Pulse signal. · The number of pulses stands for number of cells passing through the small hole. · The hight of the pulse stands for the volume of

 Open the catalog to page 5
PE-3020VET-6

Working Principle Colorimetric method for determining the HGB Electric receiver The red blood cells dissolved by lyse reagent release HGB. HGB combines with the lyse reagent to form a HGB On one side of the colorimetric pool, let the LED light pass through a monochromatic light tube with a wavelength of 525 nm, and then let the filted monochromatic light pass through the HGB derivatives. The electric receiver receives the transmitted light and converts the optical signal into an electrical signal. signal (voltage) generated by different light transmittances before and after adding the sample,...

 Open the catalog to page 6
PE-3020VET-7

PE-3020VET 3 Diff Fully Auto Hematology Analyzer SHENZHEN PROKAN ELECTRONICS INC. Production address: The 2nd Floor, Yanda Tech-park, Fenghuanggang, Xixiang, Bao'anDistrict, Shenzhen, China Email: [email protected] / [email protected] Follow us in Facebook

 Open the catalog to page 7
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.