MIMI

MIMI

MIMI

Product catalog summary
Isocenter Alignment IGRT QA
Ensures the coincidence of the isocenter prescribed by lasers, treatment beam, and image guidance systems. This is crucial for accurate patient positioning in radiation treatment.
Image Guidance Accuracy
Modern linear accelerators require precise patient positioning. Verifying that all positioning systems describe the same isocentric point as the treatment beam is essential.
Fast, Accurate Alignment Checks
The MIMI Phantom from Standard Imaging allows for quick verification of the isocenter's mechanical stability across various guidance systems (MV, kV, CBCT) within a 1mm tolerance. This test can be performed daily or weekly in minutes, ensuring accurate radiation delivery.
Design and Features
The MIMI Phantom includes five bone-equivalent rods intersecting at 90-degree angles for easy 2D/2D and 3D/3D matching. It features a 6.4 mm high contrast sphere for alignment checks and is compatible with the Varian FramelessArray Optical Guidance System for additional testing.
Specifications
Dimensions: 5.5 in (14 cm) in height, width, and length. Weight: 8.25 lbs (3.75 kg). Made from acetal copolymer with black PVC rods and stainless steel spheres.
Additional Features
Includes pre-drilled holes for the Varian FramelessArray Optical Guidance System localizer and offset cross-hair markers for verifying automatic table adjustments.
Performance Validation
Developed in collaboration with experts from the Netherlands Cancer Institute, the MIMI Phantom is validated for testing integrated system accuracy, including 3D Cone Beam registration, MV/kV isocentricity, lasers, couch table adjustments, and optical guidance systems.
See more

Catalog excerpts

MIMI-1

• ISOCENTER ALIGNMENT IGRT QA Test the coincidence of the isocenter prescribed by the lasers, treatment beam, and image guidance systems • ENSURE IMAGE GUIDANCE ACCURACY Treatment with radiation on a modern linear accelerator relies on the ability to position the patient with incredible accuracy. This is why it is critically important to verify that the systems used for positioning are all describing the same isocentric point in space as the treatment beam. • FAST, ACCURATE ALIGNMENT CHECKS The MIMI (Multiple Imaging Modality Isocentricity) Phantom from Standard Imaging allows you to establish mechanical stability of the image guidance system by verifying the isocenter described by the MV, kV, CBCT, and other guidance systems is within accepted 1mm A 2D/2D matching example with the MIMI Phantom tolerances. This daily or weekly test can be performed in just a few minutes giving you the confidence that you can deliver the radiation prescription on target. A 6.4 mm sphere at phantom center assists with virtual and physical graticule alignment checks. • EASY ALIGNMENT DUE TO UNIQUE DESIGN The MIMI Phantom incorporates five bone equivalent rods uniquely set so that four of them intersect at 90 degree angles when viewed in DRRs or a 2D projection image. The rods traverse the entire phantom making them visible in any image or slice allowing for easy 2D/2D and 3D/3D matching for fast verification of isocenter position. A 3D/3D matching example with the MIMI Phantom

 Open the catalog to page 1
MIMI-2

Additional Features Varian FramelessArray™ Optical Guidance System QA The MIMI Phantom features pre-drilled holes which precisely fit the Varian FramelessArray Optical Guidance System localizer. By fitting the localizer to the phantom, additional testing can be done to verify the isocenter prescribed by the optical guidance system is coincidental to the lasers and the treatment beam isocenter. Test Automatic Table Adjustments The MIMI Phantom features additional cross-hair markers that are offset known distances from the true isocenter. Setting up the phantom aligned to these offsets allows you...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.