
Technical Data Sheet Blood Leak Detector Original
Open the catalog to page 1Technical Data Sheet | General Blood Leak Detector BLD03/50I The Blood Leak Detector BLD03/50I is a non-invasive, optical sensor designed to detect smallest amounts of blood in a clear fluid through an optical transparent and flexible tubing. The BLD03/50I consists of a sensor head with connection line and an external electronic board (PCB) with integrated firmware. The sensor is developed as a built-in component integrated into a medical device. Article number
Open the catalog to page 2Technical Data Sheet | Sensor Head BLD03/50 Blood Leak Detector BLD03/50I Sensor Head BLD03/50 Technical data Measuring method Optical transmission measurement, LED Specification Sensor head BLD03/50 - channel width 5.0 mm Materials Housing: ABS and glass Measuring channel Width: 5.0 mm; Height: 5.8 mm Mounting Two mounting holes (0 = 3.25 mm) © The sensor must be installed on a plane surface. Working humidity 20 % ... 75 % relative humidity (not condensing) Storage humidity 10 % ... 90 % relative humidity (not condensing) Protection IP67 (Sensor head without connector) © Please provide us with...
Open the catalog to page 3Technical Data Sheet | Sensor Head BLD03/50 Blood Leak Detector BLD03/50I Technical data Special features Tube must be optically transparent within the spectral range of about 415 nm Elasticity Tube must be able to adjust flexibly Liquid requirements Optically transparent liquids Cable Sensocord®-M-UL, 5 x AWG 28, unshielded, black, Length: 500 mm ± 20 mm Connector Molex PicoBlade 51021-0500, Single Row, 5 circuits, 1.25 mm Directives/Standards The sensors were developed to be tested with respect to the following Scope of delivery Sensor Head BLD03/50 (connection cable included) Technical data...
Open the catalog to page 4Technical Data Sheet | Printed Circuit Board BLD03I Blood Leak Detector BLD03/50I Printed Circuit Board BLD03I Technical data Measuring method Optical transmission measurement Specification Printed Circuit Board BLD03I V1.1, FW V04.05 Blood threshold Light attenuation of 8 % compared to the calibrated light intensity (default (adjustable) setting) © Adjusting the blood threshold will have direct impact on blood sensitivity. Blood sensitivity At blood threshold of 8 % (default setting): Reliable detection of a blood leakage > 0.35 ml/min at a dialysis fluid flow rate of 800 ml/min, haematocrit...
Open the catalog to page 5Technical Data Sheet | Printed Circuit Board BLD03I Blood Leak Detector BLD03/50I Technical data © Stresses beyond the maximum rating may cause permanent damage to the device. Safety aspects Self-test - A self-test routine can be triggered via digital input. Working humidity 20 % ... 75 % relative humidity (not condensing) Storage humidity 10 % ... 90 % relative humidity (not condensing) Directives/Standards The sensors were developed to be tested with respect to the following • Software is developed acc. to IEC 62304:2006 + AMD1:2015 The embedded software is classified as "A" Technical Data...
Open the catalog to page 6Technical Data Sheet | Printed Circuit Board BLD03I Blood Leak Detector BLD03/50I Connector Board connector header; 5-pin; 1.25 mm Molex: 53261-0571 Board connector header; 4-pin; 1.25 mm Molex: 53261-0471 Pin Connection Supply voltage 3.1 ... 5.5 VDC Serial interface Parameter Value Bits per second 115200 Data bits 8 Parity none Flow control none
Open the catalog to page 7Technical Data Sheet | Printed Circuit Board BLD03I Blood Leak Detector BLD03/50I Connector Board connector header; 6-pin; 1.25 mm Molex: 53261-0671 4 Input "Self-Test" (high active) 6 Input "Calibration" (high active)
Open the catalog to page 8Technical Data Sheet | Printed Circuit Board BLD03I Blood Leak Detector BLD03/50I Signal level Condition Logical Output specification High Blood Low No blood (clear liquid) High No plausibility error nor device error Low Plausibility error or device error Signal at input Self-Test High Start self-test Calibration High Start calibration Signal at output Output Voh Ioh = -100 pA 3.1 ... 5.5 V Vcc-0.1 --- --- V Signal at input Input Vih High Voltage
Open the catalog to page 9Technical Data Sheet | Printed Circuit Board BLD03I Blood Leak Detector BLD03/50I Technical drawing Dimensions in mm Remnants of material up to 0,5mm (due to separation of the PCB production panel) on a sides (bottom side) (due to separation of the PCB production panel) Drawings are not to scale. Dimensions in mm, unless otherwise specified. Information is subject to change without notice. SONOTEC is a registered trademark. Manufacturer Contact USA SONOTEC GmbH Nauendorfer Str. 2 06112 Halle (Saale) Germany Suite 8, 2nd Floor [email protected] Islandia, New York 11749, USA www.sonotecusa.com...
Open the catalog to page 1016 Pages
2 Pages
2 Pages
6 Pages
6 Pages
4 Pages
4 Pages
4 Pages
4 Pages
4 Pages
10 Pages
2 Pages
2 Pages
2 Pages
6 Pages
2 Pages
2 Pages