91496 Command Module Data Sheet The Spacelabs 91496 Command Module is designed to provide a flexible selection of clinical measurements and compatibility with Spacelabs patient monitors. Choose the Command Module measurement offering based on your care unit needs or patient acuity. Durable and lightweight, Command Modules can easily be moved between monitors. With the optional Data Shuttle® feature, transfer up to 24-hours of patient demographics, physiological data, and applicable measurement settings between monitors without having to re-cable patients. ECG Available leads 3-leadwire Measurement range Adjustable from 0.5 to 10 mV/cm Numeric update rate Every 3 seconds or immediately at onset of alarm Heart rate accuracy Frequency range Gain accuracy Maximum input Sample rate Arrhythmia classifications Atrial Fibrillation, Asystole, Couplet, Paroxysmal Supraventricular Tachycardia, Pause, Ventricular Ectopic, Ventricular Fibrillation, Ventricular Run

Open the catalog to page 1
Ultraview SL™ Command Module - 2

91496 Command Module Data Sheet 12-leads; standard or Cabrera format Diagnostic sample rate Performed on up to 2 leads simultaneously, default II and V, with auto lead switch in event of lead off Duration Pacemaker detection 40 to 120 ms Indicator of pace pulse on waveform display (user selectable) Rise time 10% of width not to exceed 100 μsec Up to nine 1-second ST segment sets for up to 12 leads Defibrillator Sync Input Input level Input impedance hlo1 supports TT-253 plug, P/N 354171-001, for connection to external device Input display Defibrillator sync marker is displayed with the...

Open the catalog to page 2
Ultraview SL™ Command Module - 3

91496 Command Module Data Sheet Pacemaker pulses hlo2 supports TT-253 plug, P/N 354171-001, for connection to external device Note: Spacelabs Healthcare does not provide interface cables to third party equipment. Cables constructed for use with 354171-001 must incorporate a ferrite bead with 200 Ω impedance at 100 MHz within 25mm (1-inch) of the plug for EMI compliance. Respiration Measurement method Impedance pneumography Sensing leads User selected: RA/LA (R/L); RL/LL (N/F), RL/LA (N/L), or RA/LL (R/F) Detection sensitivity Shallow: 0.1 Ω input source impedance Normal: 0.25 Ω (normal) at...

Open the catalog to page 3
Ultraview SL™ Command Module - 4

91496 Command Module Data Sheet Initial cuff inflation Adult/Pediatric 4: 170 mmHg Pediatric 2/3: 130 mmHg Neonate/Pediatric 1: 115 mmHg Maximum cuff inflation Adult/Pediatric 2/3/4: 290 mmHg Neonate/Pediatric 1: 150 mmHg Cuff deflation rate Adult/Pediatric: <10 seconds from 260 to 15 mmHg (34.7 to 2 kPa) Neonate: <5 seconds from 150 to 5 mmHg (20 to 0.7 kPa) Adult/Pediatric: female bayonet plastic or metal fitting compliant with IEC 80369-5, with single airway tube Neonate: male bayonet plastic or metal fitting complaint with IEC 80369-5, with single airway tube Invasive Blood Pressure...

Open the catalog to page 4
Ultraview SL™ Command Module - 5

91496 Command Module Data Sheet Spacelabs Pulse Oximetry (SpO 2 ) Measurement method Functional saturation (oxygen saturation of functional hemoglobins) Display parameters %SpO2, %SpO2D (second SpO2 available with 91496-L), pulse rate Measurement range SpO2 accuracy (Arms) of Spacelabs sensors 70% to 100% ±3% 0% to 69% unspecified Established accuracy is the root-mean-square of the error between measured values and reference values obtained from a laboratory hemoximeter during adult human blood studies. Assuming a normal distribution, Arms encompasses 68% of the data population. Numeric...

Open the catalog to page 5
Ultraview SL™ Command Module - 6

91496 Command Module Data Sheet SpO2 Accuracy (70% to 100%) of Masimo sensors LNCS DC-I, LNCS DC-IP, LNCS TF-I**, LNCS Adtx, LNCS Pdtx, LNCS Inf LNOP DC-I*, LNOP DC-IP*, LNOP Adt*, LNOP Pdt*, LNOP Inf-L* LNOP Neo*, LNOP NeoPt*, LNOP NeoPt-L*, LNOP TC-I** LNOP Neo-L The accuracy specification under motion conditions is ±3%. Motion is defined as continuous rubbing and tapping motions at 2 to 4 Hz, at an amplitude of 1 to 2 cm, and continuous random frequency motion between 1 to 5 Hz, at an amplitude of 2 to 3 cm. ** These sensors were not validated under motion conditions. † Pulse amplitude...

Open the catalog to page 6
Ultraview SL™ Command Module - 7

91496 Command Module Data Sheet SpO2 Accuracy (70% to 100%) of Nellcor sensors MAX-A*, MAX-AL*, MAX-N*†, MAX-P*, MAX-I*, MAX-FAST OxiCliq A, OxiCliq P, OxiCliq N (Adult)†, OxiCliq I MAX-N*, D-YS (Infant to Adult), DS-100A, OXI-AN, OXI-P/I MAX-R**, OxiCliq N (Neonate), D-YSE * The accuracy specification under motion conditions is ±3%. ** The accuracy specification has been determined between saturations of 80% and 100%. † The MAX-N and the OxiCliq N were tested on patients >40 kg. Neonatal accuracy When sensors are used on neonatal subjects as recommended, the specified accuracy range is...

Open the catalog to page 7
Ultraview SL™ Command Module - 8

91496 Command Module Data Sheet Physical Specifications Dimensions (H x W x D) Stop NIBP Volatile memory Data is preserved for 10 minutes. Module ceases data collection when power is removed. Electrical Specifications Power source Powered directly from monitor Protection against electric shock Module is electrically isolated from host device. Defibrillator protection Type CF, meets IEC 60601-2-27, AAMI EC-13 Environmental Specifications Ambient temperature Operating: 0° to 50° C (32° to 122° F) Storage and transport: -40° to 75° C (-40° to 167° F) Relative humidity Operating: 95%...

Open the catalog to page 8
Ultraview SL™ Command Module - 9

Software options Arrhythmia* (F–Basic, G–Standard, or H–Advanced) Diagnostic 12-lead reports* (D–with interpretation or E–without interpretation) SpO2 technology (M–Masimo SET, N–Nellcor OxiMax, or U–Spacelabs) Data Shuttle* (Q) Respiration* (R) ST segment analysis* (S) Varitrend® oxycardiorespirogram* (V) Not available with 91496-I and 91496-L configurations. For further ordering information and details on compatible accessories, please contact your Spacelabs Healthcare representative or refer to the product documentation. This product may not be approved for market release in all...