DC-5110/DC-5110RS

DC-5110/DC-5110RS

DC-5110/DC-5110RS

Product catalog summary
Overview: The document provides detailed specifications and features of the WTW and Mettler Toledo Dissolved Oxygen (D.O.) sensors, specifically models DC-5110 and DC-5110RS. These sensors are designed for applications in water and wastewater treatment, fermentation, industrial processes, and environmental monitoring.
Key Features:
  • Waterproof and dustproof design with IP 65 (NEMA 4X) certification.
  • Compatible with WTW TriOxmatic and Mettler Toledo InPro D.O. sensors.
  • Measurement units include %, ppm, and mg/L with salinity and atmospheric pressure correction.
  • User-friendly interface with a large LCM display featuring auto-sense backlight and contrast settings.
  • Automatic temperature compensation and sensor self-diagnosis for TriOxmatic 700 D.O. sensor.
  • Two sets of isolated current output (DC 0/4~20mA) and selectable contact outputs with programmable set-points.
  • Enhanced security with password protection and multi-key design.
  • Automatic sensor cleaning function with programmable intervals.
Specifications:
  • Measurement range: 0-600% for %, 0-60.00 mg/L for mg/L, 0-60.00 ppm for ppm, and 0-140°C for temperature.
  • Resolution: 0.1% for %, 0.01/0.001 mg/L for mg/L, 0.01/0.001 ppm for ppm, and 0.1°C for temperature.
  • Accuracy: ±0.5%±1Digit for %, mg/L, ppm, and ±0.2°C±1Digit for temperature.
  • Temperature compensation options: NTC 30K or NTC 22K automatic or manual adjustment.
  • Salinity correction range: 0.0 ~ 45.0 ppt.
  • Pressure correction range: 0.500 ~ 2.500 bar / 7.25 ~ 36.25 psi / 500~2500 mmHg.
  • Ambient temperature range: 0~50°C; Storage temperature range: -10~70°C.
  • Power supply: 100~240VAC ±10%, 5W max., 50/60Hz.
  • Installation options: Wall, pipe, or panel mounting.
  • Dimensions: 96 mm × 96 mm × 132 mm (H×W×D); Weight: 0.5Kg.
Model DC-5110RS Additional Features:
  • Selectable figure or chart display for measuring values.
  • Trace mode for displaying historical measurement curves from 3 minutes to 4 weeks.
  • Equipped with a logbook function and RS-485 interface.
Optional Accessories:
  • D.O. sensor, membrane kit, electrolyte, cleaning solution.
  • Immersive protection holder (1M, 2M, 3M), junction box, extension cable.
  • Auto clean kit, stainless holder support base.
See more

Catalog excerpts

DC-5110/DC-5110RS-1

Applications: Water and wastewater treatment Fermentation Industrial Process Environment Features: Waterproof and dustproof design 96 x 96mm microprocessor D.O. transmitter Compatible with WTW TriOxmatic D.O. sensors or Mettler Toledo InPro D.O. sensors Measuring unit: %, ppm, mg/l Salinity correction & Atmospheric pressure correction User-friendly operation with text and graphic illustration, clear signs and easy to understand Large LCM display with auto-sense backlight and contrast setting function Sensor slope performance display Sensor self diagnosis for TriOxmatic 700 D.O. sensor Simultaneous display of measurement value and temperature Automatic temperature compensation Two sets of isolated current output DC 0/4~20mA, Max.500Ω Hi/Lo, Hi/Hi, Lo/Lo selectable two limited contact outputs with programmable set-points and hysteresis With password and multi-key design, enhanced security protection Automatic sensor clean function with programmable time interval LED indicated alarm or washing status signal IP 65 (NEMA 4X) DC-5110RS only: Measuring value can be selected for figure or chart display Trace mode can display historical measured value curve from 3-minute to 4-week program Equipped with logbook function Equipped with one analog output and RS-485 interface Mettler Toledo D.O. sensor

 Open the catalog to page 1
DC-5110/DC-5110RS-2

Intelligent Dissolved Oxygen Transmitters DC-5110/DC-5110RS Specifications: Model Measuring mode % mg/L Measuring range ppm Temp. % mg/L Resolution ppm Temp. % mg/L Accuracy ppm Temp. Temperature Compensation Salinity correction Pressure correction Ambient temperature Storage temperature % / ppm / mg/L / TEMP. 0 ~ 600% (Depends on the sensor) 0~60.00 mg/L (Depends on the sensor) 0~60.00 ppm (Depends on the sensor) 0~140°C (Depends on the sensor) 0.1 % 0.01 mg/L / 0.001 mg/L 0.01 ppm / 0.001 ppm 0.1℃ ±0.5%±1Digit ±0.2°C±1Digit with temperature error correction function NTC 30K or NTC 22K automatic...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.