Technical Data Sheet Blood Leak Detector Original
Open the catalog to page 1Technical Data Sheet | General Blood Leak Detector BLD03/50M General Description The Blood Leak Detector BLD03/50M is a non-invasive, optical sensor designed to detect smallest amounts of blood in a clear fluid through an optical transparent and flexible tubing. The sensor consists of a sensor head with connection line and an external electronic board (PCB) with integrated firmware. The BLD03/50M meets highest demands on patient safety and reliability: The firmware has been designed according to the requirements of the international standard IEC 62304:2006 + AMD1:2015, safety class C. The sensor...
Open the catalog to page 2Technical Data Sheet | Sensor Head BLD03/50 Blood Leak Detector BLD03/50M Sensor Head BLD03/50 Technical data Measuring method Optical transmission measurement, LED Specification Sensor head BLD03/50 - channel width 5.0 mm Materials Housing: ABS and glass Measuring channel Width: 5.0 mm; Height: 5.8 mm Mounting Two mounting holes (0 = 3.25 mm) © The sensor must be installed on a plane surface. Working humidity 20 % ... 75 % relative humidity (not condensing) Storage humidity 10 % ... 90 % relative humidity (not condensing) Protection IP67 (Sensor head without connector) © Please provide us with...
Open the catalog to page 3Technical Data Sheet | Sensor Head BLD03/50 Blood Leak Detector BLD03/50M Special features Tube must be optically transparent within the spectral range of about 415 nm Elasticity Tube must be able to adjust flexibly Liquid requirements Optically transparent liquids Cable Sensocord®-M-UL, 5 x AWG 28, unshielded, black, Length: 500 mm ± 20 mm Connector Molex PicoBlade 51021-0500, Single Row, 5 circuits, 1.25 mm Directives/Standards The sensors were developed to be tested with respect to the following Scope of delivery Sensor Head BLD03/50 (connection cable included) Technical data sheet Labelling...
Open the catalog to page 4Technical Data Sheet | Printed Circuit Board BLD03M Blood Leak Detector BLD03/50M Printed Circuit Board BLD03MTechnical Data Measuring method Optical transmission measurement Specification Printed Circuit Board BLD03M V1.1, FW V04.05 Blood threshold Light attenuation of 8 % compared to the calibrated light intensity Blood sensitivity Reliable detection of a blood leakage > 0.35 ml/min at a dialysis fluid flow rate of 800 ml/min, haematocrit level of 32 %. © The indication above is a worst case scenario. Depending on tube properties, application and process characteristics the sensor may detect...
Open the catalog to page 5Technical Data Sheet | Printed Circuit Board BLD03M Blood Leak Detector BLD03/50M Technical Data Safety aspects Self-test - A self-test routine can be triggered via digital input. Working humidity 20 % ... 75% relative humidity (not condensing) Storage humidity 10 % ... 90 % relative humidity (not condensing) Directives/Standards The sensors were developed to be tested with respect to the following • Software is developed acc. to IEC 62304:2006 + AMD1:2015 The embedded software is classified as "C" Technical Data Sheet Operating Manual Labelling Each Printed Circuit Board is labelled with its...
Open the catalog to page 6Technical Data Sheet | Printed Circuit Board BLD03M Blood Leak Detector BLD03/50M Electrical Connection ATTENTION The board needs to be protected against unintended contact with other conductive parts. J2 – Connection to Sensor Head Connector Board connector header; 5-pin; 1.25 mm Molex: 53261-0571 J3 – Serial Interface Intended for use with another software version. Connector Board connector header; 4-pin; 1.25 mm Molex: 53261-0471 Operating voltage
Open the catalog to page 7Technical Data Sheet | Printed Circuit Board BLD03M Blood Leak Detector BLD03/50M Connector Board connector header; 6-pin; 1.25 mm Molex: 53261-0671 3 Output "Blood Alarm" 4 Input "External Self-Test" (high active) 6 Input "Calibration" (high active) Signal level Condition High Blood Low No blood (clear liquid) High No plausibility error nor device error Low Plausibility error or device error Signal at input Self-Test S05-007 Signal level Condition High Start self-test High Start calibration
Open the catalog to page 8Technical Data Sheet | Printed Circuit Board BLD03M Blood Leak Detector BLD03/50M Logical level Parameter Condition Operating Min Typ Max Unit Signal at output Output Voh Ioh = -100 pA 3.1 ... 5.5 V Vcc-0.1 --- --- V Signal at input Input Vih High Voltage
Open the catalog to page 9Technical Data Sheet | Printed Circuit Board BLD03M Blood Leak Detector BLD03/50M Technical drawing Dimensions in mm Remnants of material up to 0,5mm (due to separation of the PCB production panel) on a sides (bottom side) (due to separation of the PCB production panel) Drawings are not to scale. Dimensions in mm, unless otherwise specified. Information is subject to change without notice. SONOTEC is a registered trademark. Manufacturer Contact USA SONOTEC GmbH Nauendorfer Str. 2 06112 Halle (Saale) Germany Suite 8, 2nd Floor [email protected] Islandia, New York 11749, USA www.sonotecusa.com...