1. Catalogs
  2. SIEVE France
  3. IM range isolator - IM1111

IM range isolator - IM1111

IM range isolator - IM1111

IM range isolator - IM1111

Product catalog summary
Field of Application
  • Sterile medications
  • Eye drops
  • Chemotherapy
  • Monoclonal antibodies
  • Clinical trials
Composition
  • 1 technical block with entrance airlock
  • 1 isolator equipped with 1 workstation
  • 1 technical block with exit airlock
Specifications of IM1111 Isolator
  • Equipped with 1 workstation, sterilizable entry, and exit airlocks
  • Supply voltage: 230 V single-phase
  • Protection: 30 mA residual current circuit breaker - 20A Type D
  • Outlet protection current: 30 mA residual current device - 16A type C
  • Average power consumption: 300 W
  • Maximum consumption: 1700 W
  • Weight: Inlet/outlet technical unit - 183 kg, Isolator - 176 kg, Output technical unit - 117 kg, Total - 476 kg
  • Ground load: 233 kg/m²
  • Sound pressure during production: < 65 dB(A)
Dimensions
  • Length: 2400 mm (94.49 inch)
  • Width: 850 mm (33.46 inch)
  • Height: 1900 mm (74.8 inch)
Electrical Specifications
  • Power supply: 2 cables, minimum 3G2.5mm²
  • Computer network: 3 network cables with 4 pairs (1 VPN + 1 spare + 1 per workstation)
Pneumatic Characteristics
  • Pressure: 4 bars minimum
  • Consumption: < 1 liter/minute
  • Air quality: Clean and dry
Air Flow Characteristics
  • Extraction via airlock: 50 m³/hour (Only during rinsing)
  • Extraction via isolator: 100 m³/hour
Other Requirements
  • Pneumatic power supply: 1 Female connector 1/4 or 1/8
  • Air extraction from the isolator: 1 PVC pressure pipe Ø 63 mm
  • Extraction of sterilizing agent vapors: 1 PVC pressure pipe Ø 63 mm
See more

Catalog excerpts

IM range isolator - IM1111-1

IM range isolator Isolator intended for handling toxic and/or sterile products. ♦ 1 technical block 1 entrance airlock ♦ 1 isolator equipped with 1 workstation ♦ 1 technical block 1 exit airlock 1 isolator equipped with: > 1 sterilizable entry airlock > 1 sterilizable exit airlock MANUDO GROUP

 Open the catalog to page 1
IM range isolator - IM1111-2

electrical specifications: 30 mA residual current circuit breaker- 20A Type D 30 mA residual current device - 16A type C Pneumatic characteristics: _ 4 bars minimum < 1 liter / minute Clean and dry Other characteristics:

 Open the catalog to page 2

All SIEVE France catalogs and technical brochures