Calibration Procedure for Tango® Exercise Stress BP Monitor
1. Access Calibration Mode: Use the 'VERIFY CALIBRATION' mode found under 'UTILITIES' in the main menu.
2. Connect Equipment: Attach a mercury manometer with a bulb to the pneumatic connector using a T-connector (SunTech part #750 0018 ATA).
3. Verify Cuff Pressure: Ensure the cuff pressure displayed is within ±2mmHg of the mercury manometer's pressure value across a range of 0-300mmHg. If discrepancies are found, contact SunTech for calibration assistance.
Recommendation: It is advised to calibrate the monitor annually to maintain the accuracy of pressure transducers and indicators.
See more

Catalog excerpts

Tango-1

It is recommended that the monitor be calibrated annually to verify the accuracy of the pressure transducers and indicators. 1. USE THE “VERIFY CALIBRATION” MODE in the main menu view under “UTILITIES.” 2. CONNECT A MERCURY MANOMETER with a bulb to the pneumatic connector using a T-connector, SunTech part # 750 0018 ATA. 3. VERIFY THAT THE CUFF PRESSURE on the display is within ±2mmHg of the pressure value on the mercury manometer through a range of pressures from 0-300mmHg. If not, contact SunTech about calibration.