Overview:
The document describes an isolator designed for handling toxic and/or sterile products, suitable for applications such as sterile medicines, eye drops, chemotherapy, monoclonal antibodies, and clinical trials.
Field of Application:- Sterile medicines
- Eye drops
- Chemotherapy
- Monoclonal antibodies
- Clinical trials
Composition:- 1 technical block with 2 entry/exit airlocks
- 1 isolator with 2 workstations facing each other
- 1 technical block with 2 exit airlocks
IM Range Isolators - IM1222EC:- 1 isolator with 2 workstations facing each other
- 2 entry/exit airlocks
- 2 exit airlocks
Dimensions:- Length: 2400 mm (94.49 inch)
- Width: 1410 mm (55.51 inch)
- Height: 1900 mm (74.80 inch)
Electrical Specifications:- Supply voltage: 230 V Single-phase
- Protection: 30 mA residual current device - 20A curve D
- Power outlet protection: 30 mA residual current device - 16A curve C
- Average power consumption: 300 W
- Maximum power consumption: 3100 W
Pneumatic Characteristics:- Pressure: 4 bars minimum
- Consumption: < 1 liter/minute
- Air quality: Clean and dry
Air Flow Characteristics:- Extraction via airlock: 50 m3/hour (only during rinsing)
- Extraction via isolator: 100 m3/hour
Other Characteristics:- Weight of technical inlet/outlet unit: 322 kg (709.89 lb)
- Isolator weight: 340 kg (749.57 lb)
- Weight of output technical unit: 210 kg (462.97 lb)
- Total weight: 872 kg (1922.43 lb)
- Ground load: 258 kg/m² (568.79 lb)
- Sound pressure during production: < 65 dB(A)
Requirements:- Power supply: 2 Cable 3G2.5mm² minimum
- Computer network: 4 network cables, 4 pairs: 1 VPN connection + 1 spare + 1 per workstation
- Pneumatic supply: 1 female 1/4 or 1/8 connector
- Air extraction from the isolator: 1 PVC pressure pipe, minimum diameter 63 mm (2.48 inch)
- Extraction of sterilizing agent vapors: 1 PVC pressure pipe, minimum diameter 63 mm (2.48 inch)
See more