Single Leads ECG Module M101B

Single Leads ECG Module M101B

Single Leads ECG Module M101B

Product catalog summary
Features
The M101B is a single-lead ECG module with dimensions of 132mm x 43mm x 17mm. It supports 3-lead ECG monitoring for adults, pediatric, and newborn patients. The module includes 26 arrhythmia analysis functions, verified by MIT and AHA databases, and complies with EC57 standardized tests. It offers four measurement modes: diagnosis (0.05Hz~130Hz), monitoring (0.5Hz~40Hz), HARDEST (5Hz~20Hz), and surgery (1Hz~25Hz). The module allows ECG calibration with a 1mV standard voltage and gain adjustment for waveform amplitude. Notch mode options include 50Hz, 60Hz, 50/60Hz, and off.
Specifications
  • ECG Algorithm: The module operates on a DC 12V±5% power supply with a power consumption of ≤3W. Communication is facilitated through TTL and USART.
  • Temperature: Operating range is 10°C~40°C (50°F~104°F), and storage range is -20°C~70°C (4°F~158°F).
  • Database Performance: The sensitivity of QRS wave detection is 99.60% (MIT) and 99.66% (AHA). The sensitivity of early detection is 99.72% (MIT) and 99.91% (AHA). The early detection accuracy rate is 92.0% (MIT) and 90.0% (AHA), with a false positive rate of 0.52% (MIT) and 0.638% (AHA).
  • ECG Range: 0.15mV-5.5mV with a resolution of 2.36uV/LSB.
  • Heart Rate (HR): Range is 15~300bpm with an accuracy of ±1bpm and a resolution of 1bpm.
  • Standards: Complies with AAMI EC57-2012 and IEC60601-2-25-2011.
Compliance
Data is subject to change without notice. Refer to the manual for contraindications and precautions.
Contact Information
Address: Building A-1201, Senyang High-tech Park, Yutang Street, Guangming District, Shenzhen, Guangdong, China
Website: www.szwitleaf.com
Tel: +86 0755-21384132
Email: [email protected]
See more

Catalog excerpts

Single Leads ECG Module M101B-1

Single Leads ECG Module M101B The M101B single-lead ECG measurement module can measure ECG and is used in patient monitoring. With 3-lead ECG monitoring function Patient model: adult, pediatric and newborn 26 arrhythmia analysis functions, verified by MIT and AHA databases, and passed EC57 standardized tests Provides four measurement modes: diagnosis, monitoring, HARDEST and surgery mode Diagnosis mode: filter range is 0.05hz~130hz Monitoring mode: filter range is 0.5hz~40hz HARDEST mode: the filtering range is 5hz~20hz Surgery mode: filter range is 1hz~25hz ECG calibration: input 1mv standard voltage to record the waveform amplitude Gain setting: adjust the amplitude of the ECG waveform Notch mode: 50Hz, 60Hz, 50/60Hz and close notch mode commands can be selected For more information, please contact us: sales@sz

 Open the catalog to page 1
Single Leads ECG Module M101B-2

lead type Electical Specification Power supply Power consumption Compliance Standard Sensitivity of QRS wave detection/ Q Se Positive predictive degree of QRS complex detection/Q +P Sensitivity of room early detection/ V Se Early detection accuracy rate /V +P False positive rate of early detection /V FPR * The data is subject to change without notice. Please refer to the manual for the contraindications and precautions Address:Building A-1201, Senyang High-tech Park, Yutang Street, Guangming District, Shenzhen, Guangdong, China Tel:+86 0755-21384132 Email:[email protected]

 Open the catalog to page 2