DATA SHEET PRIMUS Healthcare– Small Size Sterilizers These sterilizers are designed for Healthcare applications including surgery, central sterile supply, and clinical laboratories . PRIMUS sterilizers are designed and manufactured in the USA according to Quality Management Systems which are in compliance with ISO 9001:2008, ISO 13485:2003, CMDCAS and FDA’s Good Manufacturing Practice (GMP) for Medical Devices: General Regulation (21CFR Part 820). Small size sterilizers are one segment of 12 standard clinical sizes manufactured by PRIMUS and are recognized for government contracts via Federal Supply Schedule Contract #V797P-4400B. STERILIZER DESIGN The pressure vessel is the heart of any sterilizer. PRIMUS offers simple straight forward design. All PRIMUS pressure vessels are constructed of solid stainless steel and are fabricated by PRIMUS in our quality controlled, ASME facility. The interior chamber surface is polished to a mirror finish of <10μ in. Ra. The brilliant Pri-Mirror ® reflective chamber finish sets the highest standard for cleanliness and offers easy spill clean-up and resistance to staining and degradation. PRIMUS rectangular chambers eliminate the wasted space and high utility costs common to old-style cylindrical or elliptical chambers. (E.g. The PRIMUS 26 cu. ft. 26”x26”x67” sterilizer with vertical door, occupies less floor space and provides greater capacity than the conventional 24”x36”x48” hinged radial arm door sterilizers.) PRIMUS’ unique design reduces the high service cost associated with conventional hinged/radial arm doors and, most importantly, provides an added margin of operator safety. PRIMUS vertically operating doors operate with minimal hand pressure. Where hands-free door operation is desired (for single door units only), a mushroom button actuated by elbow/forearm is offered. An optional foot switch is available. For double door units ordered with the Power Door option the doors are operated through the use of the touchpad only. The PRIMUS vertically sliding door design is efficient, reliable and inherently safe STERILIZER CYCLES All PRIMUS Sterilizers include Gravity, Vacuum, and Liquid cycle settings. A Test/Vac cycle is provided for utilizing Bowie-DickTM tests. For surgical packs, select one of the vacuum cycles. For the sterilization of liquids in open or vented containers, select a liquid cycle. The time-tested and reliable PRIMUS PSS500 Microcomputer Control is an industrial microprocessor providing accuracy and automation for all customer requirements. This simple, versatile, control has become the first choice for both sterilizer users and service technicians. Sterilizer Company, LLC • 8719 South 135h Street, Ste 300 Phone (402) 344-4200 • Fax (402) 344-4242 •
Open the catalog to page 1TECHNICAL DATA SHEET PURCHASING TECHNICAL DATA SHEET GENERAL (Options italicized) All models include Vacuum, Gravity and Liquids cycles. A Low Temperature Flowing Steam option provides flowing steam at low temperatures for Research laboratories or specific applications. VESSEL MATERIAL AND CONSTRUCTION The sterilizer will have a double-wall design providing for a fully jacketed pressure vessel. The chamber cross section will be rectangular with width and height dimensions to include: EQUIPMENT WARRANTY Sterilizer pressure vessels manufactured by PRIMUS are warranted against defects in...
Open the catalog to page 2SINGLE DOOR TECHNICAL DATA SHEET PRIMUS Healthcare - Small Sterilizers VOLUME/DIMENSIONS CHART *Refer to General Arrangement (GA) drawings for details and final connection point to utility services (S-Steam, W-Water, D-Drain, E-Electrical, A-Air). MOD EL Wall Openi ng Length Fascia extends 1” beyond wall opening on each side overlapping the wall and sealing the opening. Allow minimum 2” clearance at rear of recessed unit only. Cabinet side models have rear clearance built in. Handle projects from the face of the facia to the outer radius of the handle 2.375” for Models AA & A. Models B, C,...
Open the catalog to page 3TECHNICAL DATA SHEET PRIMUS Healthcare - Small Sterilizers DOUBLE DOOR E S VOLUME/DIMENSIONS CHART (CON’T) *Refer to General Arrangement (GA) drawings for details and final connection point to utility services (S-Steam, W-Water, D-Drain, E-Electrical, A-Air). MOD EL Wall Openi ng Length Fascia extends 1” beyond wall opening on each side overlapping the wall and sealing the opening. Handle projects from the face of the facia to the outer radius of the handle 2.375” for Models AA & A. Models B, C, & D the handle projects 4.25”. Contact PRIMUS for alternative wall opening. Standard Left Side...
Open the catalog to page 4ARCHITECTURAL NOTES: UTILITY SERVICES Stub-in utility services within 6’-0” of final connection to sterilizer. Optimum sterilizer performance requires the specified utilities. ELECTRICAL AIR (A) DRAIN (D) STEAM (S) WATER (W) Instrument Air Building Drain System (E) Building Steam Supply Cold Water Supply Building Power . Connection: See Below Mimimum 2" . Pipe Siz e: 3/4" NPT . Pipe Siz e: 3/4" NPT Supply . Quality: Condensate . Temperature: < 70o F . Quality: Dry and oil free . Location: Locate Dedicated Circuit free 97% to 100% . Pressure: 50-70 PSIG . Pressure: 60-80 PSI floor sink...
Open the catalog to page 5TECHNICAL DATA SHEET Use the following pages to custom design your sterilizer by checking the boxes of the configuration and options required for your project. Transfer all items checked to the Technical Worksheet (Page 12) and fax to PRIMUS Sales Department or to your local Sales Representative for a quotation. If you do not find an option or size listed to meet your requirements, please contact our Project Manager for additional information and assistance with a custom design to meet your specifications. SD = Single Door Model Number Notation SD = Single Door DD = Double Door ...
Open the catalog to page 6HEALTHCARE LABORATORY CONFIGURATIONS CONTROLS AND RELATED OPTIONS (Contact PRIMUS for details and performance specifications) PSS500 Control C6 PSS500 microcomputer with eight (8) factory preset cycles; seven (7) programmable and 1 Test cycle are standard. A vertical touchpad contains a Liquid Crystal Display (LCD) mounted on a side-mounted control panel with a thermal printer. Cycle progression is displayed by Light Emitting Diodes (LEDs) on the operator panel and printed, 32 characters per line printing. A second copy of the entire sterilization cycle printable at the end of the run....
Open the catalog to page 7All Spire Integrated Solutions catalogs and technical brochures
-
CWC Series
4 Pages
-
Surgical Scrub Sinks
2 Pages
-
STERIPLUS™ 40
2 Pages
-
Tesalys
8 Pages
-
Cadaver Handling Solutions
16 Pages
-
Model DEM Dyna-Poise™
2 Pages
-
Model DPEM
2 Pages
-
Medicine Station
2 Pages
-
NSM5 Specifications
2 Pages
-
Model D Deluxe Autopsy Table
2 Pages
-
MC2-B-MT1
2 Pages
-
3SC General Specifications
2 Pages
-
2SC General Specifications
2 Pages
-
1SC General Specifications
2 Pages
-
3SP1W General Specifications
2 Pages
-
Model DPV
2 Pages
-
DP AUTOPSY TABLE
2 Pages
-
R1910-31L/R
2 Pages
-
SR1910-30
2 Pages
-
REMOTE MORGUE
2 Pages
-
6R2W SPECS
2 Pages
-
4R2W Specs
2 Pages
-
3R1W SUBMITTAL
1 Pages
-
2R1W SUBMITTAL
1 Pages
-
1R SUBMITTAL
1 Pages