• DOSIMETRY, ADVANCED An intuitive user-interface and a host of innovative features make the SuperMAX the premier reference-grade electrometer on the market.
Open the catalog to page 1The Premier, Reference-Grade Electrometer The SuperMAX Electrometer is the culmination of one simple notion – surpass expectations for a reference-grade electrometer. Each aspect of the SuperMAX has been engineered to integrate seamlessly into clinical quality assurance, forming the preeminent dosimetry tool available. • Superior Accuracy and Stability Standard Imaging’s legacy of accuracy, and rounds of rigorous testing, ensure the SuperMAX exceeds requirements for reference grade instruments. This exceptional stability allows for measurement after only one minute of warm-up time. • Touchscreen...
Open the catalog to page 2SuperMAX Electrometer shown with the Exradin A19 Ion Chamber and the Exradin A10 Ion Chamber
Open the catalog to page 3Two Independent Channels The SuperMAX Electrometer has two measurement channels with independent control over range, bias voltage and applied system factors. Both channels have an extensive range (0.001 pA to 500.00 nA, 0.001 pC to 999.9 pC) with automatically scaling units, ideal for a spectrum of applications, including: • Cross calibration between two chambers • Isocenter versus off-axis comparisons • In-air versus in-water comparisons These channels can be viewed individually in full screen or together in a split-screen interface, with the option to display the ratio of the two channels....
Open the catalog to page 4Flexible Collection Modes The SuperMAX Electrometer has three charge collection modes, facilitating data acquisition in a variety of clinical applications. • Timed Charge Collection: obtain measurements in 1 second intervals from 1 second – 24 hours. Perform these measurements sequentially without re-zeroing the electrometer. • Continuous Charge Collection: Manual start/stop measurement for an unlimited duration. • Triggered Charge Collection: Automatically start, stop and save measurements at custom thresholds for high and low ranges. This mode is ideal for external beam measurements. Detector...
Open the catalog to page 5Easily Save and Export Data As charge collections are completed, measurements are automatically stored in a time- and date- stamped list. Rate or dose rate measurements can be added manually by tapping the “Add Measurement” button. At any time, this list can be exported to the USB flash memory in .csv or .txt files, transferred to a PC and opened in Microsoft Excel or other spreadsheet applications. Exradin W1 Scintillator Integration The Exradin W1 Scintillator is a new detector with characteristics that closely mimic water, negating many measurement corrections required with other detectors....
Open the catalog to page 6The SuperMAX can store over 100 sources for quick projected strength calculations. Enter the known activity on a given date and the Check Source Utility can display either the current strength or calculate future strength. DISPLAY RANGE RATE: Check Source Delete Source Date SUPERMAX (REF 9001 8) SPECIFICATIONS Low Range 0.001 pA - 500.0 pA, 1 fA resolution High Range 0.001 nA - 500.0 nA, 1 pA resolution CHARGE: Low Range 0.001 pC - 999.9 jC, 1 fC resolution High Range 0.001 nC - 999.9 jC, 1 pC resolution RESOLUTION High Range: 0.001nA IEC 60731 (Reference Class) CHARGE COLLECTIONS MEASURING...
Open the catalog to page 73120 Deming Way Middleton WI 53562-1461 USA © 2016 Standard Imaging, Inc.
Open the catalog to page 87 Pages