IM Range Isolator Overview
Use: The isolator is designed for handling toxic and/or sterile products, suitable for applications such as sterile medicines, eye drops, chemotherapy, monoclonal antibodies, and clinical trials.
Field of Application:
- Sterile medicines
- Eye drops
- Chemotherapy
- Monoclonal antibodies
- Clinical trials
Composition:
- 1 block technique with 1 airlock
- 1 isolator equipped with 2 workstations
- 2 side-by-side workstations
- 2 entry/exit airlocks
- 2 exit airlocks
Layout and Dimensions:
- Length: 3,650 mm (143.7 inch)
- Width: 850 mm (33.46 inch)
- Height: 1,900 mm (74.8 inch)
Electrical Specifications:
- Supply voltage: 230 V single-phase
- Protection: 30 mA residual current device - 20A type D
- Power outlet protection: 30 mA residual current device - 16A type C
- Average power consumption: 500 W
- Maximum power consumption: 3,440 W
Pneumatic Characteristics:
- Pressure: 6 bars minimum
- Consumption: < 1 litre/minute
- Air quality: Clean and dry
Air Flow Characteristics:
- Extraction via airlock: 50 m3/hour (only during rinsing)
- Extraction via isolator: 200 m3/hour
Other Characteristics:
- Weight of left-hand technical inlet unit: 270 kg (595.25 lb)
- Isolator Weight: 180 kg (396.83 lb)
- Weight of right-hand technical input block: 270 kg (595.25 lb)
- Total weight: 720 kg (1587.33 lb)
- Load at ground: 232 kg/m² (511.47 lb)
- Sound pressure during operation: < 65 dB(A)
Requirements:
- Power supply: 2 Cables 3G2.5mm² minimum
- Computer network: 4 Network Cables, 4 pairs (1 VPN + 1 spare + 1 per Workstation)
- Pneumatic supply: 1 female connector 1/4 or 1/8
- Air extraction from the isolator: 2 PVC pressure pipes 63 mm (2.8 inch)
- Extraction of sterilizing agent vapors: 2 PVC pressure pipes 63 mm (2.8 inch)
See more