Anesthetic Gas Scavenging Inlet - Non venturi type - 1

Medical Gas Systems Since 1950 Anesthetic Gas Scavenging Inlet Non venturi type SILBERMANN Anesthetic Gas Scavenging Inlet, Non venturi type is designed to safely remove surplus anesthetic gas and explosive gas mixtures from the operating theatre. It's purpose is to avoid long-term health hazards for the exposed medical staff. The Inlet is a quick connect type, connected to a central AGSS system. It withstands pressure tests without the need for a dedicated plug. ■ Operation based on a central AGSS system. ■ Internal adjusting valve ensures the required vacuum flow. ■ Very quiet operation ■ Automatic shut off when probe is disconnected ■ Locking mechanism quick connect Medical Gas System Standards: The manufacturer reserves the right to make changes without notification

Open the catalog to page 1
Anesthetic Gas Scavenging Inlet - Non venturi type - 2

Medical Gas Pipeline Technical Data ■ Material of construction: ■ Box: Baked painted Aluminium ■ Front Panel: Aluminum ■ Inlet pressure: 3.5-5 bars ■ Free air suction capacity: ■ Vacuum level: Adjustable 20-50 cmH2O ■ Exhaust pipe diameter: OD 7/8” nominal 20 cm long, K type. The manufacturer reserves the right to make changes without notification

Open the catalog to page 2
Anesthetic Gas Scavenging Inlet - Non venturi type - 3

Anesthetic Gas Scavenging Inlet Non venturi typeCatalogue Numbers Inlet Configuration Flexible Hose Assembly for OT Hose Assembly 5 meter long includes: ■ Flexible hose - 5m long, 23 mm diameter. ■ Rear adapter (to anaesthetic machine) Address: #5 Harakevet st. POB 4605, Petah-Tikva, 4900715, ISRAEL Tel: +972-3-9309186 | Fax: +972-3-9315721 E-Mail: info@silbermnn.com, www.silbermnn.com /SSx Medical Gas Systems Since 1950

Open the catalog to page 3