Validation and Comparison of SOMNOtouch™ NIBP:
The SOMNOtouch™ NIBP device has been validated against several methods for measuring blood pressure (BP), including cuff-based methods during ergometry and over 24 hours, the Penaz method, and intra-arterial BP measurement. It is listed as a recommended device on the dabl® Educational Trust website.

Pulse Transit Time (PTT) Validations:
The device uses a patented Pulse Transit Time (PTT) technique, validated against other methods. The PTT method measures the time for a pulse wave to travel from the heart to the fingertip, allowing continuous, non-invasive BP monitoring.

Key Validation Studies:
  • ESH-IP 2010 Protocol: Validated with 33 patients across BP strata, fulfilling all requirements with strong correlations for both systolic (SBP) and diastolic blood pressure (DBP).
  • Comparison to Cuff-Based Method: In studies with healthy subjects and during exercise, PTT showed significant correlation with cuff-based methods.
  • 24-Hour Ambulatory BP Measurement: Demonstrated reliability over 24 hours, with some measurements affected by artefacts like motoric activity and arrhythmia.
  • Children and Adolescents: In a study with 90 participants aged 5-17, PTT showed minimal deviation from cuff-based methods over 24 hours.
  • Apnea-Related BP Changes: PTT reliably detected BP changes related to apnea and hypopnea.
  • Intra-Arterial Measurement Comparison: PTT showed high correlation with intra-arterial measurements in both healthy subjects and cardiology patients.

Device Features:
The SOMNOtouch™ NIBP offers cuffless, continuous BP monitoring, reducing artefacts and improving patient adherence. It is designed for ease of use and reliability, suitable for long-term diagnostics in sleep medicine, cardiology, and neurology.

About SOMNOmedics:
SOMNOmedics specializes in diagnostic devices for sleep medicine, cardiology, and neurology, providing innovative and mobile solutions for patient health insights.
See more

Catalog excerpts

PTT-Validations-1

PTT Validations (PULSE TRANSIT TIME) VALIDATION ACCORDING TO THE ESH-IP 2010 PROTOCOL AGAINST CUFF-BASED METHOD DURING ERGOMETRY AGAINST CUFF-BASED METHOD OVER 24 HOURS AGAINST THE PENAZ METHOD AGAINST INTRA-ARTERIAL BP MEASUREMENT SOMNOtouchTM NIBP is listed on the dabl®Educational Trust website as a recommended device http://www.dableducational.org/sphygmomanometers/devices_1_clinical.html#ClinTable

 Open the catalog to page 1
PTT-Validations-2

(PULSE TRANSIT TIME) We have validated our patented1 PTT technique against other current methods as described below: Validation of the Somnotouch-NIBP noninvasive continuous blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010. Bilo, G„ et al. (2015), Blood Press Monit, 20(5): p. 291 -4. ♦ 33 patients from low, medium and high BP strata as required by the ESH-IP ♦ All validation requirements of the ESH-IP were fulfilled: Table 1: Validation results (adapted from Bilo et al.) * two out of three required for the first threshold; three out...

 Open the catalog to page 2
PTT-Validations-3

♦ After exclusion of values affected by artefacts: • SBP: 1.4 mmHg, + 18/- 15 mmHg; DBP: 0.3 mmHg, + 14/- 13 mmHg ♦ This study demonstrates that the PTT method is a reliable and favorable method for BP measurement over 24 hours. Accuracy of cuff-less, continuous, and non-invasive blood pressure measurement in 24-h ABPM in children aged 5-17. Bothe, T.L., et al. (2023), Blood Pressure, 32(1): p. 2255704. ♦ 90 children and adolescents included (5-17 years), PTT (SOMNOtouch™ NIBP) against cuff-based method, 24 hours ♦ Additionally recorded: 3-channel ECG, motoric activity, finger plethymogram, oxygen...

 Open the catalog to page 3
PTT-Validations-4

Turning point in plethysmogram measurement allows us to `set` the software‘s algorithm – and measure blood pressure continuously throughout the day and night. Because every single pulse wave is detected, a continuous “Beat-to-Beat”-recording and analysis is possible. Reduce artefacts and increase the number of reliable data points – all in a continuous, non-invasive and non-reactive way. The SOMNOtouch™ NIBP (Non-Invasive Blood Pressure) is a new era in CUFFLESS ambulatory blood pressure monitoring (ABPM). About us SOMNOmedics designs, manufactures & distributes diagnostic devices that streamline...