Ultraview SL™
14Pages

{{requestButtons}}

Catalog excerpts

Ultraview SL™ - 1

Summary The Ultraview SL™ Command Module is a multiparameter module used with Spacelabs Healthcare patient monitors. Five different parameter configurations provide vital sign monitoring for any range of acuity. Configurable settings assure customized care for neonatal, pediatric, and adult patients. Features Module Configurations Note: Module use is restricted to one patient at a time. Noninvasive Parameter Set Provides multi-lead ECG, respiration, noninvasive blood pressure, pulse oximetry, and two temperature channels Invasive Parameter Set Provides multi-lead ECG, respiration, noninvasive blood pressure, pulse oximetry, and two temperature channels, plus two invasive pressure channels Invasive Parameter Set with Cardiac Output Provides multi-lead ECG, respiration, noninvasive blood pressure, pulse oximetry, and two temperature channels, plus four invasive pressure channels and thermodilution cardiac output Noninvasive Parameter Set Provides noninvasive blood pressure, pulse oximetry, and two temperature channels Invasive Parameter Set Provides four invasive blood pressure channels, pulse oximetry, and two temperature channels Note: 91496-L is intended for use in conjunction with another module to provide additional parameter monitoring, such as dual SpO2 (SpO2D).

Open the catalog to page 1
Ultraview SL™ - 4

Gain Accuracy Ventricular Beats Per Minute Counter Displays counts up to 99 beats per minute Heart Rate Range Heart Rate Resolution Heart Rate Alarm Limits 15 to 300 bpm; heart rates >300 bpm are displayed as “+++” 1 bpm • High — 5 to 300 bpm • Low — 0 to 200 bpm Numeric Update Rate Test Signal Display Size Bedside Display Waveform Sweep Speeds High Level Analog Output Connector (Front Panel) Dynamic Range Gain Output Impedance Defibrillator Sync Input Input Level Input Impedance ST Segment Analysis Resolution Range Leads Alarms Displays Snapshot Storage ±1% or 3 beats per minute (whichever...

Open the catalog to page 4
Ultraview SL™ - 5

Diagnostic ECG Analysis Instrument Type Standard Leads Input Channels Simultaneous acquisition of all standard leads Sample Rate Algorithm Features Arm lead reversal detection, lead-off detection, artifact detection, baseline correction, line voltage rejection Printer Speed Respiration Input 10-lead, 5-lead, or 3-lead ECG cable (cables use 1 kΩ resistors (±10%) in series with each electrode) Measurement Technique Impedance pneumography through ECG leads RA/LA (R/L), RL/LL (N/F), RL/LA (N/L), or RA/LL (R/F) Patient Source Impedance Excitation Frequency Excitation Amplitude 120 µAmp (±20%)...

Open the catalog to page 5
Ultraview SL™ - 6

Waveform Sweep Speeds 25, 12.5, 6.25, or 1.56 mm/sec High Level Analog Input/ Output Connector (Front Panel) 4.42 mm (0.174 in) diameter, three conductor TT-phone plug Dynamic Range ±4 V minimum Transducer Type Strain-gauge, standardized to 5 pV/V/mmHg ±1% Dynamic Waveform -50 to +500 mmHg (-6.7 to +66.7 kPa) Measurement Units mmHg or kPa Measurement Range -50 to +300 mmHg (-6.7 to +40 kPa); displays “+++” for pressures >+300 mmHg Accuracy ±2 mmHg (0.27 kPa) or 2% of reading (whichever is greater) Zero Drift (exclusive of transducer) <0.1 mmHg/° C (<0.01 kPa/° C) after a 5-minute warm Zero...

Open the catalog to page 6
Ultraview SL™ - 7

Ultraview SLCommand Module (91496)High Level Analog Input/ Output Connector (Front Panel)

Open the catalog to page 7
Ultraview SL™ - 8

NIBP Alarm Limit Ranges Neonate/Pediatric 1 • High — 20 to 140 mmHg (2.6 to 18.7 kPa) • Low — 15 to 135 mmHg (2 to 18 kPa) Pediatric 2/3 • High — 35 to 190 mmHg (4.7 to 25.4 kPa) Pediatric 4/Adult • High — 35 to 260 mmHg (4.7 to 34.7 kPa) Measurement Method Functional saturation (oxygen saturation of functional hemoglobins) Measurement Range O2 Saturation 30% to 100% Measurement Accuracy Established accuracy is the root-mean-square of the error between measured (Arms) values and reference values obtained from a laboratory hemoximeter during adult human blood studies. Assuming a normal...

Open the catalog to page 8
Ultraview SL™ - 9

Measurement Method Functional saturation (oxygen saturation of functional hemoglobins) Measurement RangeO2 Saturation 1% to 100% Pulse Rate 25 to 240 bpm Masimo SET SpO2 These sensors have been clinically validated by Masimo using the Masimo MS-11 Measurement Accuracy oximetry board. Saturation Accuracy 70 to 100% Masimo Sensor Models Weight RangeNo Motion Low Perfusion* LNCS Reusable Sensors

Open the catalog to page 9
Ultraview SL™ - 10

* The accuracy specification under motion conditions is ±3%. Motion is defined as continuous rubbing and tapping motions at 2 to 4 Hz, at an amplitude of 1 to 2 cm, and continuous random frequency motion between 1 to 5 Hz, at an amplitude of 2 to 3 cm. ** These sensors were not validated under motion conditions. f Pulse amplitude >0.2%; % transmission >5% (LNOP Y-I sensor was not validated for low perfusion). ff Saturation accuracy under no motion for neonatal, infant, or pediatric patients with congenital cyanotic cardiac lesions ±3% for 80 to 100%, ±4% for 60 to 80%, and ±3.3% for 70 to...

Open the catalog to page 10
Ultraview SL™ - 11

Measurement Method Functional saturation (oxygen saturation of functional hemoglobins) Measurement Range O2 Saturation 1% to 100% Nellcor OxiMax These sensors have been clinically validated by Nellcor using the Nellcor NELL-1 Measurement Accuracy oximetry board. Nellcor Sensor ModelsSaturation Accuracy 70% to 100% OxiMax Sensors, Single Patient Use * The accuracy specification under motion conditions is ±3%. ** The accuracy specification has been determined between saturations of 80% and 100%. f The MAX-N and the OxiCliq N were tested on patients >40 kg.

Open the catalog to page 11
Ultraview SL™ - 12

Neonatal Accuracy When sensors are used on neonatal subjects as recommended, the specified accuracy range is increased by ±1 digit as compared to adult usage, to account for the theoretical effect on oximeter measurements of fetal hemoglobin in neonatal blood. For example, MAX-N accuracy on neonates is ±3 digits, rather than ±2 digits. Saturation Resolution Pulse Rate Resolution Saturation Alarm Limits Numeric Update Nellcor Sensors Temperature Probe Type Sample Rate Measurement Range Display Parameters Accuracy Resolution Numeric Update Rate Alarms Alarm Limits Cardiac Output (CO)...

Open the catalog to page 12