Monitor AMI Codes-II O3

Monitor AMI Codes-II O3

Monitor AMI Codes-II O3

Product catalog summary
Overview: The Monitor AMI Codes-II O3 is a comprehensive system designed for the automatic and continuous measurement of ozone in pharmaceutical and industrial high purity water applications. It utilizes the DPD colorimetric method (DIN 38404-3) for ozone determination.
Specifications:
  • Measuring Range: 0 to 500 ppb with an accuracy of ±1 ppb or 5%, whichever is greater.
  • Limit of Detection: 1 ppb.
  • Cycle Time: 5 to 10 minutes (default: 6 minutes).
  • Transmitter Case: Aluminum with IP 66 / NEMA 4X protection.
  • Display: Backlit LCD, 75 x 45 mm.
  • Power Supply: AC version: 100 - 240 VAC, DC version: 10-36 VDC.
  • Operating Temperature: -10 to +50 °C, with a limit range of -25 to +65 °C.
Features:
  • Includes measurement and control electronics, photometer, flow indicator, reaction chamber, reagent dosing system, and reagent containers.
  • Compatible with usual dosing devices for disinfectants via relays or analog output signals.
  • Automatic interruption of disinfectant dosing with an external signal.
  • Alarm display and relay activation when critical limits are reached.
  • Continuous monitoring of main instrument functions.
  • Large LCD display for simultaneous viewing of measured values and status information.
  • Specific features for the pharmaceutical industry include Audit Trail, Performance Verification, and a Validation Package.
Accessories and Options:
  • Validation Package (IQ, OQ, PQ) and Verification Kit available.
  • Communication interfaces: Profibus DP & Modbus RTU, USB, and HART.
Operation and Safety:
  • Easy operation with separate menus for different functions and multilingual user menus.
  • Data storage for event logs, alarm logs, and calibration history.
  • Safety features include data retention after power failure and over-voltage protection.
  • Programmable alarm limits and real-time clock for action timestamping.
Control and Communication:
  • Relays and current outputs programmable for dosing pumps, solenoid valves, or motor valves.
  • Communication interface options include RS485 with Modbus RTU or Profibus DP, USB, and HART.
Sample Conditions:
  • Water consumption: Minimum 10 l/h.
  • Water inlet pressure: 0.15 to 2 bar.
  • Sample temperature: 5 to 50 °C.
Physical Specifications:
  • Flow cell made of acrylic glass with needle valve.
  • Panel dimensions: 400 x 850 x 200 mm, made of stainless steel, weighing 14.0 kg.
See more

Catalog excerpts

Monitor AMI Codes-II O3-1

SWAN Analytische Instrumente AG CH-8340 Hinwil/Switzerland Tel. +41 44 943 63 00 [email protected] . www.swan.ch Monitor AMI Codes-II O3 Data sheet no. DenA2544X3000 Complete monitoring system for the automatic, continuous measurement of ozone in pharma water and industrial high purity water applications. Monitor AMI Codes-II O3  For the continuous online determination of ozone based on the DPD colorimetric method (DIN 38404-3).  Measuring range: 0 to 500 ppb  Complete system including measurement and control electronics, photometer, flow indicator, reaction chamber, reagent dosing system and reagent containers.  All usual dosing devices for disinfectants control can be connected either through relays or analog output signals. Two independent controllers can operate simultaneously.  Dosing of disinfectant can be interrupted automatically with an external signal.  Alarm display and activation of alarm relay when user defined, critical limits for a measurement value are reached.  Continuous, automatic monitoring of main instrument functions (dirty photometer, sample flow, reagents).  Large back-lit LCD display showing all measured values and status information simultaneously.  Factory tested, ready for installation and operation.  Specific features for pharmaceutical industry: Audit Trail, Performance verification and a Validation Package. Accessories:  Validation Package (IQ, OQ, PQ)  Verification Kit Instrument Option:  Communication interface Monitor AMI Codes-II O3 DC Option: [ ] Profibus DP & Modbus RTU interface (RS-485) [ ] HART interface [ ] Validation Package 06/2020 Subject to change without notice

 Open the catalog to page 1
Monitor AMI Codes-II O3-2

SWAN Analytische Instrumente AG CH-8340 Hinwil/Switzerland Tel. +41 44 943 63 00 [email protected] . www.swan.ch Ozone measurement Measuring range: 0 – 500 ppb Limit of detection Cycle time: Monitor AMI Codes-II O3 Data sheet no. DenA2544X3000 Electrical Connection Scheme Accuracy:  1 ppb or 5% (whichever is greater) 1 ppb 5 – 10 min. (default: 6 min) Transmitter Specifications and Functionality Electronics case: Aluminum Protection degree: IP 66 / NEMA 4X Display: backlit LCD, 75 x 45 mm Electrical connectors: screw clamps Ambient temperature: -10 to +50 °C Limit range of operation: -25 to +65 °C...

 Open the catalog to page 2

All Swan Analytical Instruments catalogs and technical brochures

  1. AMI INSPECTOR

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.